Smartwatch irregular-pulse notifications are usually right when they fire again during monitoring — positive predictive value 0.84 (Apple) and 98.2% (Fitbit) for concurrent atrial fibrillation — but only about a third of notified users who return an ECG patch have atrial fibrillation on it, and notifications reach only about 0.5–1% of users
Shelf #10491 · Heart, vessels, blood pressure & clotting
PREDICTIVE PERFORMANCE — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2019–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults without known atrial fibrillation using consumer smartwatches, how well do photoplethysmography irregular-pulse notifications predict atrial fibrillation — what proportion of notified users have atrial fibrillation on a subsequent ECG patch, what is the positive predictive value of a notification that recurs during patch monitoring, and how many users are notified at all?
Balance of the evidence
PREDICTIVE PERFORMANCE — PARTIAL — Composition: two very large prospective remote studies (419,297 and 455,699 participants) and one small simultaneous-recording validation in 24 patients with insertable cardiac monitors; total documented N = 875,020. Diagnosis of the prediction problem first: the watches sample pulse intermittently in a mostly young, low-prevalence population, so the performance questions are separate — how many are flagged, how many flagged people truly have atrial fibrillation, and whether a flag that repeats during ECG monitoring is trustworthy. The answers split. Notifications are rare (0.52% of Apple participants over a median 117 days; 1% of Fitbit participants) and the yield of the follow-up patch is modest: atrial fibrillation was found in 34% of Apple participants who returned analysable patches and 32.2% of Fitbit participants who did. But a notification that recurs while the ECG patch is actually recording is highly predictive: positive predictive value 0.84 for a subsequent Apple tachogram notification with simultaneous atrial fibrillation, and 98.2% for a Fitbit irregular-rhythm detection during patch monitoring (97.0% in those 65 and over). The small continuous-comparison study sounds a caution from the other direction: episode sensitivity (in words: the share of true episodes caught) was 97.5% for episodes of at least 1 hour, yet the positive predictive value for episodes was only 39.9% against the implanted monitor's record. The tested products and programmes are named as the records name them: the Apple Heart Study (Apple Watch with the irregular-pulse notification algorithm; its record states no registry ID) and the Fitbit Heart Study (NCT04380415, stated in its record). PREDICTIVE PERFORMANCE — PARTIAL: a repeated wearable irregular-pulse notification during ECG monitoring is strong evidence of concurrent atrial fibrillation, but a single notification is only a prompt for testing — about two-thirds of notified users who complete patch monitoring do not have atrial fibrillation found on it.
MeSH: Adult · Aged · Algorithms · Atrial Fibrillation · Confidentiality · Electrocardiography · Female · Humans · Male · Middle Aged · Mobile Applications · Predictive Value of Tests · Prospective Studies · Telemedicine · Wearable Electronic Devices · Photoplethysmography · Electrocardiography, Ambulatory · Action Potentials · Deep Learning · Heart Atria · Heart Rate · Remote Sensing Technology · Reproducibility of Results · Signal Processing, Computer-Assisted
Cards (3)
Card 1 — Apple Heart Study, 419,297 participants: 0.52% were notified over a median 117 days; atrial fibrillation was found in 34% of those returning analysable ECG patches — but a later notification during patch wear had a positive predictive value of 0.84 for simultaneous atrial fibrillation (Perez 2019)
Study: Prospective remote predictive-performance study, N = 419,297
Population: Adults without self-reported atrial fibrillation using an Apple Watch and iPhone app; notified participants had a telemedicine visit and were mailed an ECG patch to wear for up to 7 days, applied on average 13 days after notification; 450 participants returned patches with analysable data.
Summary: Apple Heart Study, 419,297 participants: 0.52% were notified over a median 117 days; atrial fibrillation was found in 34% of those returning analysable ECG patches — but a later notification during patch wear had a positive predictive value of 0.84 for simultaneous atrial fibrillation (Perez 2019)
Result: Over a median of 117 days of monitoring, 2,161 participants (0.52%) received irregular-pulse notifications. Among the 450 who returned analysable ECG patches, atrial fibrillation was present in 34% (97.5% CI 29–39) overall and 35% in those 65 and older. Among notified participants, the positive predictive value was 0.84 (95% CI 0.76–0.92) for observing atrial fibrillation on the ECG simultaneously with a subsequent irregular-pulse notification.
Watch out: Only a minority of notified participants returned analysable patches, and the patch was worn about 13 days after the notification — so the 34% is the yield of delayed patch testing in self-selected returners, not the probability that the original notification was atrial fibrillation.
