SourceMed · Shelf #10440 in the ask interface

OCT thickness numbers are repeatable to a few micrometres — ganglion-cell test–retest standard deviation 1.45 µm on average thickness but 3.42 µm on minimum thickness, intraclass correlations 0.95–0.99 — except peripapillary RNFL, whose retest limits widen steeply once thickness falls below about 82 µm

Shelf #10440 · Eyes, skin, nose & allergy

MEASUREMENT ACCURACY — GOOD

Stamped: 2026-11-28.

Evidence published: 2013–2017.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How reproducible are spectral-domain OCT structural measurements used to monitor glaucoma — peripapillary retinal nerve fibre layer (RNFL) thickness, macular ganglion cell (GCIPL/GCC) thickness and cup/disc ratios — between visits, and how does reproducibility depend on the amount of damage and on the segmentation algorithm?

Balance of the evidence

MEASUREMENT ACCURACY — GOOD — Composition: 3 primary repeat-imaging studies (24 subjects / 41 eyes; 109 eyes; 104 eyes from 56 patients; total documented N = 189 patients, 254 eyes). The intraclass correlation coefficients reported here are reliability coefficients: as a plain-words guide, 0.70 suffices for comparing groups, while 0.90 is the ideal — a measure is acceptable for individual-patient decisions when it reaches the 90% threshold. By that standard these measurements clear the individual-patient bar: long-term GCIPL intraclass correlations ran 0.966–0.992 with coefficients of variation of 2.0–5.5%, and the two-algorithm ganglion-cell comparison gave intraclass correlations of 0.98 (Iowa algorithm) and 0.95 (Cirrus), with intervisit standard deviations of 1.55 µm and 2.45 µm — algorithm choice alone nearly doubles the noise. Absolute noise floors matter more than correlations for progression calls: average GCIPL thickness had the lowest test–retest standard deviation (1.45 µm) and minimum GCIPL the highest (3.42 µm). The exception is the workhorse peripapillary RNFL average: its repeatability held at about ±5.61 µm while thickness stayed above ~82 µm, then widened linearly as the layer thinned — fitted in the Topcon study as limits of agreement of ±(34.67 − 0.294 × thickness) µm — so a fixed change threshold overcalls progression exactly where glaucoma is advanced. Cup/disc ratio coefficients of repeatability were 0.05 (horizontal) and 0.07 (vertical), and the ganglion cell complex coefficient was 2.9 µm. MEASUREMENT ACCURACY — GOOD: macular ganglion-cell metrics are reproducible enough for individual monitoring; peripapillary RNFL change must be judged against thickness-dependent limits, not a single cut-off.

MeSH: Aged · Aged, 80 and over · Female · Glaucoma, Open-Angle · Humans · Imaging, Three-Dimensional · Intraocular Pressure · Male · Middle Aged · Nerve Fibers · Optic Disk · Optic Nerve Diseases · Reproducibility of Results · Retinal Ganglion Cells · Tomography, Optical Coherence · Adult · Disease Progression · Longitudinal Studies · Observer Variation · Retrospective Studies · Tonometry, Ocular · Visual Fields · Algorithms · Follow-Up Studies · Macula Lutea · Prospective Studies

Cards (3)

Card 1 — In 24 glaucoma subjects (41 eyes) re-imaged one week apart, ganglion cell complex repeatability was 2.9 µm and cup/disc coefficients 0.05–0.07 — but peripapillary RNFL retest limits widened as the layer thinned below about 82 µm (Pearce 2017)

Study: Primary intervisit reproducibility study, N = 24 subjects (41 eyes)

Population: Twenty-four subjects with varying glaucoma severity contributing 41 usable eyes, scanned with a Topcon 3D OCT-2000 (disc/RNFL and macular ganglion cell complex protocols) twice, one week apart at the same time of day; variability expressed as Bland–Altman limits of agreement and coefficients of repeatability.

Summary: In 24 glaucoma subjects (41 eyes) re-imaged one week apart, ganglion cell complex repeatability was 2.9 µm and cup/disc coefficients 0.05–0.07 — but peripapillary RNFL retest limits widened as the layer thinned below about 82 µm (Pearce 2017)

Result: Overall horizontal and vertical cup/disc ratio coefficients of repeatability were 0.05 and 0.07; the ganglion cell complex coefficient of repeatability was 2.9 µm. Average peripapillary RNFL test–retest variability expanded with increasing damage, with limits of agreement fitted by ±(34.67 − 0.294 × d) µm where d is RNFL thickness; a model with constant limits of ±5.61 µm above 82 µm and linearly expanding variability below it achieved marginal significance (P < .06).

Watch out: One device (Topcon 3D OCT-2000) and 24 subjects: the thickness-dependent widening is the transferable finding; the exact fitted coefficients belong to this instrument and protocol.

