Closed-loop control in type 1 diabetes (iDCL): time in the 70–180 mg/dL (3.9–10.0 mmol/L) range rose to 71% versus 59% on a sensor-augmented pump (adjusted difference 11 percentage points; P < 0.001), with HbA1c 0.33 points lower (documented N = 168)
Shelf #11584 · Diabetes, glucose & metabolism
POSSIBLE (benefit, surrogate endpoint only)
Stamped: 2026-11-28.
Evidence published: 2019.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with type 1 diabetes aged 14 to 71 years, does a closed-loop (automated insulin delivery) system increase time in the target glucose range versus a sensor-augmented pump over 6 months?
Balance of the evidence
POSSIBLE (benefit, surrogate endpoint only) — Composition: one randomised trial (total documented N = 168). Over 6 months, time in the target range of 70 to 180 mg/dL (3.9 to 10.0 mmol/L) rose from 61% to 71% with the closed-loop system and stayed at 59% with a sensor-augmented pump (adjusted difference 11 percentage points; P < 0.001); HbA1c was 0.33 percentage points lower (P = 0.001) and time below 70 mg/dL (3.9 mmol/L) was 0.88 percentage points lower (P < 0.001). The endpoints are glycaemic surrogates, not clinical outcomes, and the stamp is qualified accordingly; a single trial caps it at POSSIBLE with tierC true. One episode of diabetic ketoacidosis occurred in the closed-loop group; no serious hypoglycaemic events occurred in either group. This is a narrowed build within a crowded field: the insulin-only bionic pancreas is homed at shelf #5529, open-source automated delivery at #1764, dual-hormone closed loop at #10743, closed loop in pregnancy at #2633 and automated delivery in type 2 diabetes at #6054. The record describes the intervention as a closed-loop system under the iDCL protocol and does not name a commercial product.
MeSH: Adolescent · Adult · Aged · Blood Glucose / analysis · Diabetes Mellitus, Type 1 / blood · Diabetes Mellitus, Type 1 / drug therapy · Equipment Design · Female · Glycated Hemoglobin / analysis · Humans · Hypoglycemic Agents / administration & dosage · Hypoglycemic Agents / adverse effects · Insulin / administration & dosage · Insulin / adverse effects · Insulin Infusion Systems / adverse effects* · Male · Middle Aged · Pancreas, Artificial / adverse effects* · Young Adult
Cards (2)
Card 1 — iDCL 2019: closed-loop control raised time in range to 71% versus 59% over 6 months (adjusted difference 11 points; P < 0.001) and lowered HbA1c by 0.33 points
Study: Randomised controlled trial (168 patients randomised; iDCL)
Population: Six-month randomised multicentre trial: 168 patients with type 1 diabetes aged 14 to 71 years (baseline HbA1c 5.4% to 10.6%) assigned 2:1 to a closed-loop system (112) or a sensor-augmented pump (56). The primary outcome was the percentage of time glucose was within 70 to 180 mg/dL (3.9 to 10.0 mmol/L) on continuous glucose monitoring. All 168 patients completed the trial.
Summary: iDCL 2019: closed-loop control raised time in range to 71% versus 59% over 6 months (adjusted difference 11 points; P < 0.001) and lowered HbA1c by 0.33 points
Result: Time in range increased in the closed-loop group from 61 ± 17% at baseline to 71 ± 12% over 6 months and remained unchanged at 59 ± 14% in the control group (mean adjusted difference 11 percentage points; 95% CI 9 to 14; P < 0.001). All main secondary outcomes (time above 180 mg/dL [10.0 mmol/L], mean glucose, HbA1c, time below 70 mg/dL [3.9 mmol/L] or below 54 mg/dL [3.0 mmol/L]) met the prespecified hierarchical criterion for significance favouring closed loop; the difference in time below 70 mg/dL was −0.88 percentage points (95% CI −1.19 to −0.57; P < 0.001) and in HbA1c −0.33 percentage points (95% CI −0.53 to −0.13; P = 0.001). The system was in closed-loop mode a median 90% of the time. No serious hypoglycaemic events occurred in either group; one episode of diabetic ketoacidosis occurred in the closed-loop group.
Watch out: A surrogate-endpoint trial: no carded outcome on severe hypoglycaemia (none occurred), complications or quality of life; participants were experienced pump/CGM users at trial centres, and the record does not name the commercial system used.
Source: Brown SA, Kovatchev BP, Raghinaru D et al.. Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes. The New England journal of medicine. 2019. PMID: 31618560. Registry ID: NCT03563313 (as stated in the record). MeSH: Adolescent · Adult · Aged · Blood Glucose / analysis · Diabetes Mellitus, Type 1 / blood · Diabetes Mellitus, Type 1 / drug therapy · Equipment Design · Female · Glycated Hemoglobin / analysis · Humans · Hypoglycemic Agents / administration & dosage · Hypoglycemic Agents / adverse effects · Insulin / administration & dosage · Insulin / adverse effects · Insulin Infusion Systems / adverse effects* · Male · Middle Aged · Pancreas, Artificial / adverse effects* · Young Adult
Estimate type: adjusted estimate, as stated in the card.
Card 2 — Synthesis: one 6-month trial shows closed-loop control improving time in range and HbA1c versus a sensor-augmented pump — POSSIBLE (benefit, surrogate endpoint only)
Study: Synthesis
Population: The carded record: the iDCL trial only (168 randomised); total documented N = 168.
Summary: Synthesis: one 6-month trial shows closed-loop control improving time in range and HbA1c versus a sensor-augmented pump — POSSIBLE (benefit, surrogate endpoint only)
Result: Time in range was 11 percentage points higher and HbA1c 0.33 points lower with closed-loop control, with less time below range. A single trial on surrogate endpoints caps the stamp at POSSIBLE, qualified as surrogate-endpoint only. Adjacent automated-delivery shelves: bionic pancreas #5529, open-source systems #1764, dual-hormone closed loop #10743, pregnancy closed loop #2633, type 2 diabetes automated delivery #6054.
Watch out: This shelf does not compare commercial systems against each other and does not cover children under 14, pregnancy, or type 2 diabetes — those questions are homed at the shelves named in the synthesis.
Source: Synthesis of the carded records above (PMID 31618560). Registry ID: NCT03563313.