SourceMed · Shelf #10938 in the ask interface
Bedside ocular ultrasound identifies retinal detachment with sensitivity (in words: the share of detachments it detects) 94.2% and specificity (in words: the share of non-detachments it correctly clears) 96.3% across 11 studies and 844 patients
Shelf #10938 · Eyes, skin, nose & allergy
MEASUREMENT ACCURACY — GOOD
Stamped: 2026-11-28.
Evidence published: 2015–2019.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In emergency-department patients with ocular symptoms, how accurately does point-of-care ocular ultrasonography diagnose retinal detachment against ophthalmologist assessment — what are its pooled sensitivity and specificity (each labelled in words), and how does it perform for vitreous haemorrhage and vitreous detachment?
Balance of the evidence
MEASUREMENT ACCURACY — GOOD — Composition: two systematic reviews with meta-analysis (11 studies, 844 patients; and 4 trials, 201 patients) plus one multicentre prospective study (225 patients, 75 operators). The larger meta-analysis pools sensitivity (in words: the share of retinal detachments the ultrasound detects) at 94.2% and specificity (in words: the share of patients without detachment it correctly clears) at 96.3%; the earlier review's three low-bias-risk trials found sensitivity (in words: the share of detachments detected) ranging 97% to 100% and specificity (in words: the share of non-detachments cleared) 83% to 100%; and the prospective study found sensitivity (in words: the share of detachments detected) 96.9% and specificity (in words: the share of non-detachments cleared) 88.1% for detachment, but only sensitivity (in words: the share detected) 42.5% for vitreous detachment. Accuracy is for identifying detachment in symptomatic ED patients against ophthalmologist diagnosis, not for screening asymptomatic eyes. MEASUREMENT ACCURACY — GOOD for retinal detachment specifically; the same probe is weak for vitreous detachment.
MeSH: Adult · Emergency Service, Hospital/organization & administration · Female · Humans · Male · Middle Aged · Point-of-Care Systems/standards · Retinal Detachment/diagnostic imaging · Sensitivity and Specificity · Ultrasonography/standards
Cards (4)
Card 1 — Gottlieb 2019 meta-analysis (11 studies, 844 patients): point-of-care ocular ultrasound was 94.2% sensitive and 96.3% specific for retinal detachment
Study: Systematic review and meta-analysis of diagnostic accuracy studies
Population: 11 prospective and randomised studies of point-of-care ocular ultrasound for retinal detachment, identified from six databases and bibliographies, with dual data extraction and QUADAS-2 quality assessment; 844 patients in total.
Summary: Gottlieb 2019 meta-analysis (11 studies, 844 patients): point-of-care ocular ultrasound was 94.2% sensitive and 96.3% specific for retinal detachment
Result: Overall, ultrasound was sensitivity (in words: it detected 94.2% of retinal detachments) 94.2% (95% CI 78.4% to 98.6%) and specificity (in words: it correctly cleared 96.3% of patients without detachment) 96.3% (95% CI 89.2% to 98.8%), with a positive likelihood ratio of 25.2 (a relative measure) and a negative likelihood ratio of 0.06 (a relative measure). Subgroup analysis found ultrasound more accurate among emergency-department patients, with no significant difference between ED and non-ED providers.
Watch out: The confidence interval for sensitivity is wide (78.4% to 98.6%) because the primary studies are small; the pooled figure should be read with that interval, and the review covers symptomatic patients in acute-care settings, not screening.
Source: Gottlieb M et al.. Point-of-Care Ocular Ultrasound for the Diagnosis of Retinal Detachment: A Systematic Review and Meta-Analysis. Academic emergency medicine : official journal of the Society for Academic Emergency Medicine. 2019. PMID: 30636351. Registry ID: none in the record. MeSH: Adult · Emergency Service, Hospital/organization & administration · Female · Humans · Male · Middle Aged · Point-of-Care Systems/standards · Retinal Detachment/diagnostic imaging · Sensitivity and Specificity · Ultrasonography/standards
Card 2 — Lahham 2019: in 225 ED patients scanned by 75 emergency practitioners, POCUS sensitivity for retinal detachment was 96.9% and specificity 88.1% — but sensitivity for vitreous detachment was only 42.5%
Study: Prospective diagnostic accuracy study
Population: 225 adults presenting to 4 emergency departments (2 academic, 2 county) with ocular symptoms and concern for retinal detachment, vitreous haemorrhage or vitreous detachment, scanned by 75 emergency-medicine attending physicians, residents and physician assistants; the reference standard was the final diagnosis of an ophthalmologist masked to the ultrasound findings.
Summary: Lahham 2019: in 225 ED patients scanned by 75 emergency practitioners, POCUS sensitivity for retinal detachment was 96.9% and specificity 88.1% — but sensitivity for vitreous detachment was only 42.5%
Result: Ophthalmologists diagnosed retinal detachment in 47 patients (20.8%), vitreous haemorrhage in 54 (24.0%) and vitreous detachment in 34 (15.1%). For retinal detachment, POCUS sensitivity (in words: the share of detachments detected) was 96.9% (95% CI 80.6% to 99.6%) and specificity (in words: the share of non-detachments cleared) 88.1% (95% CI 81.8% to 92.4%). For vitreous haemorrhage, sensitivity (in words: the share detected) was 81.9% and specificity (in words: the share cleared) 82.3%; for vitreous detachment, sensitivity (in words: the share detected) was only 42.5% while specificity (in words: the share cleared) was 96.0%.
Watch out: A single (large, multicentre) primary study with many operators of mixed seniority; its specificity for detachment (88.1%) sits below the meta-analytic pool (96.3%), and both figures print — the test's weak target is vitreous detachment, which it misses more often than it finds.
Source: Lahham S et al.. Point-of-Care Ultrasonography in the Diagnosis of Retinal Detachment, Vitreous Hemorrhage, and Vitreous Detachment in the Emergency Department. JAMA network open. 2019. PMID: 30977855. Registry ID: none in the record. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Emergency Medicine/methods · Emergency Service, Hospital · Female · Humans · Male · Middle Aged · Point-of-Care Systems/statistics & numerical data · Prospective Studies · Reproducibility of Results · Retinal Detachment/diagnostic imaging · Sensitivity and Specificity · Ultrasonography/methods,statistics & numerical data · Vitreous Detachment/diagnostic imaging · Vitreous Hemorrhage/diagnostic imaging · Young Adult
Card 3 — Vrablik 2015 meta-analysis (4 trials, 201 patients): bedside ocular ultrasonography sensitivity for retinal detachment ranged 97% to 100% and specificity 83% to 100% in the trials assessed
Study: Systematic review and meta-analysis of diagnostic accuracy studies
Population: 4 trials of clinician-performed bedside ocular ultrasonography in emergency-department patients, found from 7,771 citations; 3 trials were judged at low risk of bias and enrolled 201 patients, using 7.5- or 10-MHz linear-array probes against ophthalmologic evaluation (one trial also used orbital CT). Retinal detachment prevalence ranged from 15% to 38%, and two trials included trauma-related detachments.
Summary: Vrablik 2015 meta-analysis (4 trials, 201 patients): bedside ocular ultrasonography sensitivity for retinal detachment ranged 97% to 100% and specificity 83% to 100% in the trials assessed
Result: Across the assessed trials, sensitivity (in words: the share of detachments detected) ranged from 97% to 100% and specificity (in words: the share of non-detachments cleared) from 83% to 100%. The authors conclude bedside ocular ultrasonography has a high degree of accuracy for retinal detachment on the basis of three small prospective investigations, and call for larger prospective validation.
Watch out: Only 201 patients in the low-bias trials, with detachment prevalence (15–38%) far above an unselected ED eye-complaint population; near-perfect sensitivity in small enriched samples usually falls in larger routine series, as the Lahham study's wider intervals illustrate.
Source: Vrablik ME et al.. The diagnostic accuracy of bedside ocular ultrasonography for the diagnosis of retinal detachment: a systematic review and meta-analysis. Annals of emergency medicine. 2015. PMID: 24680547. Registry ID: none in the record. MeSH: Eye/diagnostic imaging · Humans · Point-of-Care Systems · Reference Standards · Retinal Detachment/diagnostic imaging · Sensitivity and Specificity · Ultrasonography
Card 4 — Synthesis
Study: Synthesis
Population: Two meta-analyses (844 and 201 patients) and one 225-patient multicentre prospective study.
Summary: Synthesis
Result: All three sources agree that emergency-performed ocular ultrasound detects retinal detachment with sensitivity (in words: the share of detachments detected) in the mid-90% range and specificity (in words: the share of non-detachments cleared) from the high-80s to mid-90s; the same examination performs much worse for vitreous detachment. MEASUREMENT ACCURACY — GOOD.
Watch out: GOOD applies to retinal detachment in symptomatic ED patients against ophthalmologist diagnosis; it does not extend to vitreous detachment (sensitivity 42.5% in the prospective study) or to asymptomatic screening.
Source: Synthesis of the reports carded above; no additional source.