Subacromial balloon spacer for irreparable rotator cuff tears: matched partial repair in the US trial (ASES success 83% vs 81%) — but lost to debridement alone in the UK trial (Oxford Shoulder Score 30.3 vs 34.3)
Shelf #10053 · Bones & joints
SPLIT
Stamped: 2026-11-28.
Evidence published: 2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults with irreparable posterosuperior massive rotator cuff tears that have failed non-operative care, does inserting a subacromial balloon spacer (the InSpace implant named in both trials) improve pain and shoulder function compared with arthroscopic partial repair or with arthroscopic debridement alone?
Balance of the evidence
SPLIT — Composition: two multicentre randomised trials of the same device that answer against different comparators and point opposite ways — a 184-patient US trial against arthroscopic partial repair (Verma 2022) and a 117-patient UK trial against arthroscopic debridement with biceps tenotomy (Metcalfe 2022, START:REACTS). The documented base is 301 randomised patients, above the 30 minimum. Against partial repair, the balloon matched the operation on the primary score at 12 and 24 months, with faster early recovery and a shorter operation (Verma 2022). Against debridement alone, the balloon made patients worse on the primary score at 12 months — Oxford Shoulder Score 30.3 versus 34.3, adjusted difference −4.2 favouring control — and the trial stopped early at a predefined boundary (Metcalfe 2022). The split is therefore comparator-shaped: the device looks equivalent to one real operation and inferior to a simpler one. Boundary: superior capsular reconstruction, tendon transfers and reverse arthroplasty for irreparable tears are different operations, not this question. SPLIT: two adequate trials of one device disagree by comparator — non-inferior to partial repair, worse than debridement alone — so no single benefit verdict is honest.
MeSH: Arthroscopy · Humans · Pain · Range of Motion, Articular · Retrospective Studies · Rotator Cuff Injuries · Treatment Outcome · Female · Male · Muscle, Skeletal · Shoulder · Shoulder Pain
Cards (3)
Card 1 — Balloon spacer versus partial repair: ASES success 83% versus 81% at the clinically important threshold, with faster early recovery and shorter operations (Verma 2022)
Study: Multicentre single-blind randomised controlled trial (N = 184)
Population: 184 patients aged 40 or older with symptomatic, irreparable, posterosuperior massive rotator cuff tears, an intact subscapularis and failed non-operative management, randomised at 20 sites to the InSpace implant (93) or arthroscopic partial repair (91), followed for 24 months.
Summary: Balloon spacer versus partial repair: ASES success 83% versus 81% at the clinically important threshold, with faster early recovery and shorter operations (Verma 2022)
Result: Both groups improved significantly on the American Shoulder and Elbow Surgeons (ASES) score at 12 months, maintained at 24 months. Overall, 83% of the InSpace group and 81% of the partial-repair group reached the ASES minimally clinically important difference, and 82% versus 79% reached the substantial clinical benefit threshold. Forward elevation was significantly greater with the implant at day 10, week 6, month 12 and month 24, and operative time was significantly shorter (p < 0.0001). No device-related surgical complications were noted; 4 reoperations followed implantation and 3 followed partial repair.
Watch out: Single-blinded, industry-device trial against the more invasive of the two comparators used in this literature: equivalence to partial repair does not answer whether the device beats doing less, which is the question the UK trial asked.
Source: Verma N, Srikumaran U, Roden CM, Rogusky EJ, Lapner P et al.. InSpace Implant Compared with Partial Repair for the Treatment of Full-Thickness Massive Rotator Cuff Tears: A Multicenter, Single-Blinded, Randomized Controlled Trial. The Journal of bone and joint surgery. American volume. 2022. PMID: 35777921. Registry ID: not stated in the record. MeSH: Arthroscopy · Humans · Pain · Range of Motion, Articular · Retrospective Studies · Rotator Cuff Injuries · Treatment Outcome
Card 2 — Balloon spacer versus debridement alone (START:REACTS): Oxford Shoulder Score 30.3 versus 34.3 at 12 months — the device group did worse, and the trial stopped early (Metcalfe 2022)
Study: Double-blind multicentre randomised controlled trial, stopped early (N = 117)
Population: 117 participants with an irreparable rotator cuff tear not resolved by conservative treatment, randomised intraoperatively at 24 UK hospitals to arthroscopic debridement with biceps tenotomy alone (61) or the same procedure plus the InSpace balloon (56); primary outcome data were obtained for 114 (97%).
Summary: Balloon spacer versus debridement alone (START:REACTS): Oxford Shoulder Score 30.3 versus 34.3 at 12 months — the device group did worse, and the trial stopped early (Metcalfe 2022)
Result: The mean Oxford Shoulder Score at 12 months was 34.3 (SD 11.1) with debridement only and 30.3 (10.9) with debridement plus device — a mean difference adjusted for the adaptive design of −4.2 (95% CI −8.2 to −0.26; p = 0.037) favouring control. A predefined stopping boundary was met at the first interim analysis and recruitment stopped at 117 randomised. There was no difference in adverse events between the groups. The authors concluded: we do not recommend the InSpace balloon for the treatment of irreparable rotator cuff tears.
Watch out: Early stopping at an interim boundary can exaggerate an effect in either direction; even allowing for that, the point estimate favours the simpler operation, and the burden of proof sits with the device.
Source: Metcalfe A, Parsons H, Parsons N, Brown J, Fox J et al.. Subacromial balloon spacer for irreparable rotator cuff tears of the shoulder (START:REACTS): a group-sequential, double-blind, multicentre randomised controlled trial. Lancet (London, England). 2022. PMID: 35461618. Registry ID: ISRCTN17825590 (stated in the record). MeSH: Arthroscopy · Female · Humans · Male · Muscle, Skeletal · Rotator Cuff Injuries · Shoulder · Shoulder Pain · Treatment Outcome
Estimate type: adjusted estimate, as stated in the card.
Card 3 — Synthesis: one device, two comparators, two verdicts — equivalence to partial repair, inferiority to debridement alone
Study: Synthesis (2 randomised trials, N = 301)
Population: All 301 patients randomised in the two trials carded on this shelf (Verma 2022; Metcalfe 2022). No other randomised trial of the balloon spacer is carded here.
Summary: Synthesis: one device, two comparators, two verdicts — equivalence to partial repair, inferiority to debridement alone
Result: Read together, the trials do not conflict on the device's behaviour so much as on the value of the operations around it: where the alternative was partial repair, adding the balloon changed nothing material by 12 months; where the alternative was debridement alone, adding the balloon left patients about 4 Oxford Shoulder Score points worse. Any claim for the device must therefore specify 'instead of what' — the record supports no claim against the simpler operation.
Watch out: The two trials also differ in blinding (single versus double) and health system; the synthesis treats the comparator difference as the headline because it is the difference the designs themselves built in.
Source: Verma N, Srikumaran U, Roden CM, Rogusky EJ, Lapner P et al.. InSpace Implant Compared with Partial Repair for the Treatment of Full-Thickness Massive Rotator Cuff Tears: A Multicenter, Single-Blinded, Randomized Controlled Trial. The Journal of bone and joint surgery. American volume. 2022. PMID: 35777921. Registry ID: not stated in the record. MeSH: Arthroscopy · Humans · Pain · Range of Motion, Articular · Retrospective Studies · Rotator Cuff Injuries · Treatment Outcome