Biotin interference with cardiac troponin assays: biotin falsely lowered troponin T by a mean of 24–56% at supplement-relevant concentrations and interfered with the Roche Gen 5 troponin T assay above 31 ng/mL, but only 0.13–0.74% of real-world samples carried that much biotin and the modelled missed-infarction risk was 0.026% or less (documented N = 2,873 across the two prevalence cohorts — 850 emergency-department patients and 2,023 laboratory samples — plus two laboratory studies whose records state no patient count)

Shelf #11147 · Heart, vessels, blood pressure & clotting

POSSIBLE

Stamped: 2026-11-28.

Evidence published: 2018–2020.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients having cardiac troponin testing, does biotin taken as a supplement interfere with troponin immunoassays — falsely lowering the result and risking a missed myocardial infarction — and how much does the interference depend on which assay is used?

Balance of the evidence

POSSIBLE — Composition: 1 volunteer-and-spiking interference study (Frame 2019), 1 in-vitro dose–bias study (Schrapp 2018), and 1 prevalence and risk-modelling study in two cohorts (Mumma 2020: 850 emergency-department patients with suspected acute coronary syndrome and 2,023 random laboratory samples). The analytical finding is consistent across all three records: biotin falsely lowers results on the Roche Gen 5 troponin T assay at blood concentrations reachable on 10 mg/day supplements (interference threshold 31 ng/mL; mean negative bias 24% at 50 µg/L and 56% at 100 µg/L in spiked plasma), while the Gen 4 troponin T assay (threshold 315 ng/mL) and the high-sensitivity troponin I assay tested (threshold above 2,000 ng/mL) were not disturbed at supplement levels, and streptavidin pretreatment or neutralisation removed the interference. The clinical frequency is low: 0.13% of emergency-department presentation samples and 0.74% of laboratory samples exceeded the 20 ng/mL interference level, and the modelled risk of a false-negative myocardial infarction prediction was 0.026% or less. Scope note and boundary: this shelf covers cardiac troponin assays only. Biotin interference with thyroid function tests is a separate question, already owned by Shelf #5639, whose washout data are not carded here; no carded record here reports an actual missed infarction, and the one record that quantifies clinical risk (Mumma 2020) includes authors employed by the assay manufacturer — both points are stated on the cards.

MeSH: Adult · Biotin / blood · Female · Humans · Male · Middle Aged · Myocardial Infarction / blood · Troponin T / blood · Biotin / metabolism · Blood Chemical Analysis / methods · Limit of Detection · Myocardium / metabolism · Time Factors · Troponin T / metabolism · Acute Coronary Syndrome / diagnosis · Biomarkers / blood · Cohort Studies · Diagnostic Tests, Routine · False Negative Reactions · Immunoassay · Immunologic Tests · Myocardial Infarction / diagnosis · Risk Assessment

Cards (4)

Card 1 — Frame 2019: 10 mg/day biotin ingestion significantly interfered with the Roche Gen 5 troponin T assay (interference threshold 31 ng/mL) — the Gen 4 troponin T (threshold 315 ng/mL) and a high-sensitivity troponin I assay (threshold above 2,000 ng/mL) were unaffected at supplement levels

Study: Laboratory interference study with a volunteer supplementation arm (2019)

Population: Laboratory interference study in three parts: (1) plasma from healthy volunteers taking 10 mg of biotin daily, mixed with plasma of known elevated troponin; (2) plasmas with known elevated troponin spiked with reagent biotin to simulate supplementation; and (3) biotin-spiked specimens pretreated with streptavidin-agarose beads. Three assays were compared: Gen 5 cardiac troponin T, contemporary Gen 4 cardiac troponin T, and high-sensitivity cardiac troponin I. The record does not state the number of volunteers, so no volunteer count is printed on this shelf.

