SourceMed · Shelf #10192 in the ask interface

Pseudoisochromatic plate tests for red-green colour vision deficiency: the Waggoner PIP24 matched Ishihara plate-for-plate (sensitivity and specificity 1.00 against it), and Ishihara was the most sensitive conventional test against the CAD reference

Shelf #10192 · Eyes, skin, nose & allergy

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2023–2026.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do printed pseudoisochromatic plate tests (Ishihara 38-plate edition and the Waggoner PIP24) detect congenital red-green colour vision deficiency — against each other and against the Colour Assessment and Diagnosis (CAD) test as reference?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: 2 small comparative studies (120 and 60 participants). Sensitivity and specificity are labelled in words at first occurrence in each card. Diagnosis first: no live shelf or pack addresses colour-vision testing accuracy; the nearest analogues are other sensory bedside tests (pack #7813, the whispered voice test for hearing), a different sense and a different reference-standard problem. The evidence is thinner than the headline figures suggest: the larger study used the Ishihara test itself as the reference for the Waggoner plates, so its perfect sensitivity and specificity (1.00) prove the two plate sets agree, not that either is right; the smaller study used an independent reference (the CAD test) and found Ishihara the most sensitive conventional test with complete specificity, but its abstract does not print Ishihara's exact sensitivity percentage, so none is claimed here. Both studies recruited known groups (deficient versus normal trichromats), a design that flatters accuracy compared with screening an unselected population. Tested products are named exactly as published (Ishihara 38 Plates, Waggoner PIP24, DIVE Colour Test) under the tested-product clarification exception. MEASUREMENT ACCURACY — PARTIAL: plate tests agree with each other perfectly and Ishihara performs best against an independent reference, but the total evidence is 180 selected participants with no unselected-population screening study.

MeSH: Humans · Color Vision Defects · Color Perception Tests · Color Vision · Sensitivity and Specificity · Color Perception

Cards (2)

Card 1 — Waggoner PIP24 vs Ishihara: perfect agreement, sensitivity and specificity 1.00 (95% CI 0.94 to 1.00) at the three-error criterion (Alduhayan 2023)

Study: Comparative accuracy study, N = 120

Population: 120 participants: 63 with congenital red-green colour vision defects and 57 with normal trichromacy, each tested with both the Ishihara 38-plate edition and the Waggoner PIP24 (both named as published).

Summary: Waggoner PIP24 vs Ishihara: perfect agreement, sensitivity and specificity 1.00 (95% CI 0.94 to 1.00) at the three-error criterion (Alduhayan 2023)

Result: Sensitivity and specificity are labelled in words here at first occurrence. Using a failure criterion of three or more errors on the screening plates (plates 1-15 for Waggoner, plates 1-17 for Ishihara), agreement between the two tests was perfect, and the Waggoner's sensitivity, specificity, predictive pass and predictive fail values against Ishihara were all 1.00 (95% CI 0.94 to 1.00). The authors conclude the Waggoner PIP24 is equivalent to the Ishihara for screening.

Watch out: The reference standard is the Ishihara test itself — this is an equivalence result between two plate sets, not validation of either against colour-vision status measured independently.

Source: Alduhayan R, Almustanyir A. Comparison of Two Printed Pseudoisochromatic Tests for Color Vision Assessment. Optometry and vision science : official publication of the American Academy of Optometry. 2023. PMID: 38019937. Registry ID: not stated in the record. MeSH: Humans · Color Vision Defects · Color Perception Tests · Color Vision · Sensitivity and Specificity · Color Perception

Card 2 — Against the CAD reference, Ishihara had the highest sensitivity of the conventional tests with complete specificity; the digital DIVE Touch reached sensitivity 78.1% / specificity 100% (Vilella 2026)

Study: Comparative accuracy study, N = 60

Population: 60 participants: 30 with CAD-confirmed congenital red-green deficiency and 30 CAD-confirmed normal trichromats; each completed the CAD test, the DIVE Colour Test (touch and eye-tracking), the Nagel anomaloscope, the Ishihara 38 Plates Test and the Roth 28-Hue Test (all named as published).

Summary: Against the CAD reference, Ishihara had the highest sensitivity of the conventional tests with complete specificity; the digital DIVE Touch reached sensitivity 78.1% / specificity 100% (Vilella 2026)

Result: Sensitivity and specificity are labelled in words here at first occurrence. In right-eye testing, DIVE Touch achieved sensitivity 78.1% and specificity 100.0% (kappa 0.77 against CAD); DIVE eye-tracking showed sensitivity 53.1% and specificity 96.4% (kappa 0.48). Among conventional tests against CAD, Ishihara showed the highest sensitivity with complete specificity (the record's abstract does not print the exact percentage, so none is claimed), the Nagel anomaloscope achieved substantial agreement, and the Roth 28-Hue Test had the lowest sensitivity for red-green classification. Binocular DIVE Touch figures were sensitivity 69.0% / specificity 100.0%.

Watch out: A 2026 study of 60 selected participants; Ishihara's headline position is qualitative in the record, and none of these figures comes from screening an unselected population.

Source: Vilella M et al.. Comparative diagnostic performance of digital and conventional tests for congenital red-green colour vision deficiency. Acta ophthalmologica. 2026. PMID: 42990001. Registry ID: not stated in the record. MeSH: none in the record