Point-of-care G6PD testing is accurate only in its quantitative form: biosensors reached pooled sensitivity (in words: the share of truly deficient patients flagged) of 99% at the 30% activity threshold, while a qualitative rapid test's sensitivity fell to 54% at the 70% threshold where heterozygous women sit
Shelf #10623 · Cancer
MEASUREMENT ACCURACY — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2019–2026.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately do point-of-care tests for glucose-6-phosphate dehydrogenase (G6PD) deficiency — quantitative biosensors and qualitative rapid tests — classify enzyme activity against spectrophotometry, at the 30% and 70% residual-activity thresholds used before primaquine or tafenoquine treatment, and what are their sensitivity and specificity (each labelled in words)?
Balance of the evidence
MEASUREMENT ACCURACY — PARTIAL — Composition: two diagnostic-accuracy meta-analyses (13 cross-sectional studies in malaria patients; 4 studies with 7,864 participants for the STANDARD G6PD device) and one primary evaluation in 216 newborns. Diagnosis of the measurement problem first: radical-cure antimalarials can haemolyse G6PD-deficient red cells, so the test's job is to classify activity at two thresholds — 30% (deficient) and 70% (intermediate, the heterozygous-women band) — and a test can be excellent at one and poor at the other. The 2026 meta-analysis found the quantitative STANDARD biosensor at sensitivity (in words: the share of truly deficient patients it flagged) of 0.99 and specificity (in words: the share of non-deficient patients it correctly cleared) of 0.98 at 30%, and 0.96/0.96 at 70%; the qualitative CareStart rapid test managed 0.82/0.96 at 30% but its sensitivity fell to 0.54 at 70%. The device-specific 2024 pool (7,864 participants) agrees: sensitivity 99.1% at 30%, 95.7% at 70%, 90% at 80%. The primary newborn study shows the same threshold gradient in a single centre (sensitivity 100%, specificity 96% at 30%). MEASUREMENT ACCURACY — PARTIAL: quantitative biosensors measure up; qualitative rapid tests cannot be trusted to clear intermediate-activity patients.
MeSH: Humans · Glucosephosphate Dehydrogenase Deficiency · Point-of-Care Systems · Malaria · Sensitivity and Specificity · Rapid Diagnostic Tests · Glucosephosphate Dehydrogenase · Point-of-Care Testing · Biosensing Techniques · Diagnostic Tests, Routine · Adolescent · Blood Donors · Clinical Enzyme Tests · Data Accuracy · Female · Hematologic Tests · Infant, Newborn · Jaundice, Neonatal · Male · Neonatal Screening · Young Adult
Cards (4)
Card 1 — Across 13 cross-sectional studies in malaria patients, the quantitative STANDARD G6PD biosensor detected deficiency with pooled sensitivity (in words: the share of truly deficient patients it flagged) of 0.99 at the 30% activity threshold and 0.96 at 70%, while the qualitative CareStart rapid test fell to sensitivity 0.54 at the 70% threshold (2026 meta-analysis)
Study: Systematic review and meta-analysis of diagnostic accuracy, 13 cross-sectional studies
Population: People with malaria tested for glucose-6-phosphate dehydrogenase (G6PD) deficiency with point-of-care devices, against spectrophotometry as the reference standard, at the clinically used 30% and 70% residual-activity thresholds. Sensitivity and specificity are labelled in words in this pack's balance at first occurrence.
Summary: Across 13 cross-sectional studies in malaria patients, the quantitative STANDARD G6PD biosensor detected deficiency with pooled sensitivity (in words: the share of truly deficient patients it flagged) of 0.99 at the 30% activity threshold and 0.96 at 70%, while the qualitative CareStart rapid test fell to sensitivity 0.54 at the 70% threshold (2026 meta-analysis)
Result: STANDARD G6PD biosensor: pooled sensitivity 0.99 (95% CI 0.96–1.00) and specificity 0.98 (0.97–0.99) at 30%; sensitivity 0.96 (0.90–0.98) and specificity 0.96 (0.89–0.99) at 70%. CareStart rapid test: sensitivity 0.82 (0.55–0.95) and specificity 0.96 (0.93–0.98) at 30%; sensitivity 0.54 (0.31–0.76) and specificity 0.96 (0.84–0.99) at 70%.
Watch out: The two device classes are not interchangeable: the qualitative rapid test misses roughly half of patients in the intermediate (70% threshold) band — the band that matters most for heterozygous women before primaquine or tafenoquine.
Source: Unnikrishnan G, Singh A et al.. Diagnostic Accuracy of Point-of-Care Devices for Glucose-6-Phosphate Dehydrogenase Deficiency Detection Among Malaria Patients: A Systematic Review and Meta-Analysis. Tropical medicine & international health. 2026. PMID: 41774916. Registry ID: PROSPERO CRD420251122222, as stated in the record. MeSH: Humans · Glucosephosphate Dehydrogenase Deficiency · Point-of-Care Systems · Malaria · Sensitivity and Specificity · Rapid Diagnostic Tests · Glucosephosphate Dehydrogenase
Card 2 — In 7,864 participants across four studies, the STANDARD G6PD test (SD Biosensor) pooled sensitivity was 99.1% at the 30% threshold, 95.7% at 70% and 90% at 80%, with specificity 97.4%, 92.9% and 89.0% respectively (2024 meta-analysis)
Study: Systematic review and meta-analysis of diagnostic accuracy, 4 studies, N = 7,864
Population: Patients suspected of G6PD deficiency whose G6PD activity was measured with the point-of-care STANDARD G6PD test (SD Biosensor, named as in the record) and by spectrophotometry, pooled across all thresholds.