Summary: Frame 2019: 10 mg/day biotin ingestion significantly interfered with the Roche Gen 5 troponin T assay (interference threshold 31 ng/mL) — the Gen 4 troponin T (threshold 315 ng/mL) and a high-sensitivity troponin I assay (threshold above 2,000 ng/mL) were unaffected at supplement levels

Result: Daily biotin ingestion (10 mg) and the simulated-supplementation studies both produced significant interference in the Gen 5 cardiac troponin T assay; the Gen 4 troponin T and high-sensitivity troponin I assays were unaffected. The biotin interference threshold was 31 ng/mL for Gen 5 troponin T, 315 ng/mL for Gen 4 troponin T, and more than 2,000 ng/mL for high-sensitivity troponin I. Streptavidin pretreatment blocked the biotin interference in the troponin assays. The authors conclude that interference is possible at plasma concentrations achievable by over-the-counter supplementation and may lead to delayed or missed diagnosis of myocardial injury with the Gen 5 troponin T assay.

Watch out: The interference is assay-specific, not a class effect on all troponin testing: the troponin I assay tolerated more than 60 times the biotin concentration that disturbed Gen 5 troponin T. The record gives thresholds, not a count of patients actually misdiagnosed, and does not state how many volunteers took part.

Source: Frame IJ, Joshi PH, Mwangi C et al.. Susceptibility of Cardiac Troponin Assays to Biotin Interference. Am J Clin Pathol. 2019. PMID: 30715102. Registry ID: not stated in the record. MeSH: Adult · Biotin / blood · Female · Humans · Male · Middle Aged · Myocardial Infarction / blood · Troponin T / blood

Card 2 — Schrapp 2018: spiked biotin falsely lowered high-sensitivity troponin T by a mean of 24% at 50 µg/L, 56% at 100 µg/L, and 97–98% at 500–1,000 µg/L — the size of the fall did not depend on the starting troponin value

Study: In-vitro spiking and neutralisation study (2018)

Population: In-vitro spiking study: plasma samples at four troponin T concentrations (18, 59, 201 and 6,423 ng/L) were spiked with biotin to final concentrations of 50, 100, 500 and 1,000 µg/L (1 µg/L is the same concentration as 1 ng/mL) and measured on a Roche Cobas 8000 analyser; a biotin neutralisation protocol was also tested at the 59 ng/L troponin level with 1,000 µg/L biotin. No patients were enrolled; this is a laboratory study and contributes no patient count to the shelf total.

Summary: Schrapp 2018: spiked biotin falsely lowered high-sensitivity troponin T by a mean of 24% at 50 µg/L, 56% at 100 µg/L, and 97–98% at 500–1,000 µg/L — the size of the fall did not depend on the starting troponin value

Result: Mean negative bias was 24% of the expected troponin T value at a biotin concentration of 50 µg/L, 56% at 100 µg/L, 97% at 500 µg/L and 98% at 1,000 µg/L, and the interference was not dependent on the initial troponin value. The neutralisation protocol restored the 59 ng/L sample to a mean of 56.8 ng/L (coefficient of variation 1.31%) at every incubation time from 0 to 60 minutes — within a timeframe compatible with emergency testing — but the neutralisation process itself diluted the sample, losing 4.5% to 9.6% of the initial troponin value.

Watch out: Spiked concentrations of 500–1,000 µg/L are far above what supplement use usually produces in blood; the clinically relevant end of this curve is the 24–56% negative bias at 50–100 µg/L. The study used one manufacturer's high-sensitivity troponin T assay only.

Source: Schrapp A, Fraissinet F, Hervouet C et al.. Biotin and high-sensitivity cardiac troponin T assay. Biochem Med (Zagreb). 2018. PMID: 30429682. Registry ID: not stated in the record. MeSH: Biotin / metabolism · Blood Chemical Analysis / methods · Humans · Limit of Detection · Myocardium / metabolism · Time Factors · Troponin T / blood · Troponin T / metabolism

Card 3 — Mumma 2020: biotin above the 20 ng/mL interference level was rare in real testing — 1 of 797 emergency-department samples (0.13%) and 15 of 2,023 laboratory samples (0.74%) — and the modelled risk of a false-negative infarction prediction was 0.026% or less

Study: Prevalence cohorts with modelled misclassification risk (850 patients; 2,023 laboratory samples)

Population: Clinical risk assessment in two cohorts: (i) 797 samples at presentation and 646 samples at 3 hours from 850 US emergency-department patients with suspected acute coronary syndrome; and (ii) 2,023 random samples from a US laboratory network, with biotin distributions for higher values extrapolated by pharmacokinetic modelling. Biotin was measured with an Elecsys assay, and infarction misclassification risk from interference with the Troponin T Gen 5 assay was modelled at different cutoffs and timepoints.

