Glucose sensors on dialysis read high-error but clinically usable: mean absolute relative difference was about 20% for a real-time sensor and 23% for flash monitoring (29% during dialysis sessions), with nearly all values in acceptable error-grid zones

Shelf #10497 · Diabetes, glucose & metabolism

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2021–2026.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately do continuous and flash glucose monitors measure glucose in patients with diabetes on maintenance dialysis, compared with blood glucose — what are the mean absolute relative difference, bias and error-grid performance, on and off dialysis?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: three paired-measurement accuracy studies (31 + 12 + 16 patients; total documented N = 59) in maintenance haemodialysis (one study also included one peritoneal-dialysis patient), comparing sensor glucose with blood glucose. Diagnosis of the measurement problem first: dialysis shifts fluid and solutes between compartments, HbA1c is unreliable in kidney failure, and sensors sample interstitial fluid — so agreement with blood glucose, especially during the dialysis session itself, is the whole question. A real-time sensor (Dexcom G6) showed a mean absolute relative difference of about 20% overall and 22.0% median during dialysis, with a mean bias of about +15 mg/dL, yet nearly all values fell in consensus error-grid zones A and B. A later study found the newer G7 more accurate than the G6 Pro (mean absolute relative difference 16.7% vs 21.2%). Flash monitoring was worse: mean absolute relative difference 23%, rising to 29% during haemodialysis sessions. Sensitivity and specificity are not reported in these records — accuracy here is agreement (bias, mean absolute relative difference, error grids), not a diagnostic 2×2 table. MEASUREMENT ACCURACY — PARTIAL: sensor glucose on dialysis is accurate enough for pattern review and trend-based decisions (almost all readings in clinically acceptable zones), but point accuracy — roughly one fifth off the blood value on average, worse mid-session — is too loose to dose insulin from a single reading without a confirmatory blood check.

MeSH: Humans · Renal Dialysis · Blood Glucose · Male · Female · Middle Aged · Aged · Blood Glucose Self-Monitoring · Diabetes Mellitus · Continuous Glucose Monitoring · Kidney Failure, Chronic · Glycated Hemoglobin · Reproducibility of Results · Diabetes Mellitus, Type 1 · Diabetes Mellitus, Type 2 · Diabetic Nephropathies · Glucose · Prospective Studies · Quality of Life · Adult

Cards (3)

Card 1 — In 31 hospitalised haemodialysis patients with diabetes, Dexcom G6 sensor glucose had a mean absolute relative difference of about 20% against blood glucose (median 22.0% during dialysis vs 18.2% off dialysis) with a mean bias of about +15 mg/dL — yet nearly all values were in error-grid zones A and B (2024 accuracy study)

Study: Paired-measurement accuracy study, N = 31

Population: Thirty-one maintenance haemodialysis patients with diabetes, hospitalised between October 2020 and May 2021, with protocolised paired measurements: Dexcom G6 continuous glucose monitoring versus blood glucose before each meal and at night, plus every 30 minutes during haemodialysis.

Summary: In 31 hospitalised haemodialysis patients with diabetes, Dexcom G6 sensor glucose had a mean absolute relative difference of about 20% against blood glucose (median 22.0% during dialysis vs 18.2% off dialysis) with a mean bias of about +15 mg/dL — yet nearly all values were in error-grid zones A and B (2024 accuracy study)

Result: Pearson and Spearman correlations for averaged sensor versus blood glucose were 0.84 and 0.79; the Bland–Altman mean difference was about +15 mg/dL. Agreement by the %20/20 criterion was 48.7% overall, 47.2% during haemodialysis and 50.2% off dialysis. Mean absolute relative difference was about 20% across all periods; median absolute relative difference was 19.4% overall, 22.0% during haemodialysis and 18.2% off dialysis. Consensus error grids showed nearly all sensor values in clinically acceptable zones A (no harm) and B (unlikely to cause significant harm).

Watch out: Hospitalised patients and a capillary/venous blood comparator rather than a laboratory reference: part of the 20% disagreement is the comparator's own error, but the during-dialysis worsening is consistent across all three studies in this pack.