Source: Pearce JG, Maddess T. Inter-visit Test-Retest Variability of OCT in Glaucoma. Optometry and vision science. 2017. PMID: 27870778. Registry ID: not stated in the record. MeSH: Aged · Aged, 80 and over · Female · Glaucoma, Open-Angle · Humans · Imaging, Three-Dimensional · Intraocular Pressure · Male · Middle Aged · Nerve Fibers · Optic Disk · Optic Nerve Diseases · Reproducibility of Results · Retinal Ganglion Cells · Tomography, Optical Coherence

Card 2 — In 109 eyes with stable glaucoma scanned three times at 6-month intervals, macular GCIPL intraclass correlations were 0.966–0.992 and test–retest standard deviation was lowest for average thickness (1.45 µm) and highest for minimum thickness (3.42 µm) (Kim 2015)

Study: Primary long-term reproducibility study, N = 109 eyes

Population: One hundred nine eyes of 109 clinically stable open-angle glaucoma patients with a localised RNFL defect and corresponding macular ganglion cell–inner plexiform layer (GCIPL) defect, with three serial Cirrus HD-OCT peripapillary and macular scans at 6-month intervals; stability defined by no change on serial photography and visual-field progression analysis.

Summary: In 109 eyes with stable glaucoma scanned three times at 6-month intervals, macular GCIPL intraclass correlations were 0.966–0.992 and test–retest standard deviation was lowest for average thickness (1.45 µm) and highest for minimum thickness (3.42 µm) (Kim 2015)

Result: GCIPL thickness intraclass correlation coefficients ranged from 0.966 to 0.992 and coefficients of variation from 2.0% to 5.5%. Test–retest standard deviation was lowest for average GCIPL thickness (1.45 µm), highest for minimum GCIPL thickness (3.42 µm), and 1.54–2.16 µm for sectoral measurements; defect angular width gave an intraclass correlation of 0.993 (coefficient of variation 3.9%, test–retest standard deviation 5.32°) and defect area 0.930 (6.7%, 0.27 mm²). Measurement variance showed no significant association with glaucoma severity.

Watch out: Patients were selected for proven clinical stability, so these are best-case noise floors: eyes actually progressing, or with poorer fixation and signal strength, will vary more.

Source: Kim KE, Yoo BW, Jeoung JW, Park KH. Long-Term Reproducibility of Macular Ganglion Cell Analysis in Clinically Stable Glaucoma Patients. Investigative ophthalmology & visual science. 2015. PMID: 25829417. Registry ID: not stated in the record. MeSH: Adult · Aged · Disease Progression · Female · Glaucoma, Open-Angle · Humans · Intraocular Pressure · Longitudinal Studies · Male · Middle Aged · Nerve Fibers · Observer Variation · Reproducibility of Results · Retinal Ganglion Cells · Retrospective Studies · Tomography, Optical Coherence · Tonometry, Ocular · Visual Fields

Card 3 — In 104 eyes imaged on two visits, ganglion cell layer thickness intraclass correlation was 0.98 with the Iowa algorithm versus 0.95 with the Cirrus algorithm, and intervisit standard deviation 1.55 µm versus 2.45 µm — the software, not just the scanner, sets the noise (Garvin 2013)

Study: Primary reproducibility study (two segmentation algorithms), N = 56 patients (104 eyes)

Population: Both eyes of patients with open-angle glaucoma or suspected glaucoma (104 eyes from 56 patients after exclusions for low signal strength and acquisition or analysis errors), with macular SD-OCT volumes on two visits within 4 months, segmented by a published graph-theoretical (Iowa) algorithm and by the Cirrus ganglion cell analysis module.

Summary: In 104 eyes imaged on two visits, ganglion cell layer thickness intraclass correlation was 0.98 with the Iowa algorithm versus 0.95 with the Cirrus algorithm, and intervisit standard deviation 1.55 µm versus 2.45 µm — the software, not just the scanner, sets the noise (Garvin 2013)

Result: For overall elliptical-annulus ganglion cell layer plus inner plexiform layer thickness, the intraclass correlation coefficient was 0.98 (95% CI 0.97–0.99) with the Iowa algorithm and 0.95 (95% CI 0.93–0.97) with the Cirrus algorithm; intervisit standard deviations were 1.55 µm and 2.45 µm and coefficients of variation 2.2% and 3.5% respectively (P < 0.0001).

Watch out: Eyes with low signal strength or segmentation errors were excluded before analysis: in routine clinics, where such scans are often kept, reproducibility will be worse than these figures.

Source: Garvin MK, Lee K, Burns TL, Abràmoff MD, Sonka M, Kwon YH. Reproducibility of SD-OCT-based ganglion cell-layer thickness in glaucoma using two different segmentation algorithms. Investigative ophthalmology & visual science. 2013. PMID: 24045993. Registry ID: not stated in the record. MeSH: Algorithms · Female · Follow-Up Studies · Glaucoma, Open-Angle · Humans · Macula Lutea · Male · Middle Aged · Nerve Fibers · Prospective Studies · Reproducibility of Results · Retinal Ganglion Cells · Tomography, Optical Coherence