Point-of-care CRP analysers are not interchangeable: the best match the laboratory within a few percent, while one widely sold device read systematically low and failed the ±20% agreement criterion

Shelf #10339 · Critical care & sepsis

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2019–2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurate are point-of-care C-reactive protein (CRP) analysers analytically — their imprecision, bias and agreement against central-laboratory CRP methods and certified reference material — and does performance differ by device?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: 4 primary analytical evaluations (documented sample sets: 217 prospectively collected samples in the multicentre study; 100 retrospective plasma + 50 prospective capillary samples in the four-device comparison; 205 patient samples in the LumiraDx evaluation; the verification study's CRP sample count is not stated in its record and is not estimated here). This shelf is about analytical accuracy — agreement with laboratory methods and reference material — not about whether CRP testing changes prescribing (a separate, live question) and not about CRP as a diagnostic test for any single disease. The devices split: cobas b 101 and Afinion 2 agreed best with the laboratory; LumiraDx showed close agreement in its own evaluation (r = 0.992) yet read systematically low and violated the agreement criterion in the independent four-device comparison; QuikRead had the largest between-day imprecision (11.5%). MEASUREMENT ACCURACY — PARTIAL: the class can be laboratory-grade, but accuracy is a property of the individual device under local quality assurance, not of point-of-care CRP as such.

MeSH: Adult · Aged · Aged, 80 and over · C-Reactive Protein · Female · Humans · Male · Middle Aged · Point-of-Care Systems · Young Adult · Point-of-Care Testing · Glycated Hemoglobin · Cholesterol, LDL

Cards (4)

Card 1 — Multicentre evaluation of the cobas b 101 CRP test: repeatability CVs of 1.7–4.0%, agreement slope 0.97 against the laboratory reagent, and a 1.00% error rate across 9,378 tests (Scharnhorst 2019)

Study: Primary analytical evaluation, 217 method-comparison samples

Population: Analytical evaluation at 3 point-of-care sites and 1 reference laboratory; method comparison and matrix/lot comparisons used prospectively collected blood samples from 217 adults (apparently healthy or with clinically relevant conditions), with CRP spanning 3.3–360.3 mg/L in clinical samples.

Summary: Multicentre evaluation of the cobas b 101 CRP test: repeatability CVs of 1.7–4.0%, agreement slope 0.97 against the laboratory reagent, and a 1.00% error rate across 9,378 tests (Scharnhorst 2019)

Result: Within-run coefficients of variation ranged from 1.7% to 4.0% and within-laboratory (intermediate) precision from 1.9% to 4.5% for serum pools containing 4.7–350.7 mg/L; reproducibility CVs were 2.5–4.0%. Results were comparable across capillary whole blood, serum and anticoagulated whole blood/plasma (Pearson r ≥ 0.996) and across three reagent lots (r ≥ 0.993). Against the laboratory CRPNX reagent, weighted Deming regression gave y = 0.97x + 0.11 with r ≥ 0.996. Mean usability score was 4.18/5.

Watch out: A manufacturer-system evaluation conducted partly at point-of-care sites: performance under routine primary-care quality assurance, with less controlled operators, is the independent comparisons' question (next cards).

Source: Scharnhorst V, Noordzij PG, Lutz A, Graser U et al. A multicenter evaluation of a point of care CRP Test. Clinical biochemistry. 2019. PMID: 31233734. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · C-Reactive Protein · Female · Humans · Male · Middle Aged · Point-of-Care Systems · Young Adult

Card 2 — Head-to-head, cobas b 101 and Afinion 2 agreed best with the laboratory; LumiraDx read low and failed the criterion of >95% of results within ±20%, and QuikRead imprecision reached 11.5% (Van Hoovels 2024)

Study: Primary analytical evaluation, 150 comparison samples (100 plasma + 50 capillary)

Population: Independent evaluation of 4 commercial point-of-care CRP devices (QuikRead go, LumiraDx, cobas b 101, Afinion 2): imprecision on plasma pools and control material; trueness verified against ERM-DA474/IFCC reference material; method comparison against the central-laboratory cobas c 503 method using 100 retrospectively selected plasma samples and 50 prospectively collected capillary whole-blood samples.