Summary: Mumma 2020: biotin above the 20 ng/mL interference level was rare in real testing — 1 of 797 emergency-department samples (0.13%) and 15 of 2,023 laboratory samples (0.74%) — and the modelled risk of a false-negative infarction prediction was 0.026% or less

Result: In the emergency-department cohort, 1 of 797 presentation samples (0.13%) and 1 of 646 3-hour samples (0.15%) had biotin above 20.0 ng/mL; the 99th-percentile biotin value was 2.62 ng/mL at presentation and 2.38 ng/mL at 3 hours. Using conservative assumptions, the likelihood of a false-negative acute myocardial infarction prediction due to biotin interference was 0.026% for the presentation result at the 19 ng/L assay cutoff. In the laboratory cohort, 15 of 2,023 samples (0.74%) had biotin above 20.0 ng/mL and the 99th percentile was 16.6 ng/mL; modelled misclassification risk at the same cutoff was 0.025% at 0 hours, 0.0064% at 1 hour, 0.00048% at 3 hours and under 0.00001% at 6 hours. The authors conclude that biotin interference has minimal impact on the assay's clinical utility and that the likelihood of a false-negative infarction prediction is extremely low.

Watch out: The misclassification figures are modelled from biotin prevalence, not counted missed infarctions — no actual false-negative case is reported in this record. Several authors are employees of the assay's manufacturer (Roche Diagnostics), which is stated here because the paper's conclusion is reassuring about that manufacturer's assay.

Source: Mumma B, Diercks D, Twerenbold R et al.. Clinical risk assessment of biotin interference with a high-sensitivity cardiac troponin T assay. Clin Chem Lab Med. 2020. PMID: 32804676. Registry ID: not stated in the record. MeSH: Acute Coronary Syndrome / diagnosis · Biomarkers / blood · Biotin / blood · Cohort Studies · Diagnostic Tests, Routine · False Negative Reactions · Female · Humans · Immunoassay · Immunologic Tests · Male · Middle Aged · Myocardial Infarction / diagnosis · Risk Assessment · Troponin T / blood

Card 4 — Synthesis: the interference is analytically real and assay-specific — supplement-level biotin can falsely lower Roche Gen 5 troponin T — but in real-world cohorts so few samples carry enough biotin that the modelled missed-infarction risk is under 0.03%

Study: Synthesis

Population: The carded record: one volunteer-and-spiking interference study (Frame 2019), one in-vitro dose–bias study (Schrapp 2018), and one two-cohort prevalence and risk-modelling study (Mumma 2020; 850 emergency-department patients and 2,023 laboratory samples).

Summary: Synthesis: the interference is analytically real and assay-specific — supplement-level biotin can falsely lower Roche Gen 5 troponin T — but in real-world cohorts so few samples carry enough biotin that the modelled missed-infarction risk is under 0.03%

Result: All three records agree on direction and mechanism: on biotin–streptavidin sandwich assays, excess biotin pulls the reported troponin down — by a mean of 24% at 50 µg/L and 56% at 100 µg/L in spiked samples, with an interference threshold of 31 ng/mL for the Gen 5 troponin T assay against more than 2,000 ng/mL for the high-sensitivity troponin I assay tested. Streptavidin pretreatment or neutralisation removed the interference in both laboratory studies. The clinical frequency is the other half of the picture: only 0.13% of emergency-department presentation samples and 0.74% of random laboratory samples exceeded the 20 ng/mL interference level, and the modelled false-negative infarction risk was 0.026% or less. The stamp is POSSIBLE because the patient-level harm is modelled rather than observed, and the evidence is laboratory and observational, not trial evidence.

Watch out: This shelf is the troponin question only. Biotin interference with thyroid assays — a different set of analytes, platforms and washout times — is owned by Shelf #5639 and is not carded here; a biotin history matters for both, but the numbers do not transfer between the two shelves.

Source: Synthesis of the carded records above (PMIDs 30715102, 30429682, 32804676). Registry ID: not stated in the record for any carded study.