Dual-chamber leadless pacemaker (Aveir DR): both pivotal performance goals met at 3 months — 90.3% free of serious device or procedure complications and at least 70% atrioventricular synchrony in 97.3% of 300 patients, in a single-arm study
Shelf #10716 · Heart, vessels, blood pressure & clotting
POSSIBLE
Stamped: 2026-11-28.
Evidence published: 2023–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with a conventional indication for dual-chamber pacing, does a two-device leadless pacemaker system (one unit in the right atrium, one in the right ventricle, communicating implant-to-implant) implant successfully, pace the atrium adequately, maintain atrioventricular synchrony, and meet its safety goal at 3 months?
Balance of the evidence
POSSIBLE — Composition: 1 prospective multicentre single-arm pivotal study (300 patients in the primary report; a procedural report covers 452 attempted implants from the same study programme, overlapping the first 300 and not counted twice), documented total N = 300 for the primary endpoints. The system — the Aveir DR dual-chamber leadless pacemaker (Abbott), two percutaneously implanted units communicating implant-to-implant; the tested product and maker are named because leadless pacing performance is device-specific — was implanted successfully in 295 of 300 patients (98.3%). The primary safety endpoint, freedom from device- or procedure-related serious adverse events at 90 days, was met in 90.3% (performance goal 78%); 35 such events occurred in 29 patients. Adequate atrial capture and sensing at 3 months was met in 90.2% (goal 82.5%), and at least 70% atrioventricular synchrony while sitting was achieved in 97.3% (goal 83%). The procedural report found 99% implant success, complications in 11.1% of patients (mostly atrial fibrillation/flutter or complete atrioventricular block, 3.5%), and freedom from complications improving from 89% to 98% with implanter experience. There is no transvenous comparator: the study shows the system meets its own performance goals, not that it matches or beats conventional dual-chamber pacing. POSSIBLE.
MeSH: Humans · Atrial Fibrillation / therapy, etiology · Atrioventricular Block / therapy · Cardiac Pacing, Artificial / adverse effects, methods · Follow-Up Studies · Pacemaker, Artificial / adverse effects · Prospective Studies · Treatment Outcome · Arrhythmias, Cardiac / therapy
Cards (3)
Card 1 — Aveir DR pivotal study: implant success 98.3%, safety endpoint met in 90.3%, ≥70% AV synchrony in 97.3% at 3 months (Knops 2023)
Study: Prospective single-arm pivotal device study
Population: 300 patients with a conventional indication for dual-chamber pacing (63.3% sinus-node dysfunction, 33.3% atrioventricular block) in a prospective multicentre single-group study of the dual-chamber leadless pacemaker system (Aveir DR, Abbott).
Summary: Aveir DR pivotal study: implant success 98.3%, safety endpoint met in 90.3%, ≥70% AV synchrony in 97.3% at 3 months (Knops 2023)
Result: Implantation succeeded in 295 patients (98.3%). Freedom from device- or procedure-related serious adverse events at 90 days was 90.3% (95% CI 87.0 to 93.7), exceeding the 78% performance goal (P<0.001); 35 such events occurred in 29 patients. Adequate atrial capture threshold and sensing at 3 months was met in 90.2% (95% CI 86.8 to 93.6; goal 82.5%); mean atrial capture threshold was 0.82±0.70 V and mean P-wave amplitude 3.58±1.88 mV, and none of the 21 patients with P-wave amplitude below 1.0 mV needed revision for inadequate sensing. At least 70% atrioventricular synchrony while sitting was achieved in 97.3% (95% CI 95.4 to 99.3; goal 83%).
Watch out: Single-arm with performance goals rather than a randomised comparator; 3-month primary follow-up says nothing about battery longevity, retrievability or late complications.
Source: Knops RE et al.. A Dual-Chamber Leadless Pacemaker. The New England journal of medicine. 2023. PMID: 37212442. Registry ID: NCT05252702 (stated in the record). MeSH: Humans · Atrial Fibrillation / therapy, etiology · Atrioventricular Block / therapy · Cardiac Pacing, Artificial / adverse effects, methods · Follow-Up Studies · Pacemaker, Artificial / adverse effects · Prospective Studies · Treatment Outcome · Arrhythmias, Cardiac / therapy
Card 2 — Procedural report: 99% implant success across 452 attempts, complications in 11.1%, and a clear learning curve (Doshi 2025)
Study: Procedural outcomes analysis of the pivotal study programme
Population: 452 patients undergoing de novo dual-chamber leadless pacemaker implantation by 126 physicians within the same AVEIR DR i2i study programme (overlapping the pivotal 300; not an independent cohort).
Summary: Procedural report: 99% implant success across 452 attempts, complications in 11.1%, and a clear learning curve (Doshi 2025)
Result: Implantation succeeded in 446 of 452 patients (99%). Mean introducer insertion-to-removal time was 90±37 minutes and fluoroscopy 20±13 minutes; procedure times fell 19–36% from initial (1–4 implants) to advanced (9+) implanter experience. There were 62 procedural complications in 50 patients (11.1%), predominantly cardiac arrhythmias (atrial fibrillation/flutter or complete atrioventricular block in 16 patients, 3.5%); freedom from complications rose from 89% to 98% with advanced experience (P<0.05).
Watch out: Same programme as the pivotal report — its 452 attempts include the original cohort, so the two Ns must never be added; complication counting here is procedural (30 days), not the 90-day primary safety endpoint.
Source: Doshi RN et al.. Dual-chamber leadless pacemaker implant procedural outcomes: Insights from the AVEIR DR i2i study. Heart rhythm. 2025. PMID: 40089049. Registry ID: NCT05252702 (stated in the record). MeSH: Humans · Female · Male · Pacemaker, Artificial · Aged · Treatment Outcome · Equipment Design · Follow-Up Studies · Arrhythmias, Cardiac / therapy · Fluoroscopy · Middle Aged · Learning Curve · Cardiac Pacing, Artificial / methods
Card 3 — Synthesis: a dual-chamber leadless system that meets its 3-month safety and synchrony goals, with an operator learning curve and no comparator
Study: Synthesis
Population: One single-arm pivotal programme (300 patients for primary endpoints; 452 implant attempts in the procedural report, overlapping).
Summary: Synthesis: a dual-chamber leadless system that meets its 3-month safety and synchrony goals, with an operator learning curve and no comparator
Result: Implant success 98.3–99%; freedom from serious device/procedure complications 90.3% at 90 days; adequate atrial pacing/sensing 90.2%; ≥70% AV synchrony 97.3%. Procedural complications 11.1%, falling with experience. POSSIBLE.
Watch out: Nothing here compares leadless dual-chamber pacing with a transvenous system on complications, infection or outcomes beyond 3 months; those claims would need the comparative record.
Source: Synthesis of the reports carded above; no additional source.