Summary: Head-to-head, cobas b 101 and Afinion 2 agreed best with the laboratory; LumiraDx read low and failed the criterion of >95% of results within ±20%, and QuikRead imprecision reached 11.5% (Van Hoovels 2024)

Result: Between-day imprecision on plasma pools ranged from 4.5% (LumiraDx) to 11.5% (QuikRead). Traceability verification showed no significant difference between the laboratory method's results and the certified reference value. The cobas b 101 and Afinion achieved the best agreement with the central laboratory method. LumiraDx and QuikRead showed a negative mean difference, with LumiraDx violating the criterion that more than 95% of point-of-care results fall within ±20% of the comparison method. Afinion scored highest on user-friendliness. The authors conclude performance varies by manufacturer and needs central-laboratory-supervised quality assurance.

Watch out: One laboratory's evaluation with its own comparison method and acceptance criterion: it establishes device-to-device spread, and its LumiraDx finding sits in tension with the device-specific evaluation carded next — both are reported, neither is smoothed over.

Source: Van Hoovels L, Massa B, Stavelin A, De Meyer H et al. Analytical performance and user-friendliness of four commercially available point-of-care devices for C-reactive protein. Clinica chimica acta; international journal of clinical chemistry. 2024. PMID: 38768699. Registry ID: not stated in the record. MeSH: C-Reactive Protein · Humans · Point-of-Care Systems · Point-of-Care Testing

Card 3 — Verification study: both point-of-care analysers put every CRP sample within the national quality-assurance limits, with imprecision rated optimal against biological variation (Li 2023)

Study: Primary analytical verification study; CRP sample count not stated in the record

Population: Laboratory verification of the Roche cobas b 101 and Abbott Afinion 2 for CRP (alongside lipid studies and glycated haemoglobin, which are not this shelf's question), by method comparison with central-laboratory analysers, assessed against Royal College of Pathologists of Australasia quality-assurance analytical performance specifications; the record does not state the CRP comparison sample count.

Summary: Verification study: both point-of-care analysers put every CRP sample within the national quality-assurance limits, with imprecision rated optimal against biological variation (Li 2023)

Result: For CRP, both evaluated point-of-care analysers had all samples within the quality-assurance analytical performance specifications and had optimal imprecision according to biological variation. The authors judged all evaluated analysers fit for purpose for CRP. (The same study found device biases for some lipid and HbA1c measurements at extremes — outside this shelf's question and not carried here.)

Watch out: A verification against acceptance limits is a pass/fail exercise, not a bias estimate: 'all within limits' is compatible with the systematic differences the head-to-head study measured.

Source: Li B, Kumar A, Finlay C, van Drimmelen M et al. Verification of point-of-care analysers for C-reactive protein, lipid studies and glycated haemoglobin. Pathology. 2023. PMID: 37778963. Registry ID: not stated in the record. MeSH: Humans · Point-of-Care Systems · Glycated Hemoglobin · C-Reactive Protein · Cholesterol, LDL · Point-of-Care Testing

Card 4 — The LumiraDx CRP test agreed closely with the laboratory reference in its own evaluation: r = 0.992, slope 0.998 across 5.1–245.2 mg/L in 205 patient samples (Ellis 2024)

Study: Primary analytical evaluation, N = 205 samples

Population: 205 samples from patients presenting with symptoms of infection or inflammation, comparing the quantitative microfluidic LumiraDx point-of-care CRP test with a laboratory reference method (Siemens assay on Dimension Xpand), plus a 44-sample matrix-equivalence substudy (fingerstick, venous blood, plasma) and paired-replicate precision testing.

Summary: The LumiraDx CRP test agreed closely with the laboratory reference in its own evaluation: r = 0.992, slope 0.998 across 5.1–245.2 mg/L in 205 patient samples (Ellis 2024)

Result: Agreement with the laboratory reference was close (range 5.1 to 245.2 mg/L, r = 0.992, slope = 0.998, intercept = −0.476). Matrix agreement was r = 0.974–0.983. Mean replicate coefficients of variation were 6.4% (plasma), 6.6% (capillary direct) and 8.1% (venous blood); the overall error rate was 2.9%.

Watch out: This device-favourable result and the independent head-to-head's LumiraDx failure are both on this shelf deliberately: correlation coefficients near 0.99 can coexist with a systematic negative bias large enough to fail a ±20% agreement criterion, which is exactly the gap between the two studies' headline metrics.

Source: Ellis J, Harnett J, Cameron G, Moss P, Gray A. Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test. Practical laboratory medicine. 2024. PMID: 38188655. Registry ID: not stated in the record. MeSH: not stated in the record.