Point-of-care natriuretic peptide testing matches the laboratory closely: device AUCs of 0.95 and 0.98 for heart failure in the emergency department, and a primary-care NT-proBNP slope of 1.05 against the hospital assay

Shelf #10338 · Heart, vessels, blood pressure & clotting

MEASUREMENT ACCURACY — GOOD

Stamped: 2026-11-28.

Evidence published: 2011–2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do point-of-care natriuretic peptide devices (BNP and NT-proBNP) measure against central-laboratory reference assays and against the final diagnosis of heart failure, in emergency and primary care?

Balance of the evidence

MEASUREMENT ACCURACY — GOOD — Composition: 3 primary studies (documented N = 388: 250 + 94 + 44). Diagnosis first: two different reference standards are in play and are not pooled — agreement with the central-laboratory assay (the primary-care study) and discrimination of the final heart-failure diagnosis (the emergency-department and express-test studies). Sensitivity and specificity are labelled in words at first occurrence in each card that reports them. The pattern is consistent: quantitative point-of-care devices tracked the laboratory assay closely (regression slope 1.05, R² = 0.97) and discriminated heart failure with areas under the curve of 0.95–0.98, with results in 9–19 minutes and failure rates at or below the laboratory's own. The semi-quantitative express test was the weakest instrument (specificity 92% versus 100% for the quantitative test, and poor quantitation above 1800 pg/mL). MEASUREMENT ACCURACY — GOOD: quantitative point-of-care natriuretic peptide measurement is laboratory-grade for the rule-out/rule-in task; visually read semi-quantitative tests are a step below.

MeSH: Adult · Aged · Aged, 80 and over · Biomarkers · Emergency Service, Hospital · Female · Heart Failure · Humans · Male · Middle Aged · Natriuretic Peptide, Brain · Point-of-Care Systems · Predictive Value of Tests · Prospective Studies · Sensitivity and Specificity · Attitude of Health Personnel · Belgium · Clinical Laboratory Techniques · General Practitioners · Health Knowledge, Attitudes, Practice · Heart Function Tests · Peptide Fragments · Pilot Projects · Point-of-Care Testing · Primary Health Care · Reproducibility of Results · Surveys and Questionnaires · Echocardiography · Point-of-Care Testing · Severity of Illness Index · Chronic Disease

Cards (3)

Card 1 — Two emergency-department POC BNP devices reached AUCs of 0.95 (Triage) and 0.98 (i-STAT) against the final heart-failure diagnosis, with results in 19 and 9 minutes (Ro 2011)

Study: Primary diagnostic-accuracy study, N = 250

Population: 250 adults presenting to an emergency department with suspected heart failure; each venous sample was split three ways — central laboratory (Siemens Advia Centaur), Triage point-of-care device and i-STAT point-of-care device. A final hospital discharge diagnosis of heart failure was made in 108 patients (42%).

Summary: Two emergency-department POC BNP devices reached AUCs of 0.95 (Triage) and 0.98 (i-STAT) against the final heart-failure diagnosis, with results in 19 and 9 minutes (Ro 2011)

Result: Sensitivity and specificity are labelled in words here at first occurrence. Neither device differed significantly from the central laboratory in sensitivity; the i-STAT was less specific than the Triage device (p = 0.003). The area under the curve was 0.95 (95% CI 0.91 to 0.98) for Triage and 0.98 (95% CI 0.96 to 0.99) for i-STAT (p < 0.01). Median time to result was 9 minutes for i-STAT and 19 minutes for Triage (p < 0.001). Device failure rates were 0.4% (i-STAT) and 1.2% (Triage), against 3.2% for the central laboratory analyser itself.

Watch out: The reference diagnosis was the hospital discharge diagnosis, not the laboratory value alone; the record reports the sensitivity comparison without printing both devices' full sensitivity/specificity pairs, and none are reconstructed here.