Source: Narasaki Y et al. Accuracy of Continuous Glucose Monitoring in Hemodialysis Patients With Diabetes. Diabetes care. 2024. PMID: 39213372. Registry ID: not stated in the record. MeSH: Humans · Renal Dialysis · Blood Glucose · Male · Female · Middle Aged · Aged · Blood Glucose Self-Monitoring · Diabetes Mellitus · Continuous Glucose Monitoring

Card 2 — In 12 patients on maintenance dialysis wearing both sensors for 10 days, the Dexcom G7 had a mean absolute relative difference of 16.7% against fingerstick glucose versus 21.2% for the G6 Pro, with 98.0% and 96.7% of values clinically acceptable on error grids (2026 accuracy study)

Study: Paired-measurement accuracy study, N = 12

Population: Twelve participants treated with maintenance dialysis (11 haemodialysis, 1 peritoneal dialysis) with diabetes (mean age 50 ± 12 years, 50% female, 92% using insulin), wearing Dexcom G6 Pro and G7 sensors concurrently for 10 days with self-monitored blood glucose by Contour Next glucometer; 245 paired measurements over 178 sensor-days.

Summary: In 12 patients on maintenance dialysis wearing both sensors for 10 days, the Dexcom G7 had a mean absolute relative difference of 16.7% against fingerstick glucose versus 21.2% for the G6 Pro, with 98.0% and 96.7% of values clinically acceptable on error grids (2026 accuracy study)

Result: Pearson correlations with fingerstick glucose were 0.87 (G6 Pro) and 0.88 (G7), and 0.95 between the two sensors. Mean absolute relative difference versus fingerstick glucose was 21.2% for G6 Pro and 16.7% for G7; excluding one peritoneal-dialysis participant with highly variable glucose, the values were 18.3% and 13.5%. Diabetes Technology Society error grids showed 96.7% (G6 Pro) and 98.0% (G7) of measurements clinically acceptable (zones A/B). Lag times appeared greater than in non-dialysis populations, with substantial between- and within-person variability.

Watch out: Twelve participants is a small accuracy sample, and a single outlier participant moves the mean absolute relative difference by 3 points — the G7-versus-G6 gap should be read as a signal, not a settled ranking.

Source: Zelnick LR et al. Accuracy of Dexcom G6 Pro and G7 Continuous Glucose Monitors in Patients Treated with Maintenance Dialysis. Diabetes technology & therapeutics. 2026. PMID: 40824875. Registry ID: not stated in the record. MeSH: Humans · Female · Middle Aged · Male · Blood Glucose Self-Monitoring · Blood Glucose · Renal Dialysis · Adult · Kidney Failure, Chronic · Glycated Hemoglobin · Reproducibility of Results · Diabetes Mellitus, Type 1 · Diabetes Mellitus, Type 2 · Diabetic Nephropathies · Aged

Card 3 — In 16 patients with diabetes on chronic haemodialysis, flash glucose monitoring (FreeStyle Libre) had a mean absolute relative difference of 23% against capillary glucose — 29% during dialysis sessions — and read lower during dialysis (100.1 mg/dL) than after it (125.1 mg/dL), unrelated to hydration status (2021 accuracy study)

Study: Paired-measurement accuracy study, N = 16

Population: Sixteen patients with diabetes in a chronic haemodialysis programme wearing a FreeStyle Libre flash glucose monitor for 14 days, with 766 paired interstitial and capillary glucose measurements and hydration status measured by bioimpedance spectroscopy.

Summary: In 16 patients with diabetes on chronic haemodialysis, flash glucose monitoring (FreeStyle Libre) had a mean absolute relative difference of 23% against capillary glucose — 29% during dialysis sessions — and read lower during dialysis (100.1 mg/dL) than after it (125.1 mg/dL), unrelated to hydration status (2021 accuracy study)

Result: The correlation coefficient between interstitial and capillary glucose was 0.936, but the mean absolute relative difference was 23%, increasing to 29% during haemodialysis; it was not related to hydration status. Interstitial glucose values were lower during haemodialysis (100.1 ± 17 mg/dL) than in the 12 hours after (125.1 ± 39 mg/dL; P = .012) and on non-dialysis days (134.2 ± 32 mg/dL; P = .001). Patient satisfaction with the device was very high despite the reduced accuracy.

Watch out: The during-dialysis fall in sensor glucose partly reflects true glucose removal by dialysis, not pure sensor error — the study cannot separate physiology from inaccuracy, which is why the pack's claim rests on the agreement statistics, not the session means.

Source: Genua I et al. Accuracy of Flash Glucose Monitoring in Patients with Diabetes Mellitus on Hemodialysis and Its Relationship with Hydration Status. Journal of diabetes science and technology. 2021. PMID: 33251824. Registry ID: not stated in the record. MeSH: Blood Glucose · Blood Glucose Self-Monitoring · Diabetes Mellitus, Type 1 · Glucose · Humans · Prospective Studies · Quality of Life · Renal Dialysis