SourceMed · Shelf #11502 in the ask interface
Computer-aided detection at colonoscopy: adenoma detection rate 54.8% versus 40.4% in the GI Genius randomised trial (relative risk 1.30 relative), and 36.6% versus 25.2% pooled across five randomised trials (relative risk 1.44 relative) — a detection benefit concentrated in small adenomas (documented N = 5,039 across the trial and the pooled trials)
Shelf #11502 · Liver & gut
LIKELY (benefit — adenoma detection endpoint only)
Stamped: 2026-11-28.
Evidence published: 2020–2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults undergoing colonoscopy, does real-time computer-aided detection (CADe) — an artificial-intelligence system that flags suspected polyps on the endoscopy display — increase adenoma detection compared with standard high-definition colonoscopy?
Balance of the evidence
LIKELY (benefit — adenoma detection endpoint only) — Composition: one multicentre randomised trial of a named commercial system plus a meta-analysis of five randomised trials, concordant. In the Repici 2020 trial (685 patients, three Italian centres), the tested product was the GI-Genius system (Medtronic) — named here as published because detection performance is a property of a specific device and software, not of a drug class. Adenoma detection rate was 54.8% with CADe versus 40.4% without (relative risk 1.30 relative, 95% CI 1.14 to 1.45), an absolute difference of about 14 percentage points, with no increase in withdrawal time and no increase in resection of non-neoplastic lesions. The pooled estimate across five randomised trials (4,354 patients) is the same direction and larger: 36.6% versus 25.2% (relative risk 1.44 relative, 95% CI 1.27 to 1.62). The stamp says detection only: no carded trial measures colorectal cancer incidence or death, and the pooled gain is concentrated in adenomas of 5 mm or smaller, whose malignant potential is lowest; the pooled trend for advanced adenomas was not significant (relative risk 1.35 relative, 95% CI 0.74 to 2.47). Total documented N = 5,039: the Repici trial (685) was published after the meta-analysis search closed in March 2020, so the two are counted as independent. Neighbouring detection shelves are cross-pointed, not rebuilt: second forward-view or retroflexed examination of the right colon (#9677) and linked colour imaging (#10113).
MeSH: Adenoma / diagnosis* · Adenoma / epidemiology · Adenoma / pathology · Aged · Biopsy / statistics & numerical data · Colon / diagnostic imaging · Colon / pathology · Colonoscopy / instrumentation · Colonoscopy / methods* · Colonoscopy / statistics & numerical data · Colorectal Neoplasms / diagnosis* · Colorectal Neoplasms / epidemiology · Colorectal Neoplasms / pathology · Deep Learning* · Female · Humans · Image Interpretation, Computer-Assisted / instrumentation · Image Interpretation, Computer-Assisted / methods* · Image Interpretation, Computer-Assisted / statistics & numerical data · Male · Mass Screening / instrumentation · Mass Screening / methods* · Mass Screening / statistics & numerical data · Middle Aged · Time Factors · Adenoma* / diagnosis · Artificial Intelligence · Colonic Polyps* / diagnosis · Colonoscopy · Colorectal Neoplasms* / diagnosis
Cards (3)
Card 1 — Repici 2020: GI Genius computer-aided detection raised the adenoma detection rate to 54.8% versus 40.4% — without longer withdrawal time or more non-neoplastic resections
Study: Randomised controlled trial, multicentre (685 patients)
Population: Multicentre randomised trial at three centres in Italy: 685 patients undergoing screening colonoscopy, post-polypectomy surveillance, or colonoscopy for a positive faecal immunochemical test or symptoms, assigned (1:1) to high-definition colonoscopy with the CADe system or standard high-definition colonoscopy (control). The CADe system was the artificial-intelligence device GI-Genius (Medtronic), which superimposes a green box over suspected lesions in real time. Minimum withdrawal time was 6 minutes; histopathology was the reference standard. The primary outcome was adenoma detection rate (at least one histologically proven adenoma or carcinoma).
Summary: Repici 2020: GI Genius computer-aided detection raised the adenoma detection rate to 54.8% versus 40.4% — without longer withdrawal time or more non-neoplastic resections
Result: Adenoma detection rate was significantly higher with CADe (54.8%) than in controls (40.4%) (relative risk 1.30 relative, 95% CI 1.14 to 1.45), an absolute difference of about 14 percentage points. Adenomas per colonoscopy were 1.07 versus 0.71 (incidence rate ratio 1.46 relative, 95% CI 1.15 to 1.86). Detection advantages were significant for adenomas 5 mm or smaller (33.7% versus 26.5% of patients) and 6 to 9 mm (10.6% versus 5.8%). Withdrawal time did not differ (417 versus 435 seconds; P = 0.1), nor did resection of non-neoplastic lesions (26.0% versus 28.7%).
Watch out: Endoscopists cannot be blinded to a system that draws boxes on their screen, so enthusiasm and inspection effort are unblinded co-interventions; and the trial's baseline detection rate (40.4% in controls) was already high, so the absolute gain may shrink where baseline detection is low — or grow: the trial cannot say.
Source: Repici A, Badalamenti M, Maselli R et al.. Efficacy of Real-Time Computer-Aided Detection of Colorectal Neoplasia in a Randomized Trial. Gastroenterology. 2020. PMID: 32371116. Registry ID: NCT04079478 (stated in the record as ClinicalTrials.gov no. 04079478). MeSH: Adenoma / diagnosis* · Adenoma / epidemiology · Adenoma / pathology · Aged · Biopsy / statistics & numerical data · Colon / diagnostic imaging · Colon / pathology · Colonoscopy / instrumentation · Colonoscopy / methods* · Colonoscopy / statistics & numerical data · Colorectal Neoplasms / diagnosis* · Colorectal Neoplasms / epidemiology · Colorectal Neoplasms / pathology · Deep Learning* · Female · Humans · Image Interpretation, Computer-Assisted / instrumentation · Image Interpretation, Computer-Assisted / methods* · Image Interpretation, Computer-Assisted / statistics & numerical data · Male · Mass Screening / instrumentation · Mass Screening / methods* · Mass Screening / statistics & numerical data · Middle Aged · Time Factors
Card 2 — Hassan 2021 meta-analysis: across five randomised trials (4,354 patients), CADe raised the adenoma detection rate from 25.2% to 36.6% (relative risk 1.44 relative) — driven by small adenomas; advanced adenomas did not significantly increase
Study: Systematic review and meta-analysis of 5 randomised controlled trials (4,354 patients)
Population: Systematic review and meta-analysis of randomised controlled trials (searched to March 2020) of real-time computer-aided polyp detection during colonoscopy: five randomised trials, 4,354 patients. Primary outcome was pooled adenoma detection rate; secondary outcomes included adenomas per colonoscopy by size, morphology and location, polyp detection and sessile serrated lesions per colonoscopy. The reviewers graded the randomised-trial evidence as moderate quality.
Summary: Hassan 2021 meta-analysis: across five randomised trials (4,354 patients), CADe raised the adenoma detection rate from 25.2% to 36.6% (relative risk 1.44 relative) — driven by small adenomas; advanced adenomas did not significantly increase
Result: Pooled adenoma detection rate was 36.6% (791 of 2,163) with CADe versus 25.2% (558 of 2,191) in controls (relative risk 1.44 relative, 95% CI 1.27 to 1.62; P < 0.01). Adenomas per colonoscopy were 0.58 versus 0.36 (relative risk 1.70 relative, 95% CI 1.53 to 1.89). Gains were significant for adenomas 5 mm or smaller, 6 to 9 mm and 10 mm or larger, and for flat and polypoid, proximal and distal lesions; sessile serrated lesions per colonoscopy also rose (relative risk 1.52 relative, 95% CI 1.14 to 2.02). For advanced adenomas the trend was not significant (relative risk 1.35 relative, 95% CI 0.74 to 2.47; P = 0.33).
Watch out: The pooled trials used several different CADe systems and mostly enrolled in single countries with low-to-moderate baseline detection rates; a class-level pooled estimate can hide a weak individual system, which is why the named-system trial is carded separately above.
Source: Hassan C, Spadaccini M, Iannone A et al.. Performance of artificial intelligence in colonoscopy for adenoma and polyp detection: a systematic review and meta-analysis. Gastrointest Endosc. 2021. PMID: 32598963. Registry ID: not stated in the record. MeSH: Adenoma* / diagnosis · Artificial Intelligence · Colonic Polyps* / diagnosis · Colonoscopy · Colorectal Neoplasms* / diagnosis · Humans
Card 3 — Synthesis: CADe finds more adenomas — about 11 to 14 more patients with at least one adenoma per 100 colonoscopies — but the extra yield is mostly diminutive adenomas, and no trial yet shows fewer cancers
Study: Synthesis
Population: The carded record: Repici 2020 (685 randomised) and the Hassan 2021 meta-analysis of five randomised trials (4,354 patients); total documented N = 5,039, counted as independent because the trial was published after the meta-analysis search closed (March 2020).
Summary: Synthesis: CADe finds more adenomas — about 11 to 14 more patients with at least one adenoma per 100 colonoscopies — but the extra yield is mostly diminutive adenomas, and no trial yet shows fewer cancers
Result: Two independent randomised evidence sets agree on a substantial relative detection gain, meeting the two-trial gate for LIKELY — restricted to the detection endpoint, as the stamp states. Whether finding more diminutive adenomas reduces later colorectal cancer is biologically plausible and unproven in trials; surveillance-interval consequences (more patients entering surveillance) are a real downstream cost the detection trials do not weigh.
Watch out: Optical diagnosis (characterising a found polyp without histology) is a different technology question, homed on shelf #10114; CADe as carded here detects lesions, it does not characterise them.
Source: Synthesis of the carded records above (PMIDs 32371116, 32598963). Registry ID: NCT04079478 (Repici trial).