Rapid lateral-flow tests for leptospirosis miss about a third of cases: pooled sensitivity (in words: the share of cases detected) 68% and specificity (in words: the share of non-cases correctly cleared) 93% across 24 studies, with individual-test sensitivity ranging from 3.6% to 100%

Shelf #10942 · Critical care & sepsis

MEASUREMENT ACCURACY — PARTIAL

Stamped: 2026-11-28.

Evidence published: 2022–2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with suspected leptospirosis, how accurately do rapid lateral-flow immunoassays and immunochromatographic kits detect the infection against microscopic agglutination testing or immunofluorescence reference standards — what are the pooled sensitivity and specificity (each labelled in words), and how wide is the spread between commercial kits?

Balance of the evidence

MEASUREMENT ACCURACY — PARTIAL — Composition: one systematic review and meta-analysis (24 eligible records, 49 data entries) and one systematic review of immunochromatographic kits (22 included articles). The meta-analysis pools sensitivity (in words: the share of leptospirosis cases the rapid tests detect) at 68% (95% CI 57% to 78%) and specificity (in words: the share of non-cases they correctly clear) at 93% (95% CI 90% to 95%) against microscopic agglutination or immunofluorescence references, with individual entries ranging from sensitivity (in words: the share detected) 3.6% to 100% and specificity (in words: the share cleared) 53.5% to 100%; only about a third of studies were at low risk of bias. The kit review found the same spread (sensitivity 15.8% to 100.0%; specificity 37.3% to 100.0%, each labelled in words as the share detected and the share cleared) and identified one lipopolysaccharide-specific assay family at the top (sensitivity about 93–100%, specificity about 99–100%, each labelled in words). MEASUREMENT ACCURACY — PARTIAL: a positive rapid test supports leptospirosis, but the pooled test misses about one case in three and kit performance varies too widely for the class figure to describe any single product.

MeSH: Leptospirosis/diagnosis,immunology · Humans · Immunoassay/methods · Sensitivity and Specificity · Antibodies, Bacterial/blood · Leptospira/immunology,isolation & purification · Reproducibility of Results

Cards (3)

Card 1 — Nualnoi 2024 meta-analysis (24 records, 49 data entries): rapid lateral-flow immunoassays for leptospirosis pooled at sensitivity 68% and specificity 93% against MAT or immunofluorescence

Study: Systematic review and meta-analysis of diagnostic accuracy studies

Population: 24 eligible records published between 2003 and 2023 reporting the accuracy of human leptospirosis lateral-flow immunoassays against microscopic agglutination tests (MAT) or immunofluorescent assays (IFA), contributing 49 data entries; study quality was assessed with QUADAS-2 and only nine studies (32.1%) were at low risk of bias with no applicability concern.

Summary: Nualnoi 2024 meta-analysis (24 records, 49 data entries): rapid lateral-flow immunoassays for leptospirosis pooled at sensitivity 68% and specificity 93% against MAT or immunofluorescence

Result: Pooled sensitivity (in words: the share of cases detected) was 68% (95% CI 57% to 78%) and pooled specificity (in words: the share of non-cases correctly cleared) was 93% (95% CI 90% to 95%), but individual entries ranged from sensitivity (in words: the share detected) 3.6% to 100% and specificity (in words: the share cleared) 53.5% to 100%. IgM-detecting assays were more sensitive than IgG-only assays, and performance did not appear to vary by phase of infection. The authors urge caution because of substantial heterogeneity.

Watch out: The pooled pair averages across many different commercial kits and study designs with substantial heterogeneity; it describes the class, not any kit a clinician might actually hold, and two-thirds of the underlying studies carried bias or applicability concerns.

Source: Nualnoi T et al.. Accuracy of rapid lateral flow immunoassays for human leptospirosis diagnosis: A systematic review and meta-analysis. PLoS neglected tropical diseases. 2024. PMID: 38748731. Registry ID: none in the record. MeSH: Leptospirosis/diagnosis,immunology · Humans · Immunoassay/methods · Sensitivity and Specificity · Antibodies, Bacterial/blood · Leptospira/immunology,isolation & purification · Reproducibility of Results

Card 2 — Clemente 2022 review (22 articles): immunochromatographic leptospirosis kits ranged from sensitivity 15.8% to 100.0% and specificity 37.3% to 100.0%, with an LPS-specific assay family highest at about 93–100% and 99–100%

Study: Systematic review of diagnostic accuracy studies

Population: 22 articles on immunochromatographic test (ICT) kits for acute human leptospirosis, selected from 41 candidate articles across PubMed, Cochrane Library and Google Scholar including grey literature, with methodological quality assessed by QUADAS-2; almost all kits detect IgM antibodies against Leptospira species, one used IgG.

Summary: Clemente 2022 review (22 articles): immunochromatographic leptospirosis kits ranged from sensitivity 15.8% to 100.0% and specificity 37.3% to 100.0%, with an LPS-specific assay family highest at about 93–100% and 99–100%

Result: Across kits, sensitivity (in words: the share of cases detected) ranged from 15.8% to 100.0% and specificity (in words: the share of non-cases correctly cleared) from 37.3% to 100.0%. The lipopolysaccharide (LPS)-specific immunochromatographic lateral-flow assay presented the highest performance, with sensitivity (in words: the share detected) of about 93–100% and specificity (in words: the share cleared) of about 99.19–100%. The authors recommend standardisation of accuracy-study conduct and reporting.

Watch out: A narrative systematic review without its own pooling, and its PubMed record lists no MeSH terms (the source line states this rather than inventing terms); the top-performing assay's figures come from individual studies inside the review, not from a pooled estimate.

Source: Clemente BM et al.. Evaluating immunochromatographic test kits for diagnosis of acute human leptospirosis: A systematic review. Heliyon. 2022. PMID: 36458294. Registry ID: none in the record. MeSH: none listed in the record

Card 3 — Synthesis

Study: Synthesis

Population: One meta-analysis (24 records, 49 entries) and one systematic review (22 articles), with overlapping primary literature.

Summary: Synthesis

Result: Rapid leptospirosis tests as a class pool at sensitivity (in words: the share of cases detected) 68% and specificity (in words: the share cleared) 93%, but the honest headline is the spread: individual kits range from near-useless to near-perfect, and the reference standard (MAT) is itself confined to specialised laboratories. MEASUREMENT ACCURACY — PARTIAL.

Watch out: PARTIAL because pooled sensitivity of 68% cannot exclude disease, heterogeneity is substantial, and most underlying studies were not at low risk of bias.

Source: Synthesis of the reports carded above; no additional source.