SourceMed · Shelf #10093 in the ask interface

Multipoint pacing for CRT non-responders: the trial's own primary endpoint — converting non-responders — did not move in either phase, yet a secondary analysis found fewer heart-failure hospitalisations or deaths

Shelf #10093 · Heart, vessels, blood pressure & clotting

SPLIT

Stamped: 2026-11-28.

Evidence published: 2019–2025.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients who have not responded to six months of conventional biventricular pacing, does switching to multipoint left-ventricular pacing (pacing from two sites on a quadripolar lead) increase echocardiographic response or improve clinical outcomes compared with continuing conventional biventricular pacing?

Balance of the evidence

SPLIT — Composition: one randomised trial programme (MORE-CRT MPP) reported in three papers, plus one systematic review and meta-analysis (7 studies, 1,390 patients; Mehta 2021). The trial programme is a single programme, so this pack is. On the programme's own primary endpoint the answer is consistently no: in Phase I (467 non-responders randomised; Leclercq 2019), conversion to echocardiographic response was 31.8% on multipoint pacing versus 33.8% on continued biventricular pacing (P = 0.72), and in the larger 2023 report (1,111 non-responders randomised) it was 29.4% versus 30.4% (P = 0.743) — in both phases about 30% of non-responders converted simply with more time on conventional therapy (the treatment made no difference to conversion). A pre-specified subgroup with wide (≥30 mm) electrode separation and tight timing (MPP-AS) converted at 45.6% versus 26.2% with other multipoint settings in Phase I (P = 0.006), but the same subgroup showed no significant difference in the larger 2023 report. Against that, a 2025 secondary analysis of the same programme (1,421 patients with complete data) found heart-failure hospitalisation or death in 6.64% on multipoint pacing versus 10.44% on biventricular pacing over about 5 months (P = 0.0107; adjusted odds ratio 0.60). The meta-analysis mirrors the split: echocardiographic benefit appeared in non-randomised studies (odds ratio 5.33) but not in randomised studies (odds ratio 1.86, 95% CI 0.91 to 3.79). Tested-product note: MultiPoint is the named pacing feature of one manufacturer's quadripolar CRT system, named here as published under the tested-product clarification exception; the records name the feature, and results may not transfer to other platforms. SPLIT: multipoint pacing does not convert non-responders on the endpoint the trial was built around, but a secondary clinical-outcome analysis of the same trial favours it — a hypothesis about programming and patient selection, not a demonstrated rescue therapy.

MeSH: Aged · Aged, 80 and over · Cardiac Pacing, Artificial · Cardiac Resynchronization Therapy · Cardiac Resynchronization Therapy Devices · Double-Blind Method · Echocardiography · Electrocardiography · Female · Follow-Up Studies · Heart Failure · Hemodynamics · Humans · Male · Middle Aged · Prospective Studies · Surveys and Questionnaires · Treatment Outcome · Ventricular Function, Left · Hospitalization · Risk Factors · Time Factors · Cause of Death · Stroke Volume

Cards (4)

Card 1 — MORE-CRT MPP Phase I (467 non-responders randomised): conversion to response 31.8% on multipoint pacing versus 33.8% on continued biventricular pacing (P = 0.72) — the wide-separation subgroup did better (Leclercq 2019)

Study: Randomised trial, Phase I of MORE-CRT MPP (N = 467 randomised)

Population: Patients receiving a quadripolar CRT system capable of multipoint pacing (1,921 enrolled); those with less than 15% reduction in left-ventricular end-systolic volume after 6 months of biventricular pacing were randomised to multipoint pacing (236) or continued biventricular pacing (231) for a further 6 months, with multipoint programming at physician discretion.

Summary: MORE-CRT MPP Phase I (467 non-responders randomised): conversion to response 31.8% on multipoint pacing versus 33.8% on continued biventricular pacing (P = 0.72) — the wide-separation subgroup did better (Leclercq 2019)

Result: No difference in non-responder-to-responder conversion (31.8% versus 33.8%, P = 0.72). Within the multipoint arm, the 68 patients programmed with wide anatomical separation (≥30 mm) and shortest timing delays (MPP-AS) converted at 45.6%, versus 26.2% for other multipoint settings (P = 0.006) and a non-significant trend versus biventricular pacing (45.6% versus 33.8%, P = 0.10).

Watch out: The MPP-AS subgroup was identified within the trial's own data and is only 68 patients; the larger later report of the same programme did not reproduce its advantage.

Source: Leclercq C, Burri H, Curnis A, Delnoy PP et al.. Cardiac resynchronization therapy non-responder to responder conversion rate in the more response to cardiac resynchronization therapy with MultiPoint Pacing (MORE-CRT MPP) study: results from Phase I. European heart journal. 2019. PMID: 30859220. Registry ID: ClinicalTrials.gov NCT02006069. MeSH: Aged · Aged, 80 and over · Cardiac Pacing, Artificial · Cardiac Resynchronization Therapy · Cardiac Resynchronization Therapy Devices · Double-Blind Method · Echocardiography · Electrocardiography · Female · Follow-Up Studies · Heart Failure · Hemodynamics · Humans · Male · Middle Aged · Prospective Studies · Surveys and Questionnaires · Treatment Outcome

Card 2 — MORE-CRT MPP main report (1,111 non-responders randomised): conversion 29.4% versus 30.4% (P = 0.743), and the wide-separation subgroup advantage disappeared (Leclercq 2023)

Study: Randomised trial, main MORE-CRT MPP report (N = 1,111 randomised)

Population: The full MORE-CRT MPP programme: 5,850 patients planned for a CRT device; after 6 months of biventricular pacing, echocardiographic non-responders (less than 15% relative reduction in left-ventricular end-systolic volume) were randomised to multipoint pacing (541) or continued biventricular pacing (570) for 6 further months.

Summary: MORE-CRT MPP main report (1,111 non-responders randomised): conversion 29.4% versus 30.4% (P = 0.743), and the wide-separation subgroup advantage disappeared (Leclercq 2023)

Result: Conversion to echocardiographic response was 29.4% with multipoint pacing versus 30.4% with continued biventricular pacing (P = 0.743; the treatment made no difference). In patients with ≥30 mm spacing between the two left-ventricular pacing sites, no significant difference in conversion was observed. The authors note that about 30% of initial non-responders remodelled with simply more time on conventional therapy.

Watch out: This report and the 2019 Phase I report come from the same trial programme with overlapping enrolment; they are two reports of one programme, not two independent trials, and their Ns must not be added.

Source: Leclercq C, Burri H, Delnoy PP, Rinaldi CA et al.. Cardiac resynchronization therapy non-responder to responder conversion rate in the MORE-CRT MPP trial. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2023. PMID: 37776313. Registry ID: not stated in the record. MeSH: Humans · Cardiac Resynchronization Therapy · Treatment Outcome · Prospective Studies · Heart Failure · Cardiac Resynchronization Therapy Devices · Ventricular Function, Left

Card 3 — Secondary clinical-outcome analysis of the same programme (1,421 patients): heart-failure hospitalisation or death 6.64% on multipoint pacing versus 10.44% on biventricular pacing over about 5 months (Leclercq 2025)

Study: Secondary analysis of a randomised trial (N = 1,421 analysed)

Population: Of 3,724 CRT patients in the programme, 1,677 were non-responders and randomised; 1,421 (722 multipoint, 699 biventricular) had complete data for this secondary analysis of heart-failure hospitalisation or all-cause mortality, with mean follow-up of 5 ± 1 months after randomisation.

Summary: Secondary clinical-outcome analysis of the same programme (1,421 patients): heart-failure hospitalisation or death 6.64% on multipoint pacing versus 10.44% on biventricular pacing over about 5 months (Leclercq 2025)

Result: The composite of heart-failure hospitalisation or all-cause mortality occurred in 48 of 722 patients (6.64%) on multipoint pacing versus 73 of 699 (10.44%) on continued biventricular pacing (P = 0.0107); on multivariable analysis multipoint pacing was associated with lower occurrence of the endpoint (odds ratio 0.60, P = 0.0124), with lower event rates also reported in ischaemic patients and in those with long interventricular electrical delay.

Watch out: A secondary analysis with a 5-month horizon and complete-case restriction (1,421 of 1,677 randomised) sits underneath a null primary endpoint from the same programme; secondary-endpoint wins after a null primary are hypothesis-generating, and the event counts are small (48 versus 73).

Source: Leclercq C, Burri H, Calò L, Rinaldi CA et al.. Multipoint pacing is associated with reduction of heart failure hospitalizations or death in patients who do not respond to cardiac resynchronization therapy: results of the MORE-CRT MPP randomized trial. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2025. PMID: 40159375. Registry ID: not stated in the record. MeSH: Humans · Female · Male · Heart Failure · Cardiac Resynchronization Therapy · Aged · Hospitalization · Treatment Outcome · Middle Aged · Risk Factors · Time Factors · Ventricular Function, Left · Cause of Death · Stroke Volume

Card 4 — Meta-analysis (7 studies, 1,390 patients): multipoint pacing looked better in non-randomised studies (odds ratio 5.33) but not in randomised studies (odds ratio 1.86, CI crossing 1) (Mehta 2021)

Study: Systematic review and meta-analysis (7 studies, N = 1,390)

Population: Systematic review and meta-analysis of studies comparing multipoint pacing with conventional CRT with at least 6 months of follow-up, searched to January 2021; 7 studies, 1,390 patients, randomised and non-randomised designs pooled separately.

Summary: Meta-analysis (7 studies, 1,390 patients): multipoint pacing looked better in non-randomised studies (odds ratio 5.33) but not in randomised studies (odds ratio 1.86, CI crossing 1) (Mehta 2021)

Result: Echocardiographic improvement was greater with multipoint pacing in non-randomised studies (odds ratio 5.33, 95% CI 3.05 to 9.33) but not significant in randomised studies (odds ratio 1.86, 95% CI 0.91 to 3.79, p = .086). There was no significant difference in left-ventricular end-systolic volume reduction over 15% (odds ratio 1.96, 95% CI 0.69 to 5.55) or in improvement by at least one NYHA class (odds ratio 2.49, 95% CI 0.74 to 8.42). The wide-separation programming subgroup signal rested on only 120 patients.

Watch out: The synthesis pool overlaps the MORE-CRT programme carded above, so it corroborates rather than adds independent patients; the authors themselves conclude widespread multipoint programming in all CRT patients is not justified.

Source: Mehta VS, Elliott MK, Sidhu BS, Gould J et al.. Multipoint pacing for cardiac resynchronisation therapy in patients with heart failure: A systematic review and meta-analysis. Journal of cardiovascular electrophysiology. 2021. PMID: 34379350. Registry ID: not stated in the record. MeSH: Cardiac Resynchronization Therapy · Heart Failure · Humans · Stroke Volume · Time Factors · Treatment Outcome · Ventricular Function, Left