Point-of-care INR versus laboratory INR: correlation is excellent (r about 0.95-0.97) but a positive bias grows with the INR value, and about 1 in 18 results would change the warfarin dose

Shelf #10188 · Heart, vessels, blood pressure & clotting

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2018–2021.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients monitored on warfarin, how closely do point-of-care INR devices (capillary whole-blood coagulometers) agree with laboratory INR — the correlation, the size and direction of bias across the INR range, and how often the difference would change a dosing decision?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: 3 device-evaluation studies (200 paired measurements; 313 participants; 105 clinical samples). Diagnosis first: pack #3364 asks whether INR self-testing/self-management improves clinical outcomes; this pack asks the prior question — whether the number itself agrees with the laboratory — and live #7393 gives reference intervals for laboratory coagulation tests in healthy people, not device agreement in anticoagulated patients. No sensitivity/specificity pair is the natural measure here (INR is a continuous ratio), so agreement statistics are reported instead of an invented pair — stated plainly. The pattern is consistent: correlations of 0.95-0.97 and near-perfect categorical agreement (kappa 0.91-0.94), but a positive bias that grows with the INR value (mean difference 0.09 INR in the subtherapeutic range, 0.29 in the therapeutic range, 0.40 above it in the largest paired study), 60% of supratherapeutic samples differing by more than 0.5 from the laboratory analyser in one evaluation, and 5.5% of paired results leading to a different dosing decision. All three study groups conclude the devices are usable in range and should be confirmed by the laboratory at high values. Tested devices are named exactly as published (CoaguChek XS Pro, CoaguChek Pro II, CoaguChek XS Plus, microINR) under the tested-product clarification exception. MEASUREMENT ACCURACY — PARTIAL: accurate enough for routine in-range monitoring, with a systematic high-end bias that makes supratherapeutic point-of-care values unreliable unconfirmed.

MeSH: Administration, Oral · Adolescent · Adult · Anticoagulants · Female · Humans · International Normalized Ratio · Male · Middle Aged · Outpatients · Point-of-Care Systems · Warfarin · Young Adult · Aged · Aged, 80 and over · Atrial Fibrillation · Prothrombin Time · Reproducibility of Results · Thromboembolism · United States · Blood Coagulation · Blood Coagulation Tests · Sensitivity and Specificity

Cards (3)

Card 1 — CoaguChek XS Pro vs laboratory: r = 0.973 overall, but mean bias rose from 0.09 (subtherapeutic) to 0.40 INR (supratherapeutic), and 5.5% of pairs changed the dose decision (Moiz 2018)

Study: Device agreement study, 200 paired measurements

Population: 200 paired INR measurements in an outpatient oral anticoagulant clinic: point-of-care CoaguChek XS Pro (named as published) against a Sysmex CS2000i laboratory analyser.

Summary: CoaguChek XS Pro vs laboratory: r = 0.973 overall, but mean bias rose from 0.09 (subtherapeutic) to 0.40 INR (supratherapeutic), and 5.5% of pairs changed the dose decision (Moiz 2018)

Result: Correlation between devices was 0.973 and overall categorical agreement was excellent (kappa = 0.916). The overall mean difference was 0.21 ± 0.32 INR, rising by range: 0.09 in the subtherapeutic range (INR up to 1.9), 0.29 in the therapeutic range (2.0-3.0) and 0.40 in the supratherapeutic range (above 3.0). Eleven of 200 measurements (5.5%) would have produced a different dosing decision between the two instruments.

Watch out: Single clinic and single device pair; the dosing-decision analysis assumes the laboratory value is the correct one, which is the study's framing, not an independent truth standard.

Source: Moiz B et al.. Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic. Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis. 2018. PMID: 29374969. Registry ID: not stated in the record. MeSH: Administration, Oral · Adolescent · Adult · Anticoagulants · Female · Humans · International Normalized Ratio · Male · Middle Aged · Outpatients · Point-of-Care Systems · Warfarin · Young Adult

Card 2 — microINR vs laboratory ACL TOP 500: r = 0.973, predicted bias at most 5% at INR 2.0-4.5, combined agreement 96% (Refaai 2020)

Study: Device agreement study, N = 313

Population: 313 participants at three US centres: 68 non-anticoagulated normal donors and 245 warfarin-anticoagulated patients; duplicate capillary testing on the microINR and CoaguChek XS point-of-care systems (named as published) against venous laboratory testing.

Summary: microINR vs laboratory ACL TOP 500: r = 0.973, predicted bias at most 5% at INR 2.0-4.5, combined agreement 96% (Refaai 2020)

Result: Against the laboratory analyser, microINR gave a correlation of 0.973, slope 1.00 (95% CI 0.97 to 1.03) and intercept 0.08; against CoaguChek XS, correlation was 0.977. Predicted bias at INR 2.0, 3.5 and 4.5 was at most 5% against both references. Agreement was 97%, 95% and 100% across the ranges below 2.0, 2.0-4.5 and at least 4.5 respectively (combined 96%), and coefficients of variation were 5.03% (INR below 2.0) and 4.68% (therapeutic range).

Watch out: Industry-context multicentre trial of one newer device; its tidier bias profile than the 2018 study may reflect the device, the protocol (duplicate testing), or both.

Source: Refaai MA, Shah V, Fernando R. Performance of the microINR Point-of-Care System: A Multicenter Clinical Trial. Thrombosis and haemostasis. 2020. PMID: 32299116. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Anticoagulants · Atrial Fibrillation · Female · Humans · International Normalized Ratio · Male · Middle Aged · Point-of-Care Systems · Prothrombin Time · Reproducibility of Results · Thromboembolism · United States · Warfarin · Young Adult

Card 3 — CoaguChek Pro II vs laboratory STA-R Max: r = 0.95, kappa = 0.91 — but 60% of supratherapeutic samples differed by more than 0.5 INR (Nam 2021)

Study: Device agreement study, 105 samples

Population: 105 clinical samples; the CoaguChek Pro II (named as published) compared with CoaguChek XS Plus and with a laboratory STA-R Max analyser, plus capillary-versus-venous comparison on the same device.

Summary: CoaguChek Pro II vs laboratory STA-R Max: r = 0.95, kappa = 0.91 — but 60% of supratherapeutic samples differed by more than 0.5 INR (Nam 2021)

Result: Precision was acceptable (within-run coefficient of variation 2.71% and between-run 3.28% at normal level; 1.52% and 4.47% at abnormal level). Pro II values correlated very highly with XS Plus (r = 0.97, kappa = 0.94) and with the laboratory analyser (r = 0.95, kappa = 0.91), and capillary and venous values on Pro II correlated at r = 0.98. However, the mean difference against the laboratory analyser grew as INR rose, with 60% of samples in the supratherapeutic range differing by more than 0.5. The authors conclude values should be confirmed by laboratory analysers in the supratherapeutic range.

Watch out: Samples, not necessarily 105 individual patients; the supratherapeutic stratum is small in all three studies, so the high-end bias is consistent in direction but imprecisely sized.

Source: Nam M et al.. Performance evaluation of coaguchek pro II in comparison with coaguchek XS plus and sta-r Max using a sta-neoplastine CI plus. International journal of laboratory hematology. 2021. PMID: 33460254. Registry ID: not stated in the record. MeSH: Blood Coagulation · Blood Coagulation Tests · Female · Humans · International Normalized Ratio · Male · Middle Aged · Point-of-Care Systems · Sensitivity and Specificity