SourceMed · Shelf #11297 in the ask interface

Subcutaneous versus transvenous ICD (PRAETORIAN): the composite of device-related complications and inappropriate shocks occurred in 68 patients in each group over a median 49.1 months (48-month incidence 15.1% versus 15.7%; hazard ratio 0.99 relative) — noninferior, not superior (documented N = 849 randomised)

Shelf #11297 · Heart, vessels, blood pressure & clotting

POSSIBLE (noninferior to transvenous ICD)

Stamped: 2026-11-28.

Evidence published: 2020.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with an indication for an implantable cardioverter-defibrillator but no indication for pacing, is a subcutaneous ICD noninferior to a transvenous ICD for the composite of device-related complications and inappropriate shocks?

Balance of the evidence

POSSIBLE (noninferior to transvenous ICD) — Composition: one randomised noninferiority trial (PRAETORIAN, 849 patients with an ICD indication but no pacing indication), so the single-trial rule caps the stamp at POSSIBLE and makes the shelf. The primary composite of device-related complications and inappropriate shocks occurred in 68 patients in each group (48-month Kaplan-Meier incidence 15.1% subcutaneous versus 15.7% transvenous; hazard ratio 0.99 relative; P = 0.01 for noninferiority, P = 0.95 for superiority). Inside the composite the parts pulled in opposite directions: fewer device-related complications on the subcutaneous device (31 versus 44 patients) but more inappropriate shocks (41 versus 29). Death (83 versus 68 patients; hazard ratio 1.23 relative) and appropriate shocks (83 versus 57; hazard ratio 1.52 relative) are printed on the card. Whether an ICD itself prolongs life in non-ischaemic cardiomyopathy is homed at #5810; this shelf answers only the subcutaneous-versus-transvenous comparison.

MeSH: Aged · Arrhythmias, Cardiac / therapy · Cardiomyopathies / therapy · Death, Sudden, Cardiac / epidemiology · Death, Sudden, Cardiac / prevention & control · Defibrillators, Implantable / adverse effects · Electrodes, Implanted / adverse effects · Equipment Failure · Female · Follow-Up Studies · Heart Diseases / therapy · Humans · Incidence · Kaplan-Meier Estimate · Male · Middle Aged · Prosthesis Design

Cards (2)

Card 1 — PRAETORIAN 2020: subcutaneous ICD was noninferior to transvenous ICD on complications plus inappropriate shocks (68 events in each group) — fewer complications (31 vs 44) but more inappropriate shocks (41 vs 29)

Study: Randomised controlled trial, noninferiority (849 patients randomised)

Population: Randomised noninferiority trial: 849 patients with an indication for an ICD but no indication for pacing, assigned to a subcutaneous ICD (426 patients) or a transvenous ICD (423). The primary endpoint was the composite of device-related complications and inappropriate shocks, with a noninferiority margin of 1.45 for the upper boundary of the 95% confidence interval of the hazard ratio; a superiority analysis was prespecified if noninferiority was established. Median follow-up 49.1 months.

Summary: PRAETORIAN 2020: subcutaneous ICD was noninferior to transvenous ICD on complications plus inappropriate shocks (68 events in each group) — fewer complications (31 vs 44) but more inappropriate shocks (41 vs 29)

Result: A primary-endpoint event occurred in 68 patients in the subcutaneous ICD group and 68 in the transvenous ICD group (48-month Kaplan-Meier estimated cumulative incidence 15.1% and 15.7%; hazard ratio 0.99 relative; 95% CI 0.71 to 1.39; P = 0.01 for noninferiority; P = 0.95 for superiority). Device-related complications occurred in 31 versus 44 patients (hazard ratio 0.69 relative; 95% CI 0.44 to 1.09) and inappropriate shocks in 41 versus 29 (hazard ratio 1.43 relative; 95% CI 0.89 to 2.30). Death occurred in 83 versus 68 patients (hazard ratio 1.23 relative; 95% CI 0.89 to 1.70) and appropriate shocks in 83 versus 57 (hazard ratio 1.52 relative; 95% CI 1.08 to 2.12).

Watch out: Noninferior on a composite is not identical performance: the subcutaneous device trades lead complications for more inappropriate shocks, and the appropriate-shock difference (hazard ratio 1.52 relative) may reflect sensing and episode classification rather than more arrhythmia. Patients who need pacing, resynchronisation or antitachycardia pacing were excluded by design.

Source: Knops RE, Olde Nordkamp LRA, Delnoy PHM et al.. Subcutaneous or Transvenous Defibrillator Therapy. N Engl J Med. 2020. PMID: 32757521. Registry ID: NCT01296022. MeSH: Aged · Arrhythmias, Cardiac / therapy · Cardiomyopathies / therapy · Death, Sudden, Cardiac / epidemiology · Death, Sudden, Cardiac / prevention & control · Defibrillators, Implantable / adverse effects · Electrodes, Implanted / adverse effects · Equipment Failure · Female · Follow-Up Studies · Heart Diseases / therapy · Humans · Incidence · Kaplan-Meier Estimate · Male · Middle Aged · Prosthesis Design

Card 2 — Synthesis: one trial establishes noninferiority of the subcutaneous ICD on its composite in patients who do not need pacing — POSSIBLE (noninferior to transvenous ICD), Tier C

Study: Synthesis

Population: The carded record: PRAETORIAN only (849 randomised); total documented N = 849.

Summary: Synthesis: one trial establishes noninferiority of the subcutaneous ICD on its composite in patients who do not need pacing — POSSIBLE (noninferior to transvenous ICD), Tier C

Result: A single noninferiority trial caps at POSSIBLE. The honest summary is a trade-off, not a winner: fewer device-related complications, more inappropriate shocks, no shown mortality difference in either direction.

Watch out: Newer subcutaneous and extravascular systems and programming strategies postdate this trial; the comparison is between the two systems as implanted and programmed in PRAETORIAN.

Source: Synthesis of the carded record above (PMID 32757521). Registry ID: NCT01296022 (PRAETORIAN).