Implantable Eversense glucose sensors: mean absolute relative difference against laboratory reference was 9.6% for the 90-day sensor (N = 35), 9.1% for the 180-day sensor (N = 181) and 8.8% for the 365-day sensor (N = 110), with no device- or procedure-related serious adverse events in any study (documented N = 326 across three prospective accuracy studies)
Shelf #11937 · Diabetes, glucose & metabolism
POSSIBLE (benefit)
Stamped: 2026-11-28.
Evidence published: 2019–2025.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In adults with type 1 or type 2 diabetes, how accurately and safely does the implantable Eversense continuous glucose monitoring system (90-day, 180-day and 365-day sensors) measure glucose against laboratory reference?
Balance of the evidence
POSSIBLE (benefit) — Composition: three prospective single-arm accuracy studies of successive generations of one implantable sensor system (90-day PRECISION, 180-day PROMISE, 365-day ENHANCE) — accuracy evidence, hence. Mean absolute relative difference (MARD) against Yellow Springs Instrument laboratory reference was 9.6% (PRECISION, 35 participants), 9.1% for the primary sensor in PROMISE (181 participants; 8.5% for the modified sacrificial-boronic-acid sensor) and 8.8% in ENHANCE (110 participants). Confirmed alert detection at 70 mg/dL (3.9 mmol/L) was 93% in PROMISE and 96.6% in ENHANCE. Sensor survival to the full wear period was 100% at 90 days in PRECISION, 65% (primary) and 90% (modified sensor) at 180 days in PROMISE, and 90% at 365 days in ENHANCE. No device- or insertion/removal-related serious adverse events occurred in any study. Boundary stated on the face: the library's general continuous-glucose-monitor accuracy shelf, #8605, covers the factory-calibrated systems (Libre, Dexcom, Guardian) and cards the one three-way home-use study that included an earlier Eversense sensor; this shelf is confined to the implantable system's own pivotal studies, which #8605 does not card, and accuracy against a laboratory reference is not evidence that using the sensor improves HbA1c or prevents hypoglycaemia.
MeSH: Adult · Aged · Blood Glucose · Blood Glucose / analysis* · Blood Glucose Self-Monitoring · Blood Glucose Self-Monitoring / instrumentation* · Blood Glucose Self-Monitoring* / instrumentation / adverse effects / methods · Blood Glucose* / analysis · Continuous Glucose Monitoring · Diabetes Mellitus, Type 1 / blood* / drug therapy · Diabetes Mellitus, Type 1* / blood / drug therapy · Diabetes Mellitus, Type 1* / drug therapy · Diabetes Mellitus, Type 2 / blood* / drug therapy · Diabetes Mellitus, Type 2* · Female · Humans · Hypoglycemia / blood · Hypoglycemic Agents / administration & dosage / therapeutic use · Insulin Infusion Systems / adverse effects* · Insulin Infusion Systems* / adverse effects · Insulin* / administration & dosage · Male · Middle Aged · Prospective Studies · Reproducