Ready-to-use liquid glucagon autoinjector (Gvoke HypoPen, as named in the record) versus the glucagon emergency kit: non-inferior rescue from severe hypoglycaemia, with preparation and injection taking 27 seconds versus 97 (161 adults, two crossover studies)
Shelf #10010 · Diabetes, glucose & metabolism
POSSIBLE (benefit)
Stamped: 2026-11-28.
Evidence published: 2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults with type 1 diabetes and insulin-induced severe hypoglycaemia, does a ready-to-use liquid glucagon autoinjector rescue as reliably as the conventional glucagon emergency kit that must be reconstituted?
Balance of the evidence
POSSIBLE (benefit) — Composition: two phase 3 randomised blinded crossover non-inferiority studies in 161 adults with type 1 diabetes (Christiansen 2021), comparing 1 mg from the glucagon autoinjector (Gvoke HypoPen — glucagon injection, Xeris Pharmaceuticals, as named in the record) with the glucagon emergency kit in insulin-induced severe hypoglycaemia. The autoinjector was non-inferior for successful plasma glucose recovery within 30 minutes and for glucose above 3.9 mmol/L (70 mg/dL) or neuroglycopenic symptom relief; mean time from administration to recovery was 13.8 minutes for the autoinjector versus 10.0 minutes for the kit, but preparation and administration took 27.3 seconds versus 97.2 seconds (p < 0.0001), which is where the ready-to-use device gains its time. Treatment-emergent adverse events were comparable. This shelf is distinct from live #954 (nasal glucagon) and live #1777 (dasiglucagon, a different molecule): the question here is the conventional glucagon molecule in a ready-to-use liquid autoinjector versus its own reconstituted kit. POSSIBLE (benefit): equally reliable rescue with a materially simpler, faster-to-give injection, from one industry programme of induced-hypoglycaemia studies.
MeSH: Blood Glucose · Cross-Over Studies · Glucagon · Humans · Hypoglycemia · Insulin
Cards (2)
Card 1 — Glucagon autoinjector non-inferior to the emergency kit; preparation 27.3 s versus 97.2 s (Christiansen 2021)
Study: Two phase 3 randomised crossover non-inferiority studies (N = 161)
Population: 161 adults with type 1 diabetes across two phase 3 randomised controlled blinded crossover non-inferiority studies, each treated for insulin-induced severe hypoglycaemia with both 1 mg glucagon by autoinjector (Gvoke HypoPen, as named in the record) and 1 mg from a glucagon emergency kit, in crossover.
Summary: Glucagon autoinjector non-inferior to the emergency kit; preparation 27.3 s versus 97.2 s (Christiansen 2021)
Result: Successful plasma glucose recovery within 30 minutes — return to above 3.9 mmol/L (70 mg/dL) or a rise of at least 1.1 mmol/L (20 mg/dL) from a baseline below 2.9 mmol/L (50 mg/dL) — was non-inferior for the autoinjector versus the kit, as was recovery defined by glucose above 3.9 mmol/L (70 mg/dL) or neuroglycopenic symptom relief. Mean time from administration to recovery was 13.8 (SD 5.6) minutes for the autoinjector versus 10.0 (3.6) minutes for the kit, while preparation and administration took 27.3 (19.7) seconds versus 97.2 (45.1) seconds (p < 0.0001). Treatment-emergent adverse events were comparable.
Watch out: Hypoglycaemia was induced and treated in a clinic, with trained handling of both devices — the real-world advantage of the autoinjector depends on lay rescuers, whose performance this record does not test. Once the dose is in, the kit was numerically faster to raise glucose; the device's gain is entirely in preparation.
Source: Christiansen MP, Cummins M, Prestrelski S et al.. Comparison of a ready-to-use liquid glucagon injection administered by autoinjector to glucagon emergency kit for the symptomatic relief of severe hypoglycemia: two randomized crossover non-inferiority studies. BMJ open diabetes research & care. 2021. PMID: 34620618. Registry ID: NCT02656069 and NCT03439072. MeSH: Blood Glucose · Cross-Over Studies · Glucagon · Humans · Hypoglycemia · Insulin
Card 2 — Synthesis: same molecule, same rescue — the benefit is the minute saved before the injection, not after it
Study: Synthesis (2 crossover studies in 1 programme, N = 161)
Population: The 161 adults of Christiansen 2021; no other study is carded on this shelf.
Summary: Synthesis: same molecule, same rescue — the benefit is the minute saved before the injection, not after it
Result: Both delivery routes rescued reliably within 30 minutes. The autoinjector's case rests on removing reconstitution: about 70 seconds saved in preparation and administration, against about 4 minutes slower mean glucose recovery after administration in these clinic studies. Net time from deciding to treat to recovery therefore favours neither decisively in trained hands.
Watch out: A single manufacturer's programme (the product and its maker are named in the record, as printed on the card); independent replication and lay-user studies are outside this record. Children were not studied here.
Source: Christiansen MP, Cummins M, Prestrelski S et al.. Comparison of a ready-to-use liquid glucagon injection administered by autoinjector to glucagon emergency kit for the symptomatic relief of severe hypoglycemia: two randomized crossover non-inferiority studies. BMJ open diabetes research & care. 2021. PMID: 34620618. Registry ID: NCT02656069 and NCT03439072. MeSH: Blood Glucose · Cross-Over Studies · Glucagon · Humans · Hypoglycemia · Insulin