SourceMed · Shelf #10439 in the ask interface
A glaucomatous visual field cannot be read off one retest: at points with moderate loss, repeat thresholds swing from normal sensitivity to absolute defect, and variability grows as sensitivity falls — larger-stimulus perimetry (size V, Matrix, motion) flattens that growth
Shelf #10439 · Eyes, skin, nose & allergy
MEASUREMENT ACCURACY — POOR
Stamped: 2026-11-28.
Evidence published: 1989–2009.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How large is the test–retest variability of automated perimetry in glaucoma — how does pointwise variability depend on defect depth and test strategy (Full Threshold, SITA Standard, SITA Fast, larger stimuli), and what does that imply for calling progression from two fields?
Balance of the evidence
MEASUREMENT ACCURACY — POOR — Composition: 3 primary repeat-testing studies (51 + 49 + 120 patients with glaucoma; total documented N = 220). Diagnosis of the measurement problem first: perimetry is used to detect progression, so the standard it must meet is that a real change exceeds the noise between two tests at the same point. In damaged areas it does not. In the earliest study, points that initially showed moderate loss ranged on retest from normal sensitivity to absolute defect, while points starting normal or near-normal varied much less and variability was lowest centrally. Strategy comparisons quantify the same gradient: with sensitivities above 25 dB both SITA strategies had lower test–retest variability than Full Threshold; below 25 dB SITA Standard's retest intervals were slightly smaller and SITA Fast's larger, and SITA and Full Threshold threshold estimates differed by about 3 dB at midrange sensitivities (~15 dB) — small beside the retest variability itself. Stimulus size is the one lever that changes the picture: with standard size III, sensitivity explained 22% of retest variability, against 12% for size V and 2% each for Matrix and motion perimetry, whose variability barely grew as sensitivity fell. No sensitivity or specificity estimate is at issue here — this is repeatability of a threshold measurement. MEASUREMENT ACCURACY — POOR: for standard perimetry in damaged field regions, single-retest swings span the whole clinical range, so progression calls require series of fields and variability-aware analysis, not two-test comparisons.
MeSH: Adult · Aged · Aged, 80 and over · Differential Threshold · Glaucoma · Humans · Light · Middle Aged · Probability · Time Factors · Visual Field Tests · Visual Fields · Algorithms · Female · Male · Reproducibility of Results · Sensitivity and Specificity · Sensory Thresholds · Vision Disorders · Glaucoma, Open-Angle · Optic Disk · Optic Nerve Diseases
Cards (3)
Card 1 — In 51 patients tested four times within four weeks, points with moderate sensitivity loss swung on retest from normal sensitivity to absolute defect — while near-normal points and central points varied far less (Heijl 1989)
Study: Primary test–retest study, N = 51
Population: Fifty-one patients with glaucoma ranging from incipient to advanced, each tested four times within a four-week period with computerised visual fields; variability analysed by initial defect depth and by distance from fixation.
Summary: In 51 patients tested four times within four weeks, points with moderate sensitivity loss swung on retest from normal sensitivity to absolute defect — while near-normal points and central points varied far less (Heijl 1989)
Result: In areas initially found to have moderate loss of sensitivity, variation in follow-up measurements ranged from normal sensitivity to absolute defect, with little dependence on distance from fixation. Large changes were considerably more unusual in locations initially showing normal or near-normal sensitivities, and variability was lowest in the most central portion of the field. The authors conclude that separating true progression from random variation requires accounting for defect depth and location, and basing comparisons on more than two tests.
Watch out: 1989-era full-threshold testing: absolute variability with modern fast strategies differs in size (the other cards measure it), but the dependence of variability on defect depth — the clinically decisive feature — is confirmed by both.
