SourceMed · Shelf #11412 in the ask interface

Hydrus Microstent with cataract surgery (HORIZON): a ≥20% unmedicated pressure reduction at 24 months in 77.3% versus 57.8% with cataract surgery alone, mean pressure fall −7.6 versus −5.3 mmHg, and medications down to 0.3 versus 0.7 (documented N = 556 eyes randomised)

Shelf #11412 · Eyes, skin, nose & allergy

POSSIBLE (benefit)

Stamped: 2026-11-28.

Evidence published: 2019.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults with primary open-angle glaucoma having cataract surgery, does adding a Schlemm canal microstent — the Hydrus Microstent (Ivantis, Inc) — lower intraocular pressure and medication use more than cataract surgery alone?

Balance of the evidence

POSSIBLE (benefit) — Tier C: a single randomised trial (the HORIZON programme's 24-month primary report is carded). Adding the Hydrus Microstent — the exact tested product, Hydrus Microstent, Ivantis, Inc, Irvine, CA, as named in the record — to cataract surgery produced a ≥20% unmedicated pressure reduction in 77.3% versus 57.8% of eyes, an extra 2.3 mmHg of mean pressure fall, and fewer medications (0.3 versus 0.7) at 24 months, with no serious ocular adverse events related to the microstent and no significant safety differences. Boundary: Schlemm canal procedures other than this stent (GATT, canaloplasty, goniotomy and related techniques) are owned by #10125; this shelf is the Hydrus-with-cataract-surgery question only. Cataract surgery alone itself lowered pressure substantially (−5.3 mmHg) — the stent's contribution is the increment over that, not the whole fall.

MeSH: Aged · Aged, 80 and over · Antihypertensive Agents / administration & dosage · Aqueous Humor / physiology · Cataract / etiology* · Cataract / physiopathology · Female · Follow-Up Studies · Glaucoma, Open-Angle / physiopathology · Glaucoma, Open-Angle / surgery* · Gonioscopy · Humans · Intraocular Pressure / physiology* · Lens Implantation, Intraocular · Limbus Corneae / surgery* · Male · Phacoemulsification / methods* · Prospective Studies · Prosthesis Implantation · Single-Blind Method · Stents* · Tonometry, Ocular · Visual Acuity

Cards (2)

Card 1 — HORIZON 2019: Hydrus Microstent plus cataract surgery beat cataract surgery alone on pressure and medications at 24 months

Study: Randomised controlled trial (556 eyes randomised)

Population: Prospective multicentre single-masked randomised trial: 556 eyes of patients with primary open-angle glaucoma, visually significant cataract and washed-out diurnal intraocular pressure 22–34 mmHg, randomised 2:1 after uncomplicated phacoemulsification to a single Hydrus Microstent (Ivantis, Inc, Irvine, CA) in Schlemm canal (369 eyes) or no stent (187). The primary endpoint was the proportion with a ≥20% reduction in unmedicated diurnal pressure at 24 months.

Summary: HORIZON 2019: Hydrus Microstent plus cataract surgery beat cataract surgery alone on pressure and medications at 24 months

Result: At 24 months, unmedicated pressure fell by ≥20% in 77.3% of stent eyes versus 57.8% of controls (difference 19.5 percentage points; P < 0.001). Mean unmedicated pressure reduction was −7.6 mmHg versus −5.3 mmHg (difference −2.3 mmHg; P < 0.001). Mean medications fell from 1.7 at baseline to 0.3 in the stent group and 0.7 in controls (difference −0.4 medications; P < 0.001). No serious ocular adverse events were related to the microstent, and safety parameters did not differ significantly between groups.

Watch out: Single-masked, single trial programme — hence POSSIBLE and. The comparison is against cataract surgery alone in mild-to-moderate disease with washed-out pressures of 22–34 mmHg; eyes needing large pressure reductions or with prior incisional surgery are outside the record.

Source: Samuelson TW, Chang DF, Marquis R et al.; HORIZON Investigators. A Schlemm Canal Microstent for Intraocular Pressure Reduction in Primary Open-Angle Glaucoma and Cataract: The HORIZON Study. Ophthalmology. 2019. PMID: 29945799. Registry ID: NCT01539239. MeSH: Aged · Aged, 80 and over · Antihypertensive Agents / administration & dosage · Aqueous Humor / physiology · Cataract / etiology* · Cataract / physiopathology · Female · Follow-Up Studies · Glaucoma, Open-Angle / physiopathology · Glaucoma, Open-Angle / surgery* · Gonioscopy · Humans · Intraocular Pressure / physiology* · Lens Implantation, Intraocular · Limbus Corneae / surgery* · Male · Phacoemulsification / methods* · Prospective Studies · Prosthesis Implantation · Single-Blind Method · Stents* · Tonometry, Ocular · Visual Acuity

Card 2 — Synthesis: one trial shows a real but modest increment from adding a Hydrus stent to cataract surgery — about 2.3 mmHg and 0.4 fewer medications

Study: Synthesis

Population: The carded record: HORIZON 24-month report (PMID 29945799); total documented N = 556 eyes randomised.

Summary: Synthesis: one trial shows a real but modest increment from adding a Hydrus stent to cataract surgery — about 2.3 mmHg and 0.4 fewer medications

Result: The direction is consistent across the pressure, responder and medication endpoints, but a single programme cannot carry more than POSSIBLE under the two-trial gate.

Watch out: Longer HORIZON follow-up exists as separate reports and is not carded here; this shelf's numbers are the 24-month primary report only.

Source: Synthesis of the carded record above (PMID 29945799). Registry ID: NCT01539239.