The MeMed BV host-response score classifies bacterial versus viral infection at about 90% sensitivity (in words: the share of bacterial cases flagged) and 93% specificity (in words: the share of non-bacterial cases correctly cleared) — against expert panels, with about 7% of results equivocal

Shelf #10627 · Critical care & sepsis

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2023–2026.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

How accurately does the MeMed BV host-response test (TRAIL, IP-10 and CRP combined into one score) distinguish bacterial from viral or other non-bacterial infection in patients with suspected acute infection — what sensitivity and specificity (each labelled in words) does it achieve against adjudicated reference standards?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: one meta-analysis (12 accuracy studies within 16 included), one prospective multicentre accuracy study (476 patients) and one paediatric fever-without-source study (241 children). Diagnosis of the measurement problem first: there is no microbiological gold standard for 'bacterial versus viral', so every figure here is agreement with blinded expert adjudication, and the test returns an equivocal band (7.2% in Apollo) that headline figures set aside. Within those limits the performance is consistent: pooled sensitivity (in words: the share of bacterial infections flagged) of 91% and specificity (in words: the share of non-bacterial infections correctly cleared) of 92% in the meta-analysis; 90.0%/92.8% in Apollo; and AUC 0.827 in febrile children under 3 years, comparable to the Labscore. MEASUREMENT ACCURACY — PARTIAL: reproducible classification against panels, unproven as an outcome-changing decision tool, and silent in its equivocal zone.

MeSH: Humans · Child · Infant · Child, Preschool · Prospective Studies · Chemokine CXCL10 · Biomarkers · Fever · C-Reactive Protein · Bacterial Infections · Tumor Necrosis Factors

Cards (4)

Card 1 — Across 12 diagnostic accuracy studies in a 2026 meta-analysis, the MeMed BV host-response test separated bacterial from viral or other non-bacterial infection with pooled sensitivity (in words: the share of bacterial cases flagged) of 91% and specificity (in words: the share of non-bacterial cases correctly cleared) of 92%

Study: Systematic review and meta-analysis, 16 studies (12 diagnostic accuracy)

Population: Adults and children with suspected respiratory or urinary tract infection or undifferentiated fever in emergency and urgent-care settings, tested with the MeMed BV assay (MeMed, named as in the record), which combines TRAIL, IP-10 and CRP into one score, against adjudicated reference standards. Sensitivity and specificity are labelled in words in this pack's balance at first occurrence. The record carries no MeSH terms.

Summary: Across 12 diagnostic accuracy studies in a 2026 meta-analysis, the MeMed BV host-response test separated bacterial from viral or other non-bacterial infection with pooled sensitivity (in words: the share of bacterial cases flagged) of 91% and specificity (in words: the share of non-bacterial cases correctly cleared) of 92%

Result: Pooled sensitivity was 91% (95% CI 86–94) and pooled specificity 92% (91–93); in adults sensitivity was 93% and specificity 91%, and in children 88% and 93%. In four non-accuracy studies in the same review, antibiotics were prescribed in 20.6% of viral-classified versus 73.2% of bacterial-classified cases, with clinician adherence to the test result of 75–80%.

Watch out: The reference standard is expert adjudication, not a microbiological gold standard that does not exist for this question — pooled accuracy is accuracy against panels, and equivocal-zone results are excluded from the headline pair.

Source: Moola S, Carrol ED, Rothman R et al.. Diagnostic Performance of the MeMed BV Test to Distinguish Between Bacterial and Viral or Other Non-Bacterial Causes Amongst ED and Urgent Care Patients: A Systematic Review with Meta-Analysis. Diagnostics. 2026. PMID: 42351588. Registry ID: not stated in the record. MeSH: none carried in the record

Card 2 — In 476 adults and children with suspected acute infection (Apollo study), MeMed BV had sensitivity 90.0% and specificity 92.8% against blinded expert-panel adjudication, with 7.2% of results equivocal and an AUC of 0.95

Study: Prospective multicentre blinded diagnostic accuracy study, N = 476 analysed

Population: Patients older than 3 months presenting with fever and clinical suspicion of acute infection at nine emergency departments and three urgent-care centres in the United States and Israel; 476 of 563 enrolled patients formed the study population (314 adults, 162 children), 60.5% with upper respiratory infection. The record carries no MeSH terms.

Summary: In 476 adults and children with suspected acute infection (Apollo study), MeMed BV had sensitivity 90.0% and specificity 92.8% against blinded expert-panel adjudication, with 7.2% of results equivocal and an AUC of 0.95

Result: Sensitivity was 90.0% (95% CI 80.3–99.7), specificity 92.8% (90.0–95.5) and negative predictive value 98.8% (96.8–99.6) for bacterial infection; 7.2% of cases gave equivocal scores and the area under the curve was 0.95 (0.90–0.99). Treating clinicians did not see the test result.

Watch out: The wide sensitivity interval (80.3–99.7) reflects few adjudicated bacterial cases; the study measures classification against a panel, in a population dominated by upper respiratory infections.

Source: Bachur RG, Kaplan SL, Arias CA et al.. A rapid host-protein test for differentiating bacterial from viral infection: Apollo diagnostic accuracy study. Journal of the American College of Emergency Physicians open. 2024. PMID: 38721037. Registry ID: not stated in the record. MeSH: none carried in the record

Card 3 — In 241 children under 3 years with fever without source, the TRAIL/IP-10/CRP host signature (ImmunoXpert, the same three-protein combination) reached AUC 0.827 for bacterial infection, comparable to the Labscore (0.854) and better than CRP alone (0.807) (2023 study)

Study: Prospective diagnostic accuracy study, N = 241

Population: Children younger than 3 years presenting with fever without source at a single tertiary centre; 68 (28%) were adjudicated bacterial infections and 5 (2%) invasive bacterial infections, with aetiology assigned by three independent experts.

Summary: In 241 children under 3 years with fever without source, the TRAIL/IP-10/CRP host signature (ImmunoXpert, the same three-protein combination) reached AUC 0.827 for bacterial infection, comparable to the Labscore (0.854) and better than CRP alone (0.807) (2023 study)

Result: AUC values: Labscore 0.854 (0.804–0.905), the host signature 0.827 (0.764–0.890), CRP 0.807 (0.744–0.869). The signature and Labscore outperformed single biomarkers, with slightly reduced sensitivity in infants younger than 90 days.

Watch out: Only five invasive bacterial infections occurred, so the study cannot support claims about the outcome that matters most in this age group; it is carded as the paediatric fever-without-source limb, not as a second estimate of the adult emergency-department figure.

Source: Lacroix L, Papis S, Mardegan C et al.. Host biomarkers and combinatorial scores for the detection of serious and invasive bacterial infection in pediatric patients with fever without source. PloS one. 2023. PMID: 37956117. Registry ID: not stated in the record. MeSH: Humans · Child · Infant · Child, Preschool · Prospective Studies · Chemokine CXCL10 · Biomarkers · Fever · C-Reactive Protein · Bacterial Infections · Tumor Necrosis Factors

Card 4 — Synthesis — MEASUREMENT ACCURACY — PARTIAL: a consistent ~90%/~93% classification against adjudication panels, with an equivocal zone and no microbiological gold standard

Study: Synthesis

Population: One meta-analysis (12 accuracy studies), one prospective multicentre study (476 patients) and one paediatric study (241 children).

Summary: Synthesis — MEASUREMENT ACCURACY — PARTIAL: a consistent ~90%/~93% classification against adjudication panels, with an equivocal zone and no microbiological gold standard

Result: The three-protein host-response score classifies bacterial versus viral/non-bacterial infection at about 90% sensitivity and 93% specificity wherever it has been measured, with roughly 7% of results falling in an equivocal band that the headline figures exclude. What is not established by any carded record is that acting on the score improves patient outcomes; the stewardship signal comes from non-randomised implementation reports inside the review.

Watch out: Distinct from pack #10022 (FebriDx, a two-marker CRP/MxA fingerstick test with its own accuracy record): this shelf covers only the three-protein TRAIL/IP-10/CRP score and its earlier ImmunoXpert branding.

Source: Synthesis of the reports carded above; no additional source.