Source: Ro R, Thode HC, Taylor M, Gulla J et al. Comparison of the diagnostic characteristics of two B-type natriuretic peptide point-of-care devices. The Journal of emergency medicine. 2011. PMID: 21620610. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Biomarkers · Emergency Service, Hospital · Female · Heart Failure · Humans · Male · Middle Aged · Natriuretic Peptide, Brain · Point-of-Care Systems · Predictive Value of Tests · Prospective Studies · Sensitivity and Specificity

Card 2 — In primary care, the Cobas h232 POC NT-proBNP tracked the hospital assay with slope 1.05 and R² = 0.97 across 94 patients with suspected heart failure (Hex 2018)

Study: Primary method-comparison study, N = 94

Population: 94 adult patients with suspected heart failure tested by 19 general practitioners in 5 GP groups: NT-proBNP measured at the point of care on the Cobas h232, with the same samples then transported to a central hospital laboratory and analysed on a Cobas e602.

Summary: In primary care, the Cobas h232 POC NT-proBNP tracked the hospital assay with slope 1.05 and R² = 0.97 across 94 patients with suspected heart failure (Hex 2018)

Result: Passing–Bablok regression gave a slope of 1.05 (95% CI 1.00 to 1.11) with R² = 0.97 against the hospital measurement; the percentage-difference plot showed a mean difference of 15.7%. GPs rated user-friendliness and usefulness at a median of 4–5 on a five-point scale, 18 of 19 said the device influenced their practice, and their confidence in using NT-proBNP rose from 4.4 to 7.6 out of 10 during the study.

Watch out: A pilot-scale method comparison (94 patients) measuring agreement with the laboratory assay, not diagnostic accuracy against an adjudicated heart-failure diagnosis; the 15.7% mean percentage difference is not negligible at decision thresholds.

Source: Hex C, Smeets M, Penders J, Van Hoof V et al. Accuracy, user-friendliness and usefulness of the Cobas h232 point-of-care test for NT-proBNP in primary care. Journal of clinical pathology. 2018. PMID: 29263170. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Attitude of Health Personnel · Belgium · Biomarkers · Clinical Laboratory Techniques · Female · General Practitioners · Health Knowledge, Attitudes, Practice · Heart Failure · Heart Function Tests · Humans · Male · Middle Aged · Natriuretic Peptide, Brain · Peptide Fragments · Pilot Projects · Point-of-Care Testing · Predictive Value of Tests · Primary Health Care · Reproducibility of Results · Surveys and Questionnaires

Card 3 — A semi-quantitative bedside NT-proBNP express test matched the quantitative test's sensitivity (95%) but not its specificity (92% vs 100%) in 44 patients (Iosifov 2024)

Study: Primary diagnostic-accuracy study, N = 44

Population: 44 patients assessed for chronic heart failure, each tested at the bedside with a semi-quantitative express NT-proBNP test (visual read) and quantitatively in the laboratory; all underwent echocardiography, and 11 were also scored on a heart-failure clinical status scale.

Summary: A semi-quantitative bedside NT-proBNP express test matched the quantitative test's sensitivity (95%) but not its specificity (92% vs 100%) in 44 patients (Iosifov 2024)

Result: Sensitivity and specificity are labelled in words here. Sensitivity was 95% for both the semi-quantitative and the quantitative test. Specificity was 100% for the quantitative test and 92% for the semi-quantitative test; negative predictive value was 96% for both, and overall diagnostic accuracy 98% versus 93%. The express test verified values above 1800 pg/mL poorly and showed increased sub-threshold sensitivity near cut-offs; the authors caution it cannot track treatment dynamics the way quantitative measurement can.

Watch out: Forty-four patients at a single centre and a visually read semi-quantitative device: this card bounds the low end of the technology class and should not be read as the performance of quantitative POC analysers.

Source: Iosifov AV, Shtegman OA. Utility of Point-of-Care Diagnosis of Chronic Heart Failure Using an Express Test for Semi-Quantitative Determination of NT-proBNP Levels. Kardiologiia. 2024. PMID: 39102570. Registry ID: not stated in the record. MeSH: Humans · Natriuretic Peptide, Brain · Heart Failure · Peptide Fragments · Female · Male · Aged · Middle Aged · Echocardiography · Sensitivity and Specificity · Biomarkers · Point-of-Care Systems · Severity of Illness Index · Chronic Disease