Source: Perez MV, Mahaffey KW, Hedlin H, Rumsfeld JS, Garcia A, Ferris T, Balasubramanian V, Russo AM, Rajmane A, Cheung L, Hung G, Lee J, Kowey P, Talati N, Nag D, Gummidipundi SE, Beatty A, Hills MT, Desai S, Granger CB, Desai M, Turakhia MP. Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. The New England journal of medicine. 2019. PMID: 31722151. Registry ID: NCT03335800. MeSH: Adult · Aged · Algorithms · Atrial Fibrillation · Confidentiality · Electrocardiography · Female · Humans · Male · Middle Aged · Mobile Applications · Predictive Value of Tests · Prospective Studies · Telemedicine · Wearable Electronic Devices
Card 2 — Fitbit Heart Study, 455,699 participants: irregular-rhythm detections in 1%; atrial fibrillation in 32.2% of those with a detection who returned analysable patches — and a detection recurring during patch wear had a positive predictive value of 98.2% for concurrent atrial fibrillation (Lubitz 2022)
Study: Prospective remote predictive-performance study, N = 455,699
Population: Adults aged 22 and over in the United States without atrial fibrillation, using compatible Fitbit devices and smartphones; the algorithm examined overlapping 5-minute pulse tachograms, an irregular heart rhythm detection required 11 consecutive irregular tachograms, and detected participants were invited to a telehealth visit and mailed a 1-week ECG patch. Registry ID: NCT04380415, stated in the record.
Summary: Fitbit Heart Study, 455,699 participants: irregular-rhythm detections in 1%; atrial fibrillation in 32.2% of those with a detection who returned analysable patches — and a detection recurring during patch wear had a positive predictive value of 98.2% for concurrent atrial fibrillation (Lubitz 2022)
Result: A total of 455,699 participants enrolled; irregular heart rhythm detections occurred in 4,728 (1%), and in 2,070 (4%) of those aged 65 and over, over a median 122 days at risk. Among 1,057 participants with a detection and subsequent analysable patch, atrial fibrillation was present in 340 (32.2%). Of 225 participants with another detection during patch monitoring, 221 had concurrent atrial fibrillation — a positive predictive value of 98.2% (95% CI 95.5–99.5), and 97.0% in those 65 and over.
Watch out: As with the Apple study, patch return was incomplete and delayed relative to the first detection, so the 32.2% patch yield and the 98.2% concurrent-detection predictive value answer different questions — the second detection, made while an ECG is actually recording, is the trustworthy one.
Source: Lubitz SA, Faranesh AZ, Selvaggi C, Atlas SJ, McManus DD, Singer DE, Pagoto S, McConnell MV, Pantelopoulos A, Foulkes AS. Detection of Atrial Fibrillation in a Large Population Using Wearable Devices: The Fitbit Heart Study. Circulation. 2022. PMID: 36148649. Registry ID: NCT04380415. MeSH: Adult · Female · Humans · Middle Aged · Male · Atrial Fibrillation · Prospective Studies · Photoplethysmography · Electrocardiography, Ambulatory · Electrocardiography · Wearable Electronic Devices
Card 3 — Against an insertable cardiac monitor in 24 patients with paroxysmal atrial fibrillation, the watch algorithm caught 80 of 82 episodes lasting at least 1 hour (episode sensitivity, in words, 97.5%) — but its positive predictive value for episodes was only 39.9% (Wasserlauf 2019)
Study: Simultaneous-recording validation study, N = 24
Population: Twenty-four patients with a history of paroxysmal atrial fibrillation and insertable cardiac monitors (Reveal LINQ) simultaneously wearing an AF-sensing watch (Apple Watch with KardiaBand, SmartRhythm neural network), with 31,348.9 hours of simultaneous recording analysed; subjects with over 50% false-positive episodes on the implanted monitor were excluded.
Summary: Against an insertable cardiac monitor in 24 patients with paroxysmal atrial fibrillation, the watch algorithm caught 80 of 82 episodes lasting at least 1 hour (episode sensitivity, in words, 97.5%) — but its positive predictive value for episodes was only 39.9% (Wasserlauf 2019)
Result: The implanted monitor detected 82 episodes of atrial fibrillation of at least 1 hour while the watch was worn, totalling 1,127.1 hours; the network detected 80 episodes (episode sensitivity 97.5%) covering 1,101.1 hours (duration sensitivity 97.7%). Patient sensitivity was 83.3% overall and 100% during time actually worn. The positive predictive value for atrial fibrillation episodes was 39.9%.
Watch out: A different device configuration (watch plus ECG band and a neural network) in patients already known to have paroxysmal atrial fibrillation — its low episode positive predictive value reflects false detections against an implanted monitor's own imperfect record, and it does not overturn the notification-level predictive values of the two large studies.
Source: Wasserlauf J, You C, Patel R, Valys A, Albert D, Passman R. Smartwatch Performance for the Detection and Quantification of Atrial Fibrillation. Circulation. Arrhythmia and electrophysiology. 2019. PMID: 31113234. Registry ID: not stated in the record. MeSH: Action Potentials · Aged · Atrial Fibrillation · Deep Learning · Electrocardiography, Ambulatory · Female · Heart Atria · Heart Rate · Humans · Male · Mobile Applications · Predictive Value of Tests · Remote Sensing Technology · Reproducibility of Results · Signal Processing, Computer-Assisted · Wearable Electronic Devices