SourceMed · Shelf #10743 in the ask interface

Adding glucagon to a closed-loop system (dual-hormone artificial pancreas) cuts hypoglycaemic events further than insulin-only closed loop in head-to-head crossover trials — but time in range is no better, and every trial lasted hours to days

Shelf #10743 · Diabetes, glucose & metabolism

POSSIBLE

Stamped: 2026-11-28.

Evidence published: 2014–2017.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In people with type 1 diabetes, does a dual-hormone closed-loop system (automated insulin plus glucagon) control glucose better than a single-hormone (insulin-only) closed-loop system or conventional pump therapy — in particular, does the glucagon arm reduce hypoglycaemia beyond what insulin suspension and modulation already achieve?

Balance of the evidence

POSSIBLE — Composition: four randomised crossover trials of dual-hormone versus single-hormone closed loop and/or conventional pump therapy (23, 52 and 17 participants with stated numbers, plus one 24-hour trial whose abstract does not state its enrolment; no long-term parallel trial exists). The pattern is consistent: both closed-loop configurations beat conventional pump therapy on time in range (63% and 62% versus 51% over 24 hours in the Montreal trial), and the bihormonal bionic pancreas lowered mean glucose versus usual pump care in adults and adolescents (7.7 mmol/L (138 mg/dL) versus 8.8 mmol/L (159 mg/dL) after adaptation in adults). The glucagon-specific gain is narrower than the closed-loop gain: in the only trial powered around hypoglycaemia, hypoglycaemic events numbered 9 on dual-hormone, 13 on single-hormone and 52 on conventional therapy, and time in range did not differ between the two closed-loop systems (p=0.75). In the 60-hour outpatient trial the dual-hormone advantage in time below 4.0 mmol/L (72 mg/dL) was a median −2.3% and not significant (p=0.072), reaching significance only at the 3.5 mmol/L (63 mg/dL) and 3.3 mmol/L (59 mg/dL) thresholds; during announced exercise, time below 4.0 mmol/L (72 mg/dL) was a median 0% on dual-hormone versus 11% on single-hormone (p=0.0001). No trial ran longer than 11 days, none measured HbA1c as an outcome of the comparison, and glucagon's nausea and long-term safety in daily use are untested at scale. POSSIBLE: glucagon probably adds a real but modest hypoglycaemia reduction on top of insulin-only closed loop, concentrated at the lowest glucose thresholds and around exercise; it has not been shown to improve overall time in range, and the evidence base is short crossover physiology, not outcomes.

MeSH: Adolescent · Adult · Aged · Blood Glucose/metabolism · Child · Cross-Over Studies · Diabetes Mellitus, Type 1/blood,drug therapy · Drug Therapy, Combination · Glucagon/administration & dosage,therapeutic use · Humans · Hypoglycemia/prevention & control · Hypoglycemic Agents/administration & dosage,therapeutic use · Infusion Pumps, Implantable · Insulin/administration & dosage,therapeutic use · Insulin Infusion Systems · Middle Aged · Pancreas, Artificial · Treatment Outcome · Young Adult

Cards (5)

Card 1 — Montreal 24-hour crossover: dual- and single-hormone closed loop both beat pump therapy on time in range (63% and 62% vs 51%); hypoglycaemic events 9 vs 13 vs 52 — no difference between the two closed-loop systems (2015)

Study: Primary randomised crossover trial

Population: Participants aged 12 years or older with type 1 diabetes (the record's abstract does not state the number enrolled), each attending three 24-hour research-facility visits in randomised order: dual-hormone artificial pancreas, single-hormone artificial pancreas, and conventional insulin pump therapy.

Summary: Montreal 24-hour crossover: dual- and single-hormone closed loop both beat pump therapy on time in range (63% and 62% vs 51%); hypoglycaemic events 9 vs 13 vs 52 — no difference between the two closed-loop systems (2015)

Result: Mean time in the plasma glucose target range over 24 h was 62% (SD 18) on single-hormone, 63% (18) on dual-hormone and 51% (19) on conventional pump therapy; the dual-versus-single difference was not significant (p=0.75). Hypoglycaemic events (below 3.3 mmol/L (59 mg/dL) with symptoms, or below 3.0 mmol/L (54 mg/dL) regardless) numbered 52 on conventional therapy (12 symptomatic), 13 on single-hormone (5 symptomatic) and 9 on dual-hormone (0 symptomatic); nocturnal events were 13, 0 and 0 respectively.

Watch out: Enrolment number is not stated in the record's abstract, so no N is claimed for this card; the visits were supervised 24-hour research-facility stays, not free-living use.

