MRI with a non-MRI-conditional pacemaker or defibrillator: in two prospective registries totalling 3,009 documented patients and cases, no device or lead failure and no MRI-attributable death occurred when a prespecified protocol was followed; device resets to backup mode occurred in 0.4% of examinations in one registry

Shelf #11794 · Heart, vessels, blood pressure & clotting

POSSIBLE (benefit)

Stamped: 2026-11-28.

Evidence published: 2017.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with a pacemaker or implantable cardioverter-defibrillator that is not MRI-conditional ('legacy' device), can a clinically indicated MRI at 1.5 tesla be performed safely under a standardised screening and reprogramming protocol?

Balance of the evidence

POSSIBLE (benefit) — Composition: two independent prospective registries of MRI at 1.5 tesla in patients with non-MRI-conditional ('legacy') pacemakers and defibrillators, both using prespecified screening and reprogramming protocols. MagnaSafe (1,500 cases: 1,000 pacemaker, 500 defibrillator) recorded no deaths, lead failures, losses of capture or ventricular arrhythmias during scanning; one defibrillator generator that had not been programmed per protocol could not be interrogated afterwards and was replaced. The Johns Hopkins registry (1,509 patients; 2,103 thoracic and nonthoracic examinations) likewise reported no long-term clinically significant adverse events; in nine examinations (0.4%) the device reset to a backup mode — transient in eight — and the parameter changes seen at follow-up were not clinically significant and required no revision. 'Benefit' here means access to a needed scan without device harm, under protocol — not a therapeutic effect. The stamp is POSSIBLE and Tier C: both sources are observational registries without a randomised comparator, so this is a consistency-of-no-harm finding, and it applies only to the protocols and 1.5-tesla field strength studied. Device-therapy questions themselves are owned elsewhere (subcutaneous ICD #11297, conduction-system pacing #9500, remote monitoring #2294); this shelf answers only the MRI-safety question.

MeSH: Adult · Aged · Aged, 80 and over · Atrial Fibrillation / etiology · Atrial Flutter / etiology · Contraindications · Defibrillators, Implantable* · Equipment Failure · Female · Humans · Magnetic Resonance Imaging / adverse effects* · Male · Middle Aged · Pacemaker, Artificial* · Prospective Studies · Registries · Electric Power Supplies · Equipment Safety* · Follow-Up Studies

Cards (3)

Card 1 — MagnaSafe 2017: 1,500 MRI cases with non-conditional pacemakers or defibrillators — no deaths, lead failures, losses of capture or ventricular arrhythmias during scanning

Study: Prospective registry (1,500 cases: 1,000 pacemaker, 500 defibrillator)

Population: Prospective registry (MagnaSafe): patients with a pacemaker (1,000 cases) or implantable cardioverter-defibrillator (500 cases) that was not MRI-conditional, referred for clinically indicated nonthoracic MRI at 1.5 tesla. Devices were interrogated before and after MRI under a standardised protocol and reprogrammed before scanning. Primary endpoints were death, generator or lead failure, induced arrhythmia, loss of capture or electrical reset during scanning; secondary endpoints were changes in device settings.

Summary: MagnaSafe 2017: 1,500 MRI cases with non-conditional pacemakers or defibrillators — no deaths, lead failures, losses of capture or ventricular arrhythmias during scanning

Result: No deaths, lead failures, losses of capture or ventricular arrhythmias occurred during MRI. One defibrillator generator could not be interrogated after MRI and required immediate replacement — it had not been appropriately programmed per protocol before scanning. Six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset occurred. Changes in lead impedance, pacing threshold, battery voltage and sensing amplitudes exceeded prespecified thresholds in a small number of cases, and repeat MRI was not associated with more adverse events.

Watch out: MagnaSafe covered nonthoracic MRI only, at 1.5 tesla, in screened patients whose devices were reprogrammed per protocol — the single generator failure happened precisely when the protocol was not followed, which is part of the lesson, not an exception to it.

