SourceMed · Shelf #10625 in the ask interface
AI chest-X-ray reading for tuberculosis is a sensitive but leaky triage test: pooled sensitivity (in words: the share of tuberculosis cases flagged) of 86–91% across five products, specificity (in words: the share of non-cases correctly cleared) of only 59–80%, and accuracy that changes with each software version
Shelf #10625 · Lungs, breathing & sleep
MEASUREMENT ACCURACY — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2016–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
How accurately does artificial-intelligence computer-aided detection (CAD) software reading digital chest X-rays detect pulmonary tuberculosis against microbiological reference standards — what sensitivity and specificity (each labelled in words) do the products achieve, and how much does accuracy change between software versions?
Balance of the evidence
MEASUREMENT ACCURACY — PARTIAL — Composition: two systematic reviews (21 studies of five products, 2025; five early CAD4TB studies, 2016), a 12,890-image version comparison, and one prospective study of 2,190 people in Karachi. Diagnosis of the measurement problem first: CAD is deployed as a triage reader where radiologists are scarce, so its operating point trades missed cases against confirmatory-test volume. Pooled product performance in 2025 was sensitivity (in words: the share of tuberculosis cases flagged) of 86–91% with specificity (in words: the share of people without tuberculosis correctly cleared) of only 59–80%; the prospective culture-referenced study sits inside that range (sensitivity 87.7%, specificity 64.3% at the recommended threshold). The version comparison is the caution: CAD4TB improved from AUC 0.823 (version 6) to 0.903 (version 7) and only the newer versions met WHO triage targets, so accuracy belongs to a version, not a brand. MEASUREMENT ACCURACY — PARTIAL: good enough to triage, not to diagnose, and perishable with every software update.
MeSH: Adult · Deep Learning · Humans · Mycobacterium tuberculosis · Prospective Studies · Sensitivity and Specificity · Software · Tuberculosis · Tuberculosis, Pulmonary · X-Rays · Diagnosis, Computer-Assisted · Image Processing, Computer-Assisted · Prevalence · Radiography
Cards (5)
Card 1 — Across 21 studies, five commercial AI chest-X-ray products for tuberculosis had sensitivity (in words: the share of tuberculosis cases flagged) of 86–91% and specificity (in words: the share of non-cases correctly cleared) of only 59–80%, varying by product (2025 meta-analysis)
Study: Systematic review and meta-analysis of diagnostic accuracy, 21 studies; participant total not documented in the record
Population: People evaluated for pulmonary tuberculosis with digital chest radiography read by one of five artificial-intelligence computer-aided detection products (JF CXR-1, qXR, Lunit INSIGHT CXR, CAD4TB, InferRead DR Chest), pooled from studies published to December 2024. Sensitivity and specificity are labelled in words in this pack's balance at first occurrence. The review's abstract does not state a pooled participant total, and none is invented here.
Summary: Across 21 studies, five commercial AI chest-X-ray products for tuberculosis had sensitivity (in words: the share of tuberculosis cases flagged) of 86–91% and specificity (in words: the share of non-cases correctly cleared) of only 59–80%, varying by product (2025 meta-analysis)
Result: Product-level sensitivity/specificity: JF CXR-1 86.0%/80.0%; qXR 90.0%/64.0%; Lunit INSIGHT CXR 90.0%/63.0%; CAD4TB 91.0%/60.0%; InferRead DR Chest 89.0%/59.0%. The authors note that scenario-specific threshold adjustment remains a challenge.
Watch out: These are per-product pooled pairs at each study's chosen threshold, not head-to-head measurements in one population; specificity near 60% means large false-positive triage volumes at screening prevalences.
Source: Han ZL, Zhang YY, Li J et al.. A systematic review and meta-analysis of artificial intelligence software for tuberculosis diagnosis using chest X-ray imaging. Journal of thoracic disease. 2025. PMID: 40529749. Registry ID: not stated in the record. MeSH: none carried in the record
Card 2 — In 12,890 chest X-rays, newer software versions performed clearly better: CAD4TB version 7 reached AUC 0.903 and qXR version 3 AUC 0.906, meeting WHO triage target-product-profile values that their predecessor versions did not meet (2022 comparison)
Study: Retrospective version-comparison evaluation, 12,890 chest X-rays
Population: A case-control sample of 12,890 chest X-rays used to compare successive versions of two computer-aided detection products (CAD4TB versions 6 and 7; qXR versions 2 and 3) against radiologist readings and the WHO target product profile for a tuberculosis triage test. The record carries no MeSH terms.
Summary: In 12,890 chest X-rays, newer software versions performed clearly better: CAD4TB version 7 reached AUC 0.903 and qXR version 3 AUC 0.906, meeting WHO triage target-product-profile values that their predecessor versions did not meet (2022 comparison)
Result: AUCs: CAD4TB v6 0.823 (0.816–0.830), CAD4TB v7 0.903 (0.897–0.908); qXR v2 0.872 (0.866–0.878), qXR v3 0.906 (0.901–0.911). Newer versions met WHO target-product-profile values and equalled or surpassed human radiologists; all products performed worse in older age groups and in people with previous tuberculosis.
