Balloon-expandable versus self-expanding transcatheter aortic valves: CHOICE found higher device success with the balloon-expandable valve (95.9% versus 77.5%) in first-generation devices, while SOLVE-TAVI found second-generation valves equivalent at 1 year

Shelf #10347 · Heart, vessels, blood pressure & clotting

SPLIT

Stamped: 2026-11-28.

Evidence published: 2014–2021.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with severe aortic stenosis undergoing transfemoral transcatheter aortic valve replacement, does a balloon-expandable valve give better device success and clinical outcomes than a self-expanding valve?

Balance of the evidence

SPLIT — Composition: 2 randomised trials (documented N = 688: 241 + 447). Diagnosis first: the two trials test different device generations, and the answer changed with the generation. In CHOICE (2014), using the Sapien XT balloon-expandable and CoreValve self-expanding systems, device success was 95.9% versus 77.5% (relative risk 1.24), driven by less more-than-mild paravalvular regurgitation (4.1% versus 18.3%) and fewer second valves, with fewer pacemakers (17.3% versus 37.6%) and no difference in 30-day cardiovascular mortality (4.1% versus 4.3%). In SOLVE-TAVI, with second-generation Evolut R and Sapien 3 valves, the 1-year combined endpoint of death, stroke, moderate or severe paravalvular leakage and permanent pacemaker implantation was essentially identical (38.3% balloon-expandable versus 40.4% self-expanding; hazard ratio 0.94). SPLIT: the balloon-expandable advantage was a property of the first-generation pairing; with current-generation devices the randomised evidence shows no overall winner, and choice turns on anatomy and operator factors outside these trials.

MeSH: Aged · Aged, 80 and over · Anesthesia · Aortic Valve · Aortic Valve Stenosis · Cardiac Catheterization · Female · Follow-Up Studies · Germany · Heart Valve Prosthesis Implantation · Hemorrhage · Humans · Incidence · Male · Middle Aged · Postoperative Complications · Prosthesis Design · Risk · Survival Rate · Transcatheter Aortic Valve Replacement · Treatment Outcome

Cards (2)

Card 1 — CHOICE: balloon-expandable valve device success 95.9% versus 77.5% for self-expanding (relative risk 1.24), with less paravalvular regurgitation and fewer pacemakers, but equal 30-day cardiovascular mortality (Abdel-Wahab 2014)

Study: Primary randomised trial, N = 241

Population: 241 patients with high-risk severe aortic stenosis and anatomy suitable for transfemoral TAVR at 5 German centres, randomly assigned to a balloon-expandable valve (Edwards Sapien XT, n = 121) or a self-expanding valve (Medtronic CoreValve, n = 120), enrolled 2012–2013.

Summary: CHOICE: balloon-expandable valve device success 95.9% versus 77.5% for self-expanding (relative risk 1.24), with less paravalvular regurgitation and fewer pacemakers, but equal 30-day cardiovascular mortality (Abdel-Wahab 2014)

Result: Device success occurred in 116 of 121 balloon-expandable patients (95.9%) versus 93 of 120 self-expanding patients (77.5%) (relative risk 1.24; 95% CI 1.12–1.37; P < 0.001), attributed to less residual more-than-mild aortic regurgitation (4.1% versus 18.3%; relative risk 0.23; P < 0.001) and less frequent implantation of more than 1 valve (0.8% versus 5.8%; P = 0.03). Cardiovascular mortality at 30 days was 4.1% versus 4.3% (P = 0.99), bleeding and vascular complications did not differ significantly, and new permanent pacemaker placement was less frequent with the balloon-expandable valve (17.3% versus 37.6%; P = 0.001).

Watch out: Both devices are superseded generations: the Sapien XT and the original CoreValve are no longer implanted, so the size of this advantage does not transfer to current valves — the second card tests the current-generation pairing.

Source: Abdel-Wahab M, Mehilli J, Frerker C, Neumann FJ, Kurz T, Tölg R et al. Comparison of balloon-expandable vs self-expandable valves in patients undergoing transcatheter aortic valve replacement: the CHOICE randomized clinical trial. JAMA. 2014. PMID: 24682026. Registry ID: NCT01645202. MeSH: Aged · Aged, 80 and over · Aortic Valve · Aortic Valve Stenosis · Cardiac Catheterization · Female · Heart Valve Prosthesis Implantation · Hemorrhage · Humans · Male · Middle Aged · Prosthesis Design · Risk · Survival Rate · Treatment Outcome

Card 2 — SOLVE-TAVI at 1 year: second-generation self-expanding (Evolut R) and balloon-expandable (Sapien 3) valves had near-identical combined outcomes — 40.4% versus 38.3%, hazard ratio 0.94 (Feistritzer 2021)

Study: Primary randomised trial (1-year valve-comparison report), N = 447

Population: 447 intermediate- to high-risk patients with severe symptomatic aortic stenosis at 7 sites, randomly assigned in a 2 × 2 factorial design to transfemoral TAVR with the self-expanding Evolut R or the balloon-expandable Sapien 3 valve (and, separately, to conscious sedation or general anaesthesia).

Summary: SOLVE-TAVI at 1 year: second-generation self-expanding (Evolut R) and balloon-expandable (Sapien 3) valves had near-identical combined outcomes — 40.4% versus 38.3%, hazard ratio 0.94 (Feistritzer 2021)

Result: In the valve comparison, the combined endpoint of all-cause mortality, stroke, moderate or severe paravalvular leakage and permanent pacemaker implantation at 1 year occurred in 84 patients (38.3%) with the balloon-expandable valve versus 87 (40.4%) with the self-expanding valve (hazard ratio 0.94; 95% CI 0.70 to 1.26; P = 0.66). The trial's 30-day report likewise found the two valve strategies similar; the companion anaesthesia comparison in the same factorial trial is reported separately and is not carded here.

Watch out: The combined endpoint bundles pacemaker implantation and paravalvular leakage with death and stroke, so equivalence on the bundle can hide opposite movements in its parts; patients had intermediate-to-high surgical risk and transfemoral anatomy, not the low-risk or bicuspid populations.

Source: Feistritzer HJ, Kurz T, Stachel G, Hartung P, Lurz P et al. Impact of Anesthesia Strategy and Valve Type on Clinical Outcomes After Transcatheter Aortic Valve Replacement. Journal of the American College of Cardiology. 2021. PMID: 33926657. Registry ID: NCT02737150. MeSH: Aged, 80 and over · Anesthesia · Aortic Valve Stenosis · Female · Follow-Up Studies · Germany · Humans · Incidence · Male · Postoperative Complications · Transcatheter Aortic Valve Replacement · Treatment Outcome