Point-of-care lactate against laboratory lactate: handheld and blood-gas devices read slightly low (mean bias about −0.2 to −0.4 mmol/L, about −1.8 to −3.6 mg/dL) with 94–100% concordance at the sepsis cut-offs — but the negative bias grows proportionally above 6 mmol/L (54 mg/dL)

Shelf #10038 · Emergency & first hours

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2010–2018.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do point-of-care whole-blood lactate devices (handheld meters, cartridge and blood-gas analysers) measure lactate against central-laboratory plasma lactate and a mass-spectrometry reference — what bias, concordance at the 2.3 mmol/L (about 21 mg/dL) and 4 mmol/L (about 36 mg/dL) clinical cut-offs, and prognostic agreement do they show?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: 3 paired method-comparison studies — a three-device laboratory evaluation against plasma methods and a flow-injection mass-spectrometry method, using donor specimens whose count the record does not state (Tolan 2017; N-not-stated), a handheld-meter comparison in 50 septic intensive-care patients against four central-laboratory analysers (Léguillier 2018), and a 699-patient emergency-department cohort with paired point-of-care and laboratory lactate (Shapiro 2010). The documented patient base is 749, above the 30 minimum. No sensitivity/specificity pair arises — these are agreement studies — so none is reported. Every mmol/L value carries its mg/dL equivalent: mean bias ranged from −0.2 ± 0.9 mmol/L (about −1.8 mg/dL) to −0.4 ± 1.2 mmol/L (about −3.6 mg/dL) by device (Tolan 2017), and point-of-care read on average 0.32 mmol/L (about 2.9 mg/dL) below laboratory values with agreement kappa 0.97 (Shapiro 2010). Concordance at the working cut-offs was high — 94–100% within the reference interval (below 2.3 mmol/L, below 20.7 mg/dL) and above 4 mmol/L (36 mg/dL) (Tolan 2017); 98–100% against laboratory analysers in sepsis (Léguillier 2018) — but all devices showed proportional negative bias above 6 mmol/L (54 mg/dL) against plasma methods, a bias that disappeared against the mass-spectrometry whole-blood method, pointing at the plasma comparison itself. Devices named as published, under the tested-product clarification exception: Nova StatStrip, Abbott i-STAT CG4+ and Radiometer ABL90 (Tolan 2017); StatStrip Xpress Meter (Léguillier 2018); i-STAT (Shapiro 2010). MEASUREMENT ACCURACY — PARTIAL: interchangeable with the laboratory around the decision cut-offs, but a point-of-care value is a slightly low reading, and the gap widens exactly where lactate is highest.

MeSH: Blood Chemical Analysis · Humans · Lactic Acid · Point-of-Care Systems · Reference Standards · Sepsis · Female · Intensive Care Units · Male · Adult · Aged · Area Under Curve · Emergency Service, Hospital · Feasibility Studies · Middle Aged · Prognosis · Prospective Studies

Cards (3)

Card 1 — Three devices vs plasma and mass spectrometry: bias −0.2 to −0.4 mmol/L (−1.8 to −3.6 mg/dL), proportional under-reading above 6 mmol/L (54 mg/dL) (Tolan 2017)

Study: Paired method-comparison study (N-not-stated)

Population: Laboratory evaluation of three point-of-care whole-blood lactate methods (Nova StatStrip, Abbott i-STAT CG4+ and Radiometer ABL90 — as named in the record) against the mean of two plasma lactate methods and against a flow-injection mass-spectrometry assay, using specimens from normal donors at rest, after exertion and after spiking. The record does not state the number of donors or specimens — N-not-stated.

Summary: Three devices vs plasma and mass spectrometry: bias −0.2 to −0.4 mmol/L (−1.8 to −3.6 mg/dL), proportional under-reading above 6 mmol/L (54 mg/dL) (Tolan 2017)

Result: Intra- and inter-assay coefficients of variation were below 5% for i-STAT and ABL90 and 3.1–8.2% for StatStrip. Mean (± SD) bias between point-of-care and plasma lactate ranged from −0.2 ± 0.9 mmol/L (about −1.8 mg/dL) for i-STAT and ABL90 to −0.4 ± 1.2 mmol/L (about −3.6 mg/dL) for StatStrip. Above 6 mmol/L (54 mg/dL) all point-of-care methods showed proportional negative bias against plasma methods, but not against the mass-spectrometry method. Concordance with plasma lactate was 94–100% within the reference interval (below 2.3 mmol/L, below 20.7 mg/dL) and above 4 mmol/L (36 mg/dL).