Watch out: A case-control image set enriched for tuberculosis, not a consecutive screening cohort — and the central lesson is version drift: accuracy figures expire when the software is updated, so a shelf number here is a snapshot, not a property of the product name.
Source: Qin ZZ, Barrett R, Ahmed S et al.. Comparing different versions of computer-aided detection products when reading chest X-rays for tuberculosis. PLOS digital health. 2022. PMID: 36812562. Registry ID: not stated in the record. MeSH: none carried in the record
Card 3 — In 2,190 symptomatic adults and household contacts in Karachi, Lunit INSIGHT (version 3.1.0.0) at its recommended threshold of 30 had sensitivity 87.7% and specificity 64.3% against liquid culture, with sensitivity lower in smear-negative disease (2022 prospective evaluation)
Study: Prospective diagnostic accuracy study, N = 2,190
Population: Symptomatic adults and household contacts of people with tuberculosis seeking care in Karachi, Pakistan; 269 of 2,190 participants (12%) had culture-confirmed pulmonary tuberculosis.
Summary: In 2,190 symptomatic adults and household contacts in Karachi, Lunit INSIGHT (version 3.1.0.0) at its recommended threshold of 30 had sensitivity 87.7% and specificity 64.3% against liquid culture, with sensitivity lower in smear-negative disease (2022 prospective evaluation)
Result: At threshold score 30: sensitivity 87.7% (95% CI 83.2–91.4) and specificity 64.3% (62.1–66.4). At score 15, sensitivity 88.1% and specificity 57.9%; at score 45, sensitivity 86.6% and specificity 69.9%. Specificity fell with increasing age, previous tuberculosis and lower body-mass index; diabetes and smoking did not modify accuracy.
Watch out: A care-seeking population in which most tuberculosis was smear-positive flatters sensitivity; the authors themselves describe the manufacturer thresholds as giving limited sensitivity for triage.
Source: Tavaziva G, Majidulla A, Nazish A et al.. Diagnostic accuracy of a commercially available, deep learning-based chest X-ray interpretation software for detecting culture-confirmed pulmonary tuberculosis. International journal of infectious diseases. 2022. PMID: 35597555. Registry ID: not stated in the record. MeSH: Adult · Deep Learning · Humans · Mycobacterium tuberculosis · Prospective Studies · Sensitivity and Specificity · Software · Tuberculosis · Tuberculosis, Pulmonary · X-Rays
Card 4 — The earlier systematic review found only five eligible studies (2010–2015), all of the single then-available product CAD4TB, with AUC ranging from 0.71 to 0.84 and a direct trade-off: settings that raised sensitivity cost specificity (2016 review)
Study: Systematic review of diagnostic accuracy, 5 studies
Population: Studies of computer-aided detection of pulmonary tuberculosis on digital chest radiographs against a microbiological reference standard (smear, culture or PCR); tuberculosis prevalence in the included studies ranged from 18% to 60% and HIV prevalence from 33% to 68%.
Summary: The earlier systematic review found only five eligible studies (2010–2015), all of the single then-available product CAD4TB, with AUC ranging from 0.71 to 0.84 and a direct trade-off: settings that raised sensitivity cost specificity (2016 review)
Result: Five of 455 identified articles were eligible, all evaluating CAD4TB. AUC ranged from 0.71 to 0.84. Risk of bias was low in the two prospective studies and high in the three retrospective studies, and the reviewers judged the evidence base small and hard to generalise to lower-prevalence settings.
Watch out: A 2016 snapshot of a single early product; it is carded as the baseline the newer evaluations improved on, not as a current estimate for any software in use today.
Source: Pande T, Cohen C, Pai M et al.. Computer-aided detection of pulmonary tuberculosis on digital chest radiographs: a systematic review. The international journal of tuberculosis and lung disease. 2016. PMID: 27510250. Registry ID: not stated in the record. MeSH: Diagnosis, Computer-Assisted · Humans · Image Processing, Computer-Assisted · Prevalence · Radiography · Sensitivity and Specificity · Software · Tuberculosis, Pulmonary
Card 5 — Synthesis — MEASUREMENT ACCURACY — PARTIAL: modern CAD reads chest X-rays for tuberculosis with high sensitivity but modest specificity, and accuracy is version-specific
Study: Synthesis
Population: Two meta-analyses/reviews (21 studies; 5 early studies), one 12,890-image version comparison and one prospective study of 2,190 people. Documented participant total: 2,190 people plus 12,890 images; review participant totals are not documented in their records.
Summary: Synthesis — MEASUREMENT ACCURACY — PARTIAL: modern CAD reads chest X-rays for tuberculosis with high sensitivity but modest specificity, and accuracy is version-specific
Result: Current products flag roughly 86–91% of tuberculosis cases but clear only about 59–80% of non-cases at studied thresholds, so CAD works as a sensitive triage reader that generates substantial confirmatory-testing volume. Version upgrades materially changed accuracy (AUC 0.82 to 0.90 for one product), which caps any accuracy claim at the version actually evaluated.
Watch out: Distinct from live shelf #8570 (Xpert cartridge accuracy on sputum) and live #2157 (point-of-care Xpert implementation): this shelf is only about software reading of the chest image itself, before any sputum test.
Source: Synthesis of the reports carded above; no additional source.