Source: Heijl A, Lindgren A, Lindgren G. Test-retest variability in glaucomatous visual fields. American journal of ophthalmology. 1989. PMID: 2757094. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Differential Threshold · Glaucoma · Humans · Light · Middle Aged · Probability · Time Factors · Visual Field Tests · Visual Fields
Card 2 — In 49 patients tested four times with each strategy, SITA and Full Threshold thresholds differed by about 3 dB at midrange sensitivity (~15 dB); above 25 dB SITA varied less on retest, below 25 dB SITA Fast varied more (Artes 2002)
Study: Primary test–retest strategy comparison, N = 49
Population: One eye of 49 patients with glaucoma (mean age 61.6 years; Mean Deviation mean −7.13 dB, range +1.8 to −23.9 dB), each examined four times with Full Threshold, SITA Standard and SITA Fast on the Humphrey Field Analyzer; pointwise 90% retest limits (5th and 95th percentiles) derived for each strategy.
Summary: In 49 patients tested four times with each strategy, SITA and Full Threshold thresholds differed by about 3 dB at midrange sensitivity (~15 dB); above 25 dB SITA varied less on retest, below 25 dB SITA Fast varied more (Artes 2002)
Result: Threshold differences between SITA and Full Threshold were largest (approximately 3 dB) for midrange sensitivities (approximately 15 dB), and the threshold distributions of SITA differed considerably from Full Threshold. With sensitivities above 25 dB, both SITA strategies showed lower test–retest variability than Full Threshold; below 25 dB, SITA Standard's retest intervals were slightly smaller than Full Threshold's, whereas SITA Fast's were larger. The authors conclude SITA Standard may be superior for monitoring, and that the strategy differences are small compared with the test–retest variability itself.
Watch out: The record's MeSH includes the terms Sensitivity and Specificity, but the study reports threshold distributions and retest limits, not diagnostic sensitivity or specificity values — none are claimed on this card.
Source: Artes PH, Iwase A, Ohno Y, Kitazawa Y, Chauhan BC. Properties of perimetric threshold estimates from Full Threshold, SITA Standard, and SITA Fast strategies. Investigative ophthalmology & visual science. 2002. PMID: 12147599. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Algorithms · Female · Glaucoma · Humans · Male · Middle Aged · Reproducibility of Results · Sensitivity and Specificity · Sensory Thresholds · Vision Disorders · Visual Field Tests · Visual Fields
Card 3 — In 120 patients, standard size-III perimetry's retest variability rose steeply as sensitivity fell (sensitivity explained 22% of variability), while size V explained 12% and Matrix and motion perimetry only 2% each (Wall 2009)
Study: Primary test–retest comparison of four perimetry methods, N = 120
Population: One eye each of 120 patients with glaucoma, examined on the same day with four tests — standard automated perimetry size III (SITA Standard), size V (full threshold), Matrix (FDT II) and motion perimetry — and retested 1 to 8 weeks later, with decibel scales adjusted to be numerically similar.
Summary: In 120 patients, standard size-III perimetry's retest variability rose steeply as sensitivity fell (sensitivity explained 22% of variability), while size V explained 12% and Matrix and motion perimetry only 2% each (Wall 2009)
Result: With size III, retest variability increased substantially with reducing sensitivity; corresponding increases with size V, Matrix and motion perimetry were considerably smaller or absent. Sensitivity explained 22% of retest variability (r²) for size III, 12% for size V, and 2% each for Matrix and motion perimetry. The authors attribute the flatter variability of the larger-stimulus tests to increased sampling, and propose them as candidates for earlier detection of progression.
Watch out: As with the Artes card, the record's MeSH lists Sensitivity and Specificity as indexing terms; the study itself reports retest distributions and variance explained, not diagnostic accuracy values.
Source: Wall M, Woodward KR, Doyle CK, Artes PH. Repeatability of automated perimetry: a comparison between standard automated perimetry with stimulus size III and V, matrix, and motion perimetry. Investigative ophthalmology & visual science. 2009. PMID: 18952921. Registry ID: not stated in the record. MeSH: Adult · Aged · Aged, 80 and over · Female · Glaucoma, Open-Angle · Humans · Male · Middle Aged · Optic Disk · Optic Nerve Diseases · Reproducibility of Results · Sensitivity and Specificity · Sensory Thresholds · Vision Disorders · Visual Field Tests · Visual Fields