DXA vertebral fracture assessment finds most vertebral fractures without a separate spine X-ray — pooled sensitivity 0.65–1.00 per patient with specificity 0.74–1.00 — but it systematically misses mild fractures, and performance swings with the scanner generation and the reader

Shelf #10752 · Bones & joints

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2009–2016.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults undergoing bone densitometry, how accurately does vertebral fracture assessment (VFA) by dual-energy X-ray absorptiometry detect prevalent vertebral fractures compared with conventional spinal radiography read by the semiquantitative method — in sensitivity and specificity (each labelled in words), per vertebra and per patient, and for mild versus moderate-to-severe fractures?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: a systematic review of 12 accuracy studies plus two primary multicentre or single-scanner cohorts (350 and 203 patients). The spread between studies is the finding. Across the review, sensitivity and specificity were 0.70–0.93 and 0.95–1.00 per vertebra, and 0.65–1.00 and 0.74–1.00 per patient; in the low-risk-of-bias subgroups, per-vertebra sensitivity was 0.70–0.84 with specificity 0.96–0.99. The best primary study — a modern Lunar iDXA scanner, two readers — reached per-vertebra sensitivity 98.68% and specificity 99.91% and per-patient sensitivity 96.83% and specificity 98.66% against radiography. The multicentre study on earlier-generation equipment found reader-dependent sensitivity of only 0.65–0.84 at specificity 0.97–0.98, improving for moderate and severe fractures (sensitivity 0.70–0.86), with unreadable vertebrae and missed mild fractures the stated failure modes. MEASUREMENT ACCURACY — PARTIAL: on current-generation scanners VFA can stand in for radiography to find prevalent vertebral fractures at the densitometry visit, particularly moderate-to-severe ones; a negative or borderline VFA does not exclude a mild fracture, and quoted accuracy must name the scanner generation and reader.

MeSH: Absorptiometry, Photon/methods · Female · Humans · Male · Osteoporosis, Postmenopausal/complications · Osteoporotic Fractures/diagnostic imaging,etiology · Radiography · Sensitivity and Specificity · Spinal Fractures/diagnostic imaging,etiology

Cards (4)

Card 1 — Systematic review of 12 studies: VFA sensitivity/specificity versus radiography ranged 0.70–0.93 / 0.95–1.00 per vertebra and 0.65–1.00 / 0.74–1.00 per patient; in low-bias subgroups per-vertebra sensitivity was 0.70–0.84 (2016)

Study: Systematic review of diagnostic accuracy

Population: 12 diagnostic-accuracy studies of DXA-based vertebral fracture assessment against spinal radiography in postmenopausal women and elderly men, identified from Medline, EMBASE, the Cochrane Library and four Korean databases to May 2013, with subgroup analyses restricted to studies at low risk of bias.

Summary: Systematic review of 12 studies: VFA sensitivity/specificity versus radiography ranged 0.70–0.93 / 0.95–1.00 per vertebra and 0.65–1.00 / 0.74–1.00 per patient; in low-bias subgroups per-vertebra sensitivity was 0.70–0.84 (2016)

Result: Across studies, sensitivity and specificity were 0.70–0.93 and 0.95–1.00 analysed per vertebra, and 0.65–1.00 and 0.74–1.00 per patient. In the five studies at low risk of bias for the index test, sensitivity and specificity were 0.70–0.84 and 0.96–0.99; in the three low-bias patient-selection studies analysed per vertebra, 0.70–0.93 and 0.96–1.00. The authors conclude VFA has moderate sensitivity and high specificity, and that the evidence is insufficient to say whether radiography should be replaced by VFA.

Watch out: The ranges span scanner generations and reader experience pooled together; the review's own conclusion is that replacement of radiography is not established — the two primary cards below show the best and the multicentre-typical ends of that range.

Source: Lee JH et al.. A systematic review of diagnostic accuracy of vertebral fracture assessment (VFA) in postmenopausal women and elderly men. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2016. PMID: 26782682. Registry ID: not stated in the record. MeSH: Absorptiometry, Photon/methods · Female · Humans · Male · Osteoporosis, Postmenopausal/complications · Osteoporotic Fractures/diagnostic imaging,etiology · Radiography · Sensitivity and Specificity · Spinal Fractures/diagnostic imaging,etiology

Card 2 — Lunar iDXA cohort (350 patients, 4,476 vertebrae visualised): per-vertebra sensitivity 98.68% / specificity 99.91% and per-patient 96.83% / 98.66% against semiquantitative radiography (2012)

Study: Primary diagnostic accuracy study

Population: 350 subjects (269 women, 81 men) who had both conventional spinal radiographs and DXA vertebral fracture assessment on a Lunar iDXA scanner, each read by two readers using the semiquantitative method; 4,476 of 4,550 vertebrae (98.4%) were visualised on VFA versus 4,535 (99.7%) on radiographs, with 231 fractures in 126 patients by radiography and 228 in 125 patients by VFA.

Summary: Lunar iDXA cohort (350 patients, 4,476 vertebrae visualised): per-vertebra sensitivity 98.68% / specificity 99.91% and per-patient 96.83% / 98.66% against semiquantitative radiography (2012)

Result: Agreement between VFA and radiography was excellent (kappa 0.984 per vertebra, 0.957 per patient). Diagnostic performance of VFA was sensitivity 98.68%, specificity 99.91%, positive predictive value 98.25% and negative predictive value 99.93% per vertebra; and sensitivity 96.83%, specificity 98.66%, positive predictive value 97.60% and negative predictive value 98.22% per patient. In patients aged 65 or older with moderate or severe osteoarthritis, radiography and VFA identified 96 and 95 fractures respectively, against 90 vertebral deformities. The authors conclude VFA may be an appropriate alternative to radiography in patients at high fracture risk undergoing densitometry.

Watch out: One scanner model, one centre, expert readers — this is the ceiling of VFA performance, not the expectation for older machines or occasional readers (contrast the multicentre card).

Source: Diacinti D et al.. Diagnostic performance of vertebral fracture assessment by the lunar iDXA scanner compared to conventional radiography. Calcified tissue international. 2012. PMID: 22965625. Registry ID: not stated in the record. MeSH: Absorptiometry, Photon/methods · Adult · Aged · Aged, 80 and over · Female · Humans · Male · Middle Aged · Spinal Fractures/diagnostic imaging · Spine/diagnostic imaging

Card 3 — Multicentre evaluation (203 postmenopausal women, three expert radiologists): sensitivity only 0.65–0.84 at specificity 0.97–0.98 across readers, improving to 0.70–0.86 for moderate and severe fractures; mild fractures and unreadable vertebrae were the failure modes (2009)

Study: Primary diagnostic accuracy study (multicentre)

Population: 203 postmenopausal women imaged by both DXA vertebral fracture assessment and conventional radiography across multiple centres; three radiologists experienced in vertebral fracture assessment read both modalities independently on two occasions using the Genant semiquantitative method.

Summary: Multicentre evaluation (203 postmenopausal women, three expert radiologists): sensitivity only 0.65–0.84 at specificity 0.97–0.98 across readers, improving to 0.70–0.86 for moderate and severe fractures; mild fractures and unreadable vertebrae were the failure modes (2009)

Result: Analysing all readable vertebrae, the kappa statistic ranged from 0.64 to 0.77 between readers and modalities, sensitivity from 0.65 to 0.84, and specificity from 0.97 to 0.98. Restricting to moderate and severe fractures improved kappa to 0.80–0.91 and sensitivity to 0.70–0.86. The authors conclude that despite inferior image quality, VFA correctly identified most moderate and severe fractures, but underperformed because of unreadable vertebrae and reduced sensitivity for mild fractures.

Watch out: Earlier-generation DXA equipment and a 2009 image chain: the gap between this card and the iDXA card is largely technological, so neither number should be quoted without its scanner context.

Source: Fuerst T et al.. Evaluation of vertebral fracture assessment by dual X-ray absorptiometry in a multicenter setting. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2009. PMID: 19083074. Registry ID: not stated in the record. MeSH: Absorptiometry, Photon/methods · Aged · Female · Femur Neck/diagnostic imaging · Humans · Lumbar Vertebrae/diagnostic imaging · Middle Aged · Osteoporosis, Postmenopausal/diagnostic imaging · Retrospective Studies · Sensitivity and Specificity · Spinal Fractures/diagnostic imaging

Card 4 — Synthesis: VFA is highly specific everywhere (0.95–1.00) but its sensitivity is conditional — 0.65–0.84 in multicentre practice on older scanners, 0.97–0.99 per vertebra on a modern scanner with expert readers — and mild (grade 1) fractures are the ones it misses

Study: Synthesis

Population: A 12-study systematic review plus primary cohorts of 350 and 203 patients, all against semiquantitative radiography as the reference standard.

Summary: Synthesis: VFA is highly specific everywhere (0.95–1.00) but its sensitivity is conditional — 0.65–0.84 in multicentre practice on older scanners, 0.97–0.99 per vertebra on a modern scanner with expert readers — and mild (grade 1) fractures are the ones it misses

Result: Per-vertebra sensitivity/specificity: 0.70–0.93 / 0.95–1.00 (review); 98.68% / 99.91% (iDXA cohort); sensitivity 0.65–0.84 / specificity 0.97–0.98 (multicentre, earlier scanners), rising to 0.70–0.86 sensitivity for moderate-to-severe fractures. MEASUREMENT ACCURACY — PARTIAL.

Watch out: Radiography itself is an imperfect reference (the semiquantitative method has its own reader variability, visible in the multicentre kappa range), so 'accuracy' here means agreement with the conventional standard, not with truth.

Source: Synthesis of the reports carded above; no additional source.