Watch out: Donor and spiked specimens, not patients: the high-range negative bias may be a plasma-processing artefact (it vanishes against whole-blood mass spectrometry), so the study locates the disagreement without settling which side is wrong.

Source: Tolan NV, Wockenfus AM, Koch CD, Crews BO et al.. Analytical performance of three whole blood point-of-care lactate devices compared to plasma lactate comparison methods and a flow-injection mass spectrometry method. Clinical biochemistry. 2017. PMID: 27836621. Registry ID: not stated in the record. MeSH: Blood Chemical Analysis · Humans · Lactic Acid · Point-of-Care Systems · Reference Standards · Sepsis

Card 2 — Handheld meter in 50 septic patients: correlations R² 0.96–0.98, slopes 0.83–0.90, concordance 98–100% (Léguillier 2018)

Study: Paired method-comparison study (N = 50)

Population: 50 identified septic patients admitted to intensive care, sampled in a mobile intensive care unit setting; capillary point-of-care lactate on the StatStrip Xpress Meter (as named in the record) compared with four central-laboratory analysers (one whole-blood, three plasma-based).

Summary: Handheld meter in 50 septic patients: correlations R² 0.96–0.98, slopes 0.83–0.90, concordance 98–100% (Léguillier 2018)

Result: The device showed repeatability coefficients of variation below 3.8% and reproducibility below 5.0%. Correlation with the central-laboratory analysers was R² 0.96–0.98, with regression slopes of 0.83–0.90 and intercepts of 0.02–0.03, and concordance within clinically relevant lactate ranges was 98–100%. The authors concluded the values are comparable and transferable between point-of-care and central laboratory.

Watch out: Slopes of 0.83–0.90 mean the meter reads progressively lower as lactate climbs, even while range concordance stays near-perfect: agreement at cut-offs and agreement in level are different claims, and this study supports the first more than the second.

Source: Léguillier T, Jouffroy R, Boisson M, Boussaroque A et al.. Lactate POCT in mobile intensive care units for septic patients? A comparison of capillary blood method versus venous blood and plasma-based reference methods. Clinical biochemistry. 2018. PMID: 29522709. Registry ID: not stated in the record. MeSH: Blood Chemical Analysis · Female · Humans · Intensive Care Units · Lactic Acid · Male · Point-of-Care Systems · Sepsis

Card 3 — 699 emergency patients: point-of-care lactate 0.32 mmol/L (2.9 mg/dL) below laboratory, kappa 0.97, mortality AUC 0.72 vs 0.70 (Shapiro 2010)

Study: Prospective paired-measurement cohort (N = 699)

Population: 699 adult emergency-department patients with suspected infection (34 deaths, 4.9%), each with a point-of-care lactate (i-STAT, Abbott Point of Care — as named in the record) and a concomitant central-laboratory lactate.

Summary: 699 emergency patients: point-of-care lactate 0.32 mmol/L (2.9 mg/dL) below laboratory, kappa 0.97, mortality AUC 0.72 vs 0.70 (Shapiro 2010)

Result: Point-of-care lactate was on average 0.32 mmol/L (about 2.9 mg/dL) lower than laboratory lactate (95% confidence interval −0.35 to 0.98 mmol/L, about −3.2 to 8.8 mg/dL), with agreement kappa = 0.97. Areas under the curve for mortality prediction were 0.72 for point-of-care lactate and 0.70 for laboratory lactate (pH 0.60, base excess 0.60).

Watch out: A single-device, single-centre deployment with standardised training and quality assurance: the near-perfect kappa describes a well-run programme, and the small average bias still conceals wide individual limits of agreement.

Source: Shapiro NI, Fisher C, Donnino M, Cataldo L et al.. The feasibility and accuracy of point-of-care lactate measurement in emergency department patients with suspected infection. The Journal of emergency medicine. 2010. PMID: 19717267. Registry ID: not stated in the record. MeSH: Adult · Aged · Area Under Curve · Emergency Service, Hospital · Feasibility Studies · Female · Humans · Lactic Acid · Male · Middle Aged · Point-of-Care Systems · Prognosis · Prospective Studies · Sepsis