Primary-prevention ICD after myocardial infarction: mortality 14.2% versus 19.8% in MADIT-II (hazard ratio 0.69 relative) and a 23% relative reduction in SCD-HeFT, whose results did not vary by ischaemic versus nonischaemic cause (documented N = 3,753 across two trials)

Shelf #11497 · Heart, vessels, blood pressure & clotting

LIKELY (benefit)

Stamped: 2026-11-28.

Evidence published: 2002–2005.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with prior myocardial infarction (ischaemic cardiomyopathy) and a reduced left ventricular ejection fraction, does a prophylactic implantable cardioverter-defibrillator reduce all-cause mortality compared with conventional medical therapy?

Balance of the evidence

LIKELY (benefit) — Composition: two independent randomised trials (total documented N = 3,753), concordant on all-cause mortality. MADIT-II (1,232 patients, all with prior infarction and ejection fraction 30% or less) found mortality of 14.2% with a defibrillator versus 19.8% with conventional therapy over an average 20 months (hazard ratio 0.69 relative). SCD-HeFT (2,521 patients with NYHA class II–III heart failure and ejection fraction 35% or less) found a 23% relative mortality reduction with a shock-only ICD (hazard ratio 0.77 relative; absolute decrease 7.2 percentage points at 5 years), with results that did not vary between ischaemic (52% of the trial) and nonischaemic causes; its amiodarone arm made no difference (hazard ratio 1.06 relative). Two-trial gate met; not. This is a narrowed build: the nonischaemic-cardiomyopathy ICD question is owned by shelf #5810 and is cross-pointed, not rebuilt; SCD-HeFT is carded here for its ischaemic-bearing result as its own record reports it. Neither abstract quantifies device harms, and the face says so.

MeSH: Adrenergic beta-Antagonists / therapeutic use · Aged · Amiodarone / therapeutic use · Angiotensin-Converting Enzyme Inhibitors / therapeutic use · Anti-Arrhythmia Agents / therapeutic use · Arrhythmias, Cardiac / etiology / prevention & control* · Combined Modality Therapy · Defibrillators, Implantable* / adverse effects · Female · Humans · Hypolipidemic Agents / therapeutic use · Male · Middle Aged · Myocardial Infarction / complications / mortality / physiopathology / therapy* · Proportional Hazards Models · Stroke Volume · Survival Analysis · Ventricular Dysfunction, Left / etiology / mortality / therapy* · Amiodarone / therapeutic use* · Anti-Arrhythmia Agents / therapeutic use* · Cause of Death · Cross-Over Studies · Death, Sudden, Cardiac / prevention & control* · Defibrillators, Implantable* · Heart Failure / drug therapy / mortality / therapy* · Risk

Cards (3)

Card 1 — MADIT-II 2002: a prophylactic defibrillator after myocardial infarction with ejection fraction 30% or less cut mortality from 19.8% to 14.2% over an average 20 months (hazard ratio 0.69 relative)

Study: Randomised controlled trial (1,232 patients randomised; MADIT-II)

Population: Randomised trial: 1,232 patients with a prior myocardial infarction and a left ventricular ejection fraction of 0.30 or less, assigned 3:2 to an implantable cardioverter-defibrillator (742) or conventional medical therapy (490), without electrophysiological risk stratification; the endpoint was death from any cause.

Summary: MADIT-II 2002: a prophylactic defibrillator after myocardial infarction with ejection fraction 30% or less cut mortality from 19.8% to 14.2% over an average 20 months (hazard ratio 0.69 relative)

Result: During an average follow-up of 20 months, mortality was 19.8% in the conventional-therapy group versus 14.2% in the defibrillator group (hazard ratio 0.69 relative; 95% CI 0.51 to 0.93; P = 0.016). The survival effect was similar across subgroups by age, sex, ejection fraction, NYHA class and QRS interval.

Watch out: A 2002-era background: beta-blocker and statin use were those of the time, follow-up averaged only 20 months, and the record's abstract does not report inappropriate shocks or device complications — the harm side of implantation is not quantified in this record.

Source: Moss AJ, Zareba W, Hall WJ et al.. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002. PMID: 11907286. Registry ID: not stated in the record. MeSH: Adrenergic beta-Antagonists / therapeutic use · Aged · Amiodarone / therapeutic use · Angiotensin-Converting Enzyme Inhibitors / therapeutic use · Anti-Arrhythmia Agents / therapeutic use · Arrhythmias, Cardiac / etiology / prevention & control* · Combined Modality Therapy · Defibrillators, Implantable* / adverse effects · Female · Humans · Hypolipidemic Agents / therapeutic use · Male · Middle Aged · Myocardial Infarction / complications / mortality / physiopathology / therapy* · Proportional Hazards Models · Stroke Volume · Survival Analysis · Ventricular Dysfunction, Left / etiology / mortality / therapy*

Card 2 — SCD-HeFT 2005: a shock-only ICD reduced mortality by 23% relative (absolute decrease 7.2 percentage points at 5 years) in NYHA class II–III heart failure; amiodarone did not (hazard ratio 1.06 relative)

Study: Randomised controlled trial (2,521 patients randomised; SCD-HeFT)

Population: Randomised double-blind (drug) trial: 2,521 patients with NYHA class II (70%) or III (30%) congestive heart failure and ejection fraction 35% or less — ischaemic in 52%, nonischaemic in 48%; median ejection fraction 25% — assigned to conventional therapy plus placebo (847), plus amiodarone (845), or plus a conservatively programmed, shock-only, single-lead ICD (829); median follow-up 45.5 months. The primary endpoint was death from any cause.

Summary: SCD-HeFT 2005: a shock-only ICD reduced mortality by 23% relative (absolute decrease 7.2 percentage points at 5 years) in NYHA class II–III heart failure; amiodarone did not (hazard ratio 1.06 relative)

Result: Deaths numbered 244 (29%) on placebo, 240 (28%) on amiodarone and 182 (22%) on ICD. Amiodarone did not change mortality (hazard ratio 1.06 relative; 97.5% CI 0.86 to 1.30; P = 0.53). ICD therapy reduced mortality by 23% relative (hazard ratio 0.77 relative; 97.5% CI 0.62 to 0.96; P = 0.007), an absolute decrease of 7.2 percentage points at 5 years. Results did not vary by ischaemic versus nonischaemic cause, though they did vary by NYHA class.

Watch out: SCD-HeFT is a mixed-aetiology trial carded here because its own analysis found no variation by cause; the purely nonischaemic question (DEFINITE, DANISH) is homed at shelf #5810 and is not re-carded. The NYHA-class variation is flagged in the abstract without subgroup rates, and none are invented here.

Source: Bardy GH, Lee KL, Mark DB et al.. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005. PMID: 15659722. Registry ID: not stated in the record. MeSH: Aged · Amiodarone / therapeutic use* · Anti-Arrhythmia Agents / therapeutic use* · Cause of Death · Cross-Over Studies · Death, Sudden, Cardiac / prevention & control* · Defibrillators, Implantable* · Female · Heart Failure / drug therapy / mortality / therapy* · Humans · Male · Middle Aged · Risk · Stroke Volume · Survival Analysis

Card 3 — Synthesis: two concordant mortality trials establish ICD benefit for primary prevention after myocardial infarction with reduced ejection fraction — LIKELY (benefit); the amiodarone arm of SCD-HeFT was null

Study: Synthesis

Population: The carded record: MADIT-II (1,232 randomised, all post-infarction) and SCD-HeFT (2,521 randomised, 52% ischaemic, no effect variation by cause); total documented N = 3,753.

Summary: Synthesis: two concordant mortality trials establish ICD benefit for primary prevention after myocardial infarction with reduced ejection fraction — LIKELY (benefit); the amiodarone arm of SCD-HeFT was null

Result: Both trials show a defibrillator prolonging life in this population (hazard ratios 0.69 and 0.77 relative), meeting the two-trial gate for LIKELY (benefit). This shelf is the ischaemic / post-infarction slice: the nonischaemic-cardiomyopathy ICD question is homed at shelf #5810, ICD programming at shelf #2293, and the wearable defibrillator after infarction at shelf #11294. Amiodarone for mortality prevention in heart failure was tested in SCD-HeFT and made no difference.

Watch out: Neither abstract quantifies inappropriate shocks, infections or lead complications; a patient weighing implantation gets the mortality side only from this record, and that gap is stated rather than filled from outside sources.

Source: Synthesis of the carded records above (PMIDs 11907286, 15659722). Registry ID: not stated in the records (both trials predate routine registration reporting in abstracts).