Summary: In 7,864 participants across four studies, the STANDARD G6PD test (SD Biosensor) pooled sensitivity was 99.1% at the 30% threshold, 95.7% at 70% and 90% at 80%, with specificity 97.4%, 92.9% and 89.0% respectively (2024 meta-analysis)
Result: Pooled sensitivity was 99.1%, 95.7% and 90% at the 30%, 70% and 80% enzyme-activity thresholds; pooled specificity was 97.4%, 92.9% and 89.0% at the same thresholds. Two of the four included studies had high risk of bias in QUADAS-2 domains 2 and 3, and heterogeneity was high at the 80% threshold.
Watch out: Only four studies underpin this device-specific pool, and accuracy falls as the threshold rises toward normal activity — a test that is excellent at ruling in severe deficiency is weaker at classifying intermediate activity.
Source: Martínez JC, Vélez-Marín V, Lopez-Perez M et al.. Diagnostic accuracy of the point-of-care standard G6PD test™ (SD Biosensor) for glucose-6-phosphate dehydrogenase deficiency: a systematic review and meta-analysis. Malaria journal. 2024. PMID: 39488711. Registry ID: not stated in the record. MeSH: Glucosephosphate Dehydrogenase Deficiency · Humans · Point-of-Care Systems · Sensitivity and Specificity · Glucosephosphate Dehydrogenase · Point-of-Care Testing · Biosensing Techniques · Diagnostic Tests, Routine
Card 3 — In 216 newborns, the careSTART G6PD biosensor (Access Bio, named as in the record) at 30% residual activity had sensitivity 100% and specificity 96% against the standard quantitative assay, with enzyme activity strongly dependent on haematocrit (2019 evaluation)
Study: Primary diagnostic evaluation, N = 216 newborns analysed (plus 40 adults in pre-analytic work)
Population: Two hundred and sixteen newborns screened for G6PD deficiency (plus 40 adults in pre-analytic experiments on anticoagulant, haematocrit, storage temperature and time), tested with the point-of-care careSTART biosensor against the fluorescent spot test and the standard quantitative enzyme assay.
Summary: In 216 newborns, the careSTART G6PD biosensor (Access Bio, named as in the record) at 30% residual activity had sensitivity 100% and specificity 96% against the standard quantitative assay, with enzyme activity strongly dependent on haematocrit (2019 evaluation)
Result: At 30% residual activity the biosensor showed sensitivity 100%, specificity 96%, positive predictive value 73%, negative predictive value 100% and accuracy 97%; correlation with the standard assay was r = 0.82 with intraclass correlation 0.90. Finger-prick and venous values did not differ (p = 0.152), activity rose with haematocrit, and samples were stable at 4°C for 3 days. Sex-stratified identification of deficiency was weaker (sensitivity 78%, specificity 89%).
Watch out: A newborn-screening population, not malaria patients before radical cure: the 100% sensitivity figure comes from the 30% threshold in this single-centre sample and should not be read across to intermediate-activity adults.
Source: Pengboon P, Thamwarokun A, Changsri K et al.. Evaluation of quantitative biosensor for glucose-6-phosphate dehydrogenase activity detection. PloS one. 2019. PMID: 31860695. Registry ID: not stated in the record. MeSH: Adolescent · Biosensing Techniques · Blood Donors · Clinical Enzyme Tests · Data Accuracy · Female · Glucosephosphate Dehydrogenase · Glucosephosphate Dehydrogenase Deficiency · Hematologic Tests · Humans · Infant, Newborn · Jaundice, Neonatal · Male · Neonatal Screening · Point-of-Care Systems · Sensitivity and Specificity · Young Adult
Card 4 — Synthesis — MEASUREMENT ACCURACY — PARTIAL: quantitative biosensors are accurate, qualitative rapid tests miss intermediate deficiency
Study: Synthesis
Population: Two meta-analyses (13 studies in malaria patients; 4 studies, 7,864 participants, for one named device) and one primary newborn evaluation (216 newborns).
Summary: Synthesis — MEASUREMENT ACCURACY — PARTIAL: quantitative biosensors are accurate, qualitative rapid tests miss intermediate deficiency
Result: Quantitative point-of-care biosensors classify G6PD activity well at the 30% threshold (pooled sensitivity about 99%) and acceptably at 70% (about 96%), whereas the qualitative CareStart rapid test's sensitivity collapses to 0.54 at the 70% threshold — precisely where heterozygous women sit before 8-aminoquinoline treatment. Accuracy is therefore device- and threshold-specific, not a class property.
Watch out: Distinct from live shelf #1182 (tafenoquine/primaquine treatment regimens for vivax relapse) and pack #1703 (single low-dose primaquine for gametocyte clearance): this shelf is only about how accurately the pre-treatment safety test measures enzyme activity.
Source: Synthesis of the reports carded above; no additional source.