Source: Russo RJ, Costa HS, Silva PD, Anderson JL et al. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. N Engl J Med. 2017. PMID: 28225684. Registry ID: NCT00907361. MeSH: Adult · Aged · Aged, 80 and over · Atrial Fibrillation / etiology · Atrial Flutter / etiology · Contraindications · Defibrillators, Implantable* · Equipment Failure · Female · Humans · Magnetic Resonance Imaging / adverse effects* · Male · Middle Aged · Pacemaker, Artificial* · Prospective Studies · Registries

Card 2 — Nazarian 2017: 1,509 patients with legacy devices underwent 2,103 thoracic and nonthoracic MRIs — no long-term clinically significant adverse events; device reset in 0.4% of examinations

Study: Prospective observational study (1,509 patients; 2,103 MRI examinations)

Population: Prospective nonrandomised study: 1,509 patients with a non-MRI-conditional ('legacy') pacemaker (58%) or defibrillator (42%) underwent 2,103 clinically necessary thoracic and nonthoracic MRI examinations at 1.5 tesla. Pacing mode was set to asynchronous in pacing-dependent patients and demand mode in others, and tachyarrhythmia functions were disabled. Outcomes were adverse events and changes in lead and generator parameters, including at long-term follow-up (available for 63% of patients).

Summary: Nazarian 2017: 1,509 patients with legacy devices underwent 2,103 thoracic and nonthoracic MRIs — no long-term clinically significant adverse events; device reset in 0.4% of examinations

Result: No long-term clinically significant adverse events were reported. In nine examinations (0.4%; 95% CI 0.2–0.7) the device reset to a backup mode — transient in eight; in one case a pacemaker with under a month of battery life left reset and could not be reprogrammed, and was replaced. The commonest parameter change immediately after MRI was a decrease in P-wave amplitude (1% of patients); at long-term follow-up, decreases in P-wave amplitude and increases in capture thresholds (3–4%) were the commonest changes, none clinically significant and none requiring device revision or reprogramming.

Watch out: Long-term follow-up was available for only 63% of patients, and patients with abandoned, fractured or epicardial leads and very recent implants are not the population these registries establish safety for; scanning outside a protocol with device expertise available is not what was studied.

Source: Nazarian S, Hansford R, Rahsepar AA, Weltin V et al. Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices. N Engl J Med. 2017. PMID: 29281579. Registry ID: NCT01130896. MeSH: Aged · Defibrillators, Implantable* · Electric Power Supplies · Equipment Failure · Equipment Safety* · Female · Follow-Up Studies · Humans · Magnetic Resonance Imaging / adverse effects* · Male · Middle Aged · Pacemaker, Artificial* · Prospective Studies

Card 3 — Synthesis: two independent registries, 3,009 documented patients and cases, zero device or lead failures under protocol — MRI access without device harm is POSSIBLE (benefit), observational Tier C

Study: Synthesis

Population: The carded record: MagnaSafe (1,500 cases) and the Nazarian registry (1,509 patients). Total documented N = 3,009 — the registries count differently (cases versus patients), which is stated on the face; the cohorts are independent and are added with that caveat.

Summary: Synthesis: two independent registries, 3,009 documented patients and cases, zero device or lead failures under protocol — MRI access without device harm is POSSIBLE (benefit), observational Tier C

Result: Concordant no-harm findings in two large prospective registries support performing protocol-guided 1.5-tesla MRI in legacy-device patients, but with no randomised or untreated comparator the stamp is POSSIBLE and the pack. The resets (0.4% of examinations) and the one protocol-deviation generator replacement define the residual risk.

Watch out: Nothing here covers 3-tesla scanners, MRI-conditional devices (a different, manufacturer-tested question), or scanning without interrogation and reprogramming support.

Source: Synthesis of the carded records above (PMIDs 28225684, 29281579). Registry IDs: NCT00907361 (MagnaSafe), NCT01130896 (Nazarian).