Source: Haidar A et al.. Comparison of dual-hormone artificial pancreas, single-hormone artificial pancreas, and conventional insulin pump therapy for glycaemic control in patients with type 1 diabetes: an open-label randomised controlled crossover trial. The lancet. Diabetes & endocrinology. 2015. PMID: 25434967. Registry ID: NCT01754337. MeSH: Adolescent · Adult · Aged · Blood Glucose/metabolism · Child · Cross-Over Studies · Diabetes Mellitus, Type 1/blood,drug therapy · Drug Therapy, Combination · Glucagon/administration & dosage,therapeutic use · Humans · Hypoglycemia/prevention & control · Hypoglycemic Agents/administration & dosage,therapeutic use · Infusion Pumps, Implantable · Insulin/administration & dosage,therapeutic use · Insulin Infusion Systems · Middle Aged · Pancreas, Artificial · Treatment Outcome · Young Adult

Card 2 — Bihormonal bionic pancreas over 5 days (adults) and 5 days at a diabetes camp (adolescents): mean glucose 7.7 mmol/L (138 mg/dL) in both groups, with less time below 3.9 mmol/L (70 mg/dL) in adults (2014)

Study: Primary randomised crossover trial

Population: 20 adults (5-day outpatient period) and 32 adolescents (5-day period at a diabetes camp) with type 1 diabetes, each randomised in crossover between the wearable bihormonal bionic pancreas (insulin plus glucagon, driven by continuous glucose monitoring) and conventional insulin pump therapy.

Summary: Bihormonal bionic pancreas over 5 days (adults) and 5 days at a diabetes camp (adolescents): mean glucose 7.7 mmol/L (138 mg/dL) in both groups, with less time below 3.9 mmol/L (70 mg/dL) in adults (2014)

Result: In adults, mean glucose over the bionic-pancreas period was 7.7 mmol/L (138 mg/dL) and time below 3.9 mmol/L (70 mg/dL) was 4.8%; after 1 day of adaptation, mean glucose was 7.4 mmol/L (133 mg/dL) on the bionic pancreas versus 8.8 mmol/L (159 mg/dL) on control (P<0.001), with low-glucose time 4.1% versus 7.3% (P=0.01). In adolescents, mean glucose was 7.7 mmol/L (138 mg/dL) versus 8.7 mmol/L (157 mg/dL) (P=0.004), low-glucose time was similar (6.1% vs 7.6%, P=0.23), and hypoglycaemia interventions were less frequent (one per 1.6 days vs one per 0.8 days, P<0.001).

Watch out: The comparator is conventional pump therapy, not an insulin-only closed-loop system — this card shows the bihormonal package works, not that glucagon is the ingredient that matters; the Montreal cards make that comparison.

Source: Russell SJ et al.. Outpatient glycemic control with a bionic pancreas in type 1 diabetes. The New England journal of medicine. 2014. PMID: 24931572. Registry ID: NCT01762059 and NCT01833988. MeSH: Adolescent · Adult · Blood Glucose/analysis · Diabetes Mellitus, Type 1/drug therapy · Equipment Design · Female · Glucagon/administration & dosage · Humans · Hypoglycemia/prevention & control · Hypoglycemic Agents/administration & dosage · Insulin/administration & dosage · Insulin Infusion Systems · Male · Middle Aged · Monitoring, Ambulatory/instrumentation · Pancreas, Artificial · Young Adult

Card 3 — 60-hour free-living crossover in 23 adults: dual-hormone had the fewest hypoglycaemic events (6 vs 14 single-hormone vs 34 control), but the headline difference in time below 4.0 mmol/L (72 mg/dL) was not significant (2017)

Study: Primary randomised crossover trial

Population: 23 adults with type 1 diabetes, each completing three 60-hour outpatient free-living periods in randomised crossover: dual-hormone artificial pancreas, single-hormone artificial pancreas, and sensor-augmented pump therapy.

Summary: 60-hour free-living crossover in 23 adults: dual-hormone had the fewest hypoglycaemic events (6 vs 14 single-hormone vs 34 control), but the headline difference in time below 4.0 mmol/L (72 mg/dL) was not significant (2017)

Result: The median difference between dual- and single-hormone systems in time below 4.0 mmol/L (72 mg/dL) was −2.3% (P=0.072, not significant); at lower thresholds the differences were significant (−1.3% below 3.5 mmol/L (63 mg/dL), P=0.017; −0.7% below 3.3 mmol/L (59 mg/dL), P=0.031). Hypoglycaemic events (below 3.0 mmol/L (54 mg/dL) for 20 minutes) numbered 34 on control therapy, 14 on single-hormone and 6 on dual-hormone, at comparable mean glucose levels. The authors conclude larger studies are needed.

Watch out: 23 participants and 60 hours per arm: the trial is a physiology comparison, and its own primary contrast (time below 4.0 mmol/L (72 mg/dL), dual vs single) missed significance.

Source: Haidar A et al.. Outpatient 60-hour day-and-night glucose control with dual-hormone artificial pancreas, single-hormone artificial pancreas, or sensor-augmented pump therapy in adults with type 1 diabetes: An open-label, randomised, crossover, controlled trial. Diabetes, obesity & metabolism. 2017. PMID: 28094472. Registry ID: not stated in the record. MeSH: Activities of Daily Living · Adult · Blood Glucose/analysis · Combined Modality Therapy/adverse effects · Cross-Over Studies · Diabetes Mellitus, Type 1/blood,drug therapy,therapy · Glucagon/administration & dosage,adverse effects,therapeutic use · Glycated Hemoglobin/analysis · Hormone Replacement Therapy/adverse effects,instrumentation · Humans · Hyperglycemia/prevention & control · Hypoglycemia/chemically induced,epidemiology,prevention & control · Infusion Pumps/adverse effects · Insulin/administration & dosage,adverse effects,therapeutic use · Insulin Infusion Systems/adverse effects · Middle Aged · Monitoring, Ambulatory/instrumentation · Pancreas, Artificial/adverse effects · Patient Education as Topic · Quebec/epidemiology · Risk

Card 4 — Announced-exercise crossover in 17 adults: dual-hormone eliminated time below range during exercise (median 0% vs 11% below 4.0 mmol/L (72 mg/dL) on single-hormone, p=0.0001) (2016)

Study: Primary randomised crossover trial

Population: 17 adults with type 1 diabetes studied in randomised crossover on single-hormone and dual-hormone artificial pancreas during two announced, energy-expenditure-matched exercise sessions (continuous and interval cycling).

Summary: Announced-exercise crossover in 17 adults: dual-hormone eliminated time below range during exercise (median 0% vs 11% below 4.0 mmol/L (72 mg/dL) on single-hormone, p=0.0001) (2016)

Result: Exercise-induced hypoglycaemia occurred in 4 of 17 (23.5%) single-hormone versus 2 of 17 (11.8%) dual-hormone interventions during continuous exercise (p=0.5), and 6 of 15 (40%) versus 1 of 16 (6.25%) during interval exercise (p=0.07). Pooled, median time below 4.0 mmol/L (72 mg/dL) was 11% on single-hormone versus 0% on dual-hormone (p=0.0001), and time in the 4.0–10.0 mmol/L (72–180 mg/dL) range was 71.4% versus 100% (p=0.003). No glucagon side-effects were reported.

Watch out: 17 participants, single supervised exercise sessions with the exercise announced 20 minutes ahead — the setting most favourable to a glucagon effect; unannounced real-world exercise is untested here.

Source: Taleb N et al.. Efficacy of single-hormone and dual-hormone artificial pancreas during continuous and interval exercise in adult patients with type 1 diabetes: randomised controlled crossover trial. Diabetologia. 2016. PMID: 27704167. Registry ID: NCT01930110. MeSH: Adult · Algorithms · Blood Glucose/drug effects · Cross-Over Studies · Diabetes Mellitus, Type 1/drug therapy,physiopathology · Exercise/physiology · Female · Glucagon/therapeutic use · Humans · Hypoglycemic Agents/therapeutic use · Insulin/therapeutic use · Male · Middle Aged · Pancreas, Artificial · Young Adult

Card 5 — Synthesis: closed loop beats pump therapy with or without glucagon; glucagon's own contribution is a further cut in hypoglycaemic events and in time at the lowest thresholds, clearest around exercise — shown only in trials of hours to days

Study: Synthesis

Population: Four randomised crossover trials; 92 participants in the three trials whose records state enrolment (52 + 23 + 17), plus one 24-hour trial whose abstract does not state its number.

Summary: Synthesis: closed loop beats pump therapy with or without glucagon; glucagon's own contribution is a further cut in hypoglycaemic events and in time at the lowest thresholds, clearest around exercise — shown only in trials of hours to days

Result: Versus conventional pump therapy, both configurations raised time in range by about 11–12 percentage points and cut hypoglycaemic events roughly four- to six-fold. Dual versus single hormone: no difference in time in range (p=0.75); fewer hypoglycaemic events (9 vs 13; 6 vs 14 in the two head-to-head trials); significantly less time below 3.5 mmol/L (63 mg/dL) and 3.3 mmol/L (59 mg/dL) but not below 4.0 mmol/L (72 mg/dL) in free-living use; and elimination of below-range time during announced exercise (median 0% vs 11%). POSSIBLE.

Watch out: No trial exceeded 11 days, none was powered for severe hypoglycaemia or HbA1c, and the systems used research glucagon formulations whose stability and tolerability for chronic daily use are not established by these records.

Source: Synthesis of the reports carded above; no additional source.