Electrode position for cardioverting atrial fibrillation — SPLIT by shock protocol: the largest trial found anterior-lateral placement clearly better on the first shock (54% versus 33%) and after the full sequence (93% versus 85%), while a five-trial pooled analysis found position made no significant difference and shock energy of 200 J or more mattered more (documented N = 1,078 in the pooled trials)

Shelf #11791 · Heart, vessels, blood pressure & clotting

SPLIT

Stamped: 2026-11-28.

Evidence published: 2021–2023.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In adults with atrial fibrillation undergoing elective electrical cardioversion with a biphasic shock, does anterior-lateral electrode (pad) placement restore sinus rhythm more reliably than anterior-posterior placement?

Balance of the evidence

SPLIT — Composition: one large multicentre randomised trial (468 patients) and a systematic review pooling five randomised trials (1,078 patients, including that trial; the pooled figure is counted once and the trial is not added again). The trial found anterior-lateral placement superior: sinus rhythm after the first shock in 54% versus 33% (risk difference 22 percentage points) and after up to four escalating shocks in 93% versus 85%. The pooled analysis found no significant association between position and success after a first low-energy shock (pooled relative risk 1.28 relative, with considerable heterogeneity) or after a high-energy final shock (relative risk 1.05 relative), and found instead that shocks under 200 J succeeded less often than shocks of 200 J or more regardless of position. The honest summary is split: position mattered in the largest trial's escalating-energy protocol; across protocols, delivered energy looks like the stronger lever. The stamp is SPLIT and the pack Tier C, because the synthesis and its largest constituent trial are one overlapping body of evidence, not two independent ones. Drug cardioversion questions are owned elsewhere (vernakalant #6655, antazoline #11789); this shelf answers only the pad-position question.

MeSH: Atrial Fibrillation / therapy* · Electric Countershock / methods* · Electrodes / standards* · Female · Humans · Male · Randomized Controlled Trials as Topic · Treatment Outcome · Atrial Fibrillation* / diagnosis · Atrial Fibrillation* / therapy · Electric Countershock / adverse effects · Electrodes

Cards (3)

Card 1 — Schmidt 2021: anterior-lateral pads beat anterior-posterior on the first shock (54% versus 33%) and after the full escalating sequence (93% versus 85%)

Study: Randomised controlled trial, multicentre (468 patients randomised)

Population: Multicentre investigator-initiated open-label randomised trial: 468 patients with atrial fibrillation scheduled for elective cardioversion assigned to anterior-lateral or anterior-posterior electrode positioning with biphasic shocks. The primary outcome was sinus rhythm after the first shock; the secondary outcome was sinus rhythm after up to four shocks escalating to maximum energy. Safety outcomes were arrhythmia during or after cardioversion, skin redness and patient-reported pain.

Summary: Schmidt 2021: anterior-lateral pads beat anterior-posterior on the first shock (54% versus 33%) and after the full escalating sequence (93% versus 85%)

Result: The primary outcome occurred in 126 patients (54%) with anterior-lateral positioning versus 77 (33%) with anterior-posterior positioning — a risk difference of 22 percentage points (95% CI 13–30; P < 0.001). After the final shock, 216 patients (93%) versus 200 (85%) were in sinus rhythm (risk difference 7 points; 95% CI 2–12). No safety outcome differed significantly between groups.

Watch out: The trial was open-label and used self-adhesive pads with an escalating-energy protocol; its first-shock advantage may partly reflect the energy sequence rather than position alone — exactly the ambiguity the pooled analysis below raises.

Source: Schmidt AS, Lauridsen KG, Møller DS, Christensen PD et al. Anterior-Lateral Versus Anterior-Posterior Electrode Position for Cardioverting Atrial Fibrillation. Circulation. 2021. PMID: 34814700. Registry ID: NCT03817372. MeSH: Atrial Fibrillation / therapy* · Electric Countershock / methods* · Electrodes / standards* · Female · Humans · Male · Randomized Controlled Trials as Topic · Treatment Outcome

Card 2 — Pooled analysis 2023: across five biphasic trials, pad position was not significantly associated with success after the first or final shock — energy under 200 J was the weaker choice in either position

Study: Systematic review and meta-analysis (5 randomised trials, 1,078 patients; counted once as the pack total — the carded Schmidt trial sits inside it)

Population: Systematic review and dose-response meta-analysis of randomised trials of elective cardioversion of atrial fibrillation using biphasic shocks and self-adhesive patches, published 2000–2023: five trials, 1,078 patients (including the Schmidt trial carded above). Pooled relative risks compared anterolateral with anteroposterior placement after the first and final shocks, and modelled success against shock energy.

Summary: Pooled analysis 2023: across five biphasic trials, pad position was not significantly associated with success after the first or final shock — energy under 200 J was the weaker choice in either position

Result: After the first low-energy shock, position was not significantly associated with success (pooled relative risk 1.28 relative; 95% CI 0.93–1.76, with considerable heterogeneity). After a high-energy final shock there was no evidence of an association (relative risk 1.05 relative; 95% CI 0.97–1.14). Regardless of position, cardioversion success was significantly less likely with shock energy under 200 J than with 200 J. The reviewers concluded that positioning can be based on convenience and that energy selection matters.

Watch out: The pooled first-shock estimate is heterogeneous and its confidence interval (0.93–1.76) does not exclude the benefit Schmidt found; 'no significant association' in the pool is not proof that position never matters.

Source: Vinter N, Holst-Hansen MZB, Johnsen SP, Lip GYH et al. Electrical energy by electrode placement for cardioversion of atrial fibrillation: a systematic review and meta-analysis. Open Heart. 2023. PMID: 37945283. Registry ID: not stated in the record. MeSH: Humans · Atrial Fibrillation* / diagnosis · Atrial Fibrillation* / therapy · Electric Countershock / adverse effects · Electrodes

Card 3 — Synthesis: if the first shock counts, anterior-lateral placement won the biggest trial; if the full protocol counts, position fades and delivering at least 200 J is the consistent lesson — SPLIT

Study: Synthesis

Population: The carded record: the Schmidt trial (468 randomised) inside a five-trial pooled analysis (1,078 patients). Total documented N = 1,078, counted once from the synthesis.

Summary: Synthesis: if the first shock counts, anterior-lateral placement won the biggest trial; if the full protocol counts, position fades and delivering at least 200 J is the consistent lesson — SPLIT

Result: The two carded sources answer slightly different questions — one protocol's first shock versus pooled success across protocols — and disagree in emphasis, so the stamp is SPLIT and the pack Tier C (one overlapping body of evidence). The practical reading printed on the face: choose anterior-lateral placement when convenient, and do not start low on energy.

Watch out: These were elective cardioversions with biphasic waveforms and adhesive pads; the findings do not automatically transfer to emergency cardioversion, monophasic devices, paddles, or atrial flutter.

Source: Synthesis of the carded records above (PMIDs 34814700, 37945283). Registry ID: NCT03817372 (Schmidt trial); the review states none.