SourceMed · Shelf #10941 in the ask interface
Automated urine flow cytometry screens for urinary infection moderately well: bacterial count area under the ROC curve 0.86 in 671 specimens, and an interpretable UF-1000i decision tree reached sensitivity (in words: the share of positive cultures it flagged) 94.5% across 15,312 samples
Shelf #10941 · Urology & men’s health
MEASUREMENT ACCURACY — PARTIAL
Stamped: 2026-11-28.
Evidence published: 2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In urine specimens sent for culture, how accurately do automated urine flow cytometers (Sysmex UF series, named as published) screen for urinary tract infection or bacteriuria against urine culture — what discrimination do bacterial and white-cell counts achieve, and what sensitivity and specificity (each labelled in words) do rule-out strategies reach?
Balance of the evidence
MEASUREMENT ACCURACY — PARTIAL — Composition: two primary validation studies, no meta-analysis (15,312 samples from 10,534 patients; 671 specimens of which 246 were culture-positive infections). On the newer UF-5000 analyser, the bacterial count discriminated urinary infection from non-infection with an area under the ROC curve of 0.86 and the white-cell count 0.74; its decision tree reached sensitivity (in words: the share of infections flagged) 0.69 and specificity (in words: the share of non-infections cleared) 0.95. On the earlier UF-1000i, an interpretable decision-tree rule using age, bacteria, mucus and scatter parameters reached sensitivity (in words: the share of positive samples flagged) 94.5% for identifying samples needing culture, cutting laboratory workload by an additional 16% over the existing procedure. The devices named are the tested products (Sysmex UF-5000 and UF-1000i, as published); results do not transfer automatically to other analysers. MEASUREMENT ACCURACY — PARTIAL: useful as a rule-out screen in the laboratories studied, not a replacement for culture.
MeSH: Humans · Nitrates · Urinary Tract Infections/diagnosis,microbiology · Urinalysis · Bacteria · Leukocytes · Sensitivity and Specificity · Flow Cytometry · Urine/microbiology
Cards (3)
Card 1 — Wang 2023: on the Sysmex UF-5000, urine bacterial count discriminated UTI with AUC 0.86 (white-cell count 0.74); a decision tree gave sensitivity 0.69 and specificity 0.95 in 671 specimens
Study: Primary diagnostic accuracy study
Population: 671 consecutively collected urine specimens sent to one hospital's microbiology laboratory for culture over about 2 months (246 urinary tract infections and 425 non-infections, with infection defined as culture above 10⁴ CFU/mL); the UF-5000 measured white blood cells and bacteria simultaneously, and a chemical strip assessed nitrate.
Summary: Wang 2023: on the Sysmex UF-5000, urine bacterial count discriminated UTI with AUC 0.86 (white-cell count 0.74); a decision tree gave sensitivity 0.69 and specificity 0.95 in 671 specimens
Result: Areas under the ROC curve were 0.86 for the bacterial count and 0.74 for the white-cell count; decision-curve analysis showed the bacterial count gave higher net benefit, and the nomogram gave bacteria the largest effect on infection probability. The decision tree's sensitivity (in words: the share of infections it flagged) was 0.69 and its specificity (in words: the share of non-infections it cleared) was 0.95. In specimens with bacteria above 1,367/µL, the analyser's Gram-negative flag had a positive predictive value of 0.93.
Watch out: Single-centre, single-analyser study over one short period, with infection defined by a 10⁴ CFU/mL culture threshold; the tree's sensitivity of 0.69 means it misses about 3 in 10 infections at that operating point — it is a triage rule, not a stand-alone diagnosis.
Source: Wang H et al.. Accuracy of the Sysmex UF-5000 analyzer for urinary tract infection screening and pathogen classification. PloS one. 2023. PMID: 36724192. Registry ID: none in the record. MeSH: Humans · Nitrates · Urinary Tract Infections/diagnosis,microbiology · Urinalysis · Bacteria · Leukocytes · Sensitivity and Specificity · Flow Cytometry · Urine/microbiology
Card 2 — Del Ben 2023: an interpretable decision tree on Sysmex UF-1000i data from 15,312 samples reached sensitivity 94.5% for flagging samples that need culture and cut workload by a further 16%
Study: Primary diagnostic accuracy study
Population: 15,312 urine samples from 10,534 patients at one laboratory, analysed with the Sysmex UF-1000i automated analyser together with clinical features; decision-tree models (with and without a lookahead strategy) were built to identify negative samples before culture, deliberately favouring interpretable rules over black-box models.
Summary: Del Ben 2023: an interpretable decision tree on Sysmex UF-1000i data from 15,312 samples reached sensitivity 94.5% for flagging samples that need culture and cut workload by a further 16%
Result: The best model achieved a sensitivity (in words: the share of culture-positive samples it flagged for culture) of 94.5%, classifying samples using age, bacteria count, mucus and two scattering parameters. It reduced laboratory workload by an additional 16% compared with the laboratory's current procedure, with an estimated saving of about €40,000 per year at 15,000 samples per year. The abstract does not state the model's specificity, so none is printed here.
Watch out: A model-development study in one laboratory's sample stream; sensitivity is reported without a paired specificity in the abstract, so the false-positive cost of the 94.5% figure cannot be judged from the record, and the rule needs external validation before the workload claim travels.
Source: Del Ben F et al.. A fully interpretable machine learning model for increasing the effectiveness of urine screening. American journal of clinical pathology. 2023. PMID: 37658807. Registry ID: none in the record. MeSH: Humans · Urinary Tract Infections/diagnosis,microbiology,urine · Flow Cytometry/methods · Urinalysis/methods · Bacteria · Machine Learning
Card 3 — Synthesis
Study: Synthesis
Population: Two single-centre primary studies (671 specimens; 15,312 samples), on two generations of one manufacturer's analyser; no meta-analysis was found for this question.
Summary: Synthesis
Result: Bacterial counting by urine flow cytometry discriminates infection moderately (AUC 0.86); operating points trade sensitivity against specificity — 0.69/0.95 for one tree, 94.5% sensitivity (specificity unstated in the abstract) for another. MEASUREMENT ACCURACY — PARTIAL.
Watch out: PARTIAL because the base is two single-centre studies without independent external validation or a synthesis, and one card's specificity is not stated in its record — both limits are printed rather than filled in.
Source: Synthesis of the reports carded above; no additional source.