The 13C-octanoate breath test against scintigraphy: reproducible within a person (coefficient of variation 12%) but not accurate between tests — half-emptying-time differences from −33 to +170 minutes and correlations from none to weak (rs = 0.54)

Shelf #10090 · Liver & gut

MEASUREMENT ACCURACY — POOR

Stamped: 2026-11-28.

Evidence published: 1997–1998.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately does the 13C-octanoic acid breath test measure gastric emptying of solids against simultaneous scintigraphy — its correlation, its half-emptying-time agreement, and its reproducibility?

Balance of the evidence

MEASUREMENT ACCURACY — POOR — Composition: 2 primary validation studies from a single research programme, in healthy volunteers — 15 volunteers with simultaneous scintigraphy and breath testing, each breath test repeated three times within 3 weeks (Choi 1997), and 30 healthy volunteers in a reproducibility and sampling-simplification study against simultaneous scintigraphy (Choi 1998). The two cohorts are carded as reported (documented base 45 people); the programme is one group's, so this is a single-programme grade, flagged here and on the cards. No sensitivity/specificity pair arises — both records report agreement and correlation for a continuous emptying measure, and the pack says so instead of inventing a pair. The agreement fails the substitution test: breath-test and scintigraphic parameters were not correlated significantly in the 15-volunteer study, and the half-emptying-time difference between the tests ranged from −33.1 to +169.6 minutes (mean 48.0 minutes); in the 30-volunteer study the half-time correlation was significant but weak (rs = 0.54, P < 0.005) and the lag phase did not correlate. The paradox is precision without accuracy: within-person reproducibility was high (mean coefficient of variation 12%, and no worse than scintigraphy's own variation), so the breath test tracks a person's change over time while mis-stating the scintigraphic value. Extending sampling from 4 to 6 hours improved the breath test's estimates, at higher cost. Boundary: scintigraphic emptying norms themselves are homed in packs #7260 and #9984, and wireless-capsule transit norms in pack #8628; this pack covers only the breath test's agreement with scintigraphy. MEASUREMENT ACCURACY — POOR: as a substitute for scintigraphy the 13C-octanoate breath test mis-measures the half-emptying time by tens of minutes in either direction, whatever its within-person repeatability.

MeSH: Adult · Aged · Breath Tests · Caprylates · Carbon Isotopes · Evaluation Studies as Topic · Female · Gastric Emptying · Humans · Male · Middle Aged · Radionuclide Imaging · Reproducibility of Results · Stomach · Adolescent · Chromatography, Gas · Linear Models · Models, Biological · Predictive Value of Tests

Cards (2)

Card 1 — Breath-test and scintigraphic emptying did not correlate significantly, and half-time differences ranged from −33 to +170 minutes, in 15 healthy volunteers (Choi 1997)

Study: Validation study against scintigraphy (N = 15)

Population: 15 healthy volunteers (8 men, 7 women; mean age 41 ± 13 years) who underwent simultaneous scintigraphy and 13C-octanoic acid breath testing (scans every 15 minutes for 4 hours, breath samples for 6 hours), with the breath test repeated three times within a 3-week period.

Summary: Breath-test and scintigraphic emptying did not correlate significantly, and half-time differences ranged from −33 to +170 minutes, in 15 healthy volunteers (Choi 1997)

Result: Parameters from scintigraphy and the breath test were not correlated significantly, contrary to the earlier claim that breath-test half-emptying time minus 66 minutes estimates the scintigraphic half-time. Differences between the tests in lag phase and half-emptying time were highly variable (half-time range −33.1 to 169.6 minutes; mean 48.0 minutes). Longer breath sampling (over 4, 5 or 6 hours) yielded decreasing estimates of lag phase and half-time. Repeated breath tests were highly reproducible within individuals (mean coefficient of variation 12%). The authors concluded the test requires further validation before it can substitute for scintigraphy, though it seems useful for intraindividual comparisons.

Watch out: Single-programme evidence in healthy volunteers only: agreement in patients with gastroparesis — the population the test is aimed at — is not established by this pack's records.

Source: Choi, Camilleri, Burton, Zinsmeister et al.. [13C]octanoic acid breath test for gastric emptying of solids: accuracy, reproducibility, and comparison with scintigraphy. Gastroenterology. 1997. PMID: 9097998. Registry ID: not stated in the record. MeSH: Adult · Aged · Breath Tests · Caprylates · Carbon Isotopes · Evaluation Studies as Topic · Female · Gastric Emptying · Humans · Male · Middle Aged · Radionuclide Imaging · Reproducibility of Results · Stomach

Card 2 — With 6-hour sampling the half-time correlation reached only rs = 0.54, and the lag phase did not correlate at all, in 30 healthy volunteers (Choi 1998)

Study: Validation and reproducibility study (N = 30)

Population: 30 healthy volunteers who performed the 13C-octanoic acid breath test with simultaneous scintigraphy, in a study of reproducibility and of whether fewer breath samples could characterise the emptying curve.

Summary: With 6-hour sampling the half-time correlation reached only rs = 0.54, and the lag phase did not correlate at all, in 30 healthy volunteers (Choi 1998)

Result: There was a significant but weak correlation between half-emptying time measured by the two tests (rs = 0.54, P < 0.005), but not between the durations of the lag phase. Differences between the tests differed between subjects but were highly reproducible within subjects, and within- and between-subject variation of the breath test was not significantly different from scintigraphy's. A subset of 11 breath samples over 6 hours (of 24) predicted the cumulative 13CO2 appearance variables with r² above 0.95, but the sampling period should still extend to 6 hours after dosing.

Watch out: Same programme as the 1997 study and again healthy volunteers: the two cards corroborate one laboratory's finding, they are not independent validations.

Source: Choi, Camilleri, Burton, Zinsmeister et al.. Reproducibility and simplification of 13C-octanoic acid breath test for gastric emptying of solids. The American journal of gastroenterology. 1998. PMID: 9448183. Registry ID: not stated in the record. MeSH: Adolescent · Adult · Aged · Breath Tests · Caprylates · Carbon Isotopes · Chromatography, Gas · Female · Gastric Emptying · Humans · Linear Models · Male · Middle Aged · Models, Biological · Predictive Value of Tests · Radionuclide Imaging · Reproducibility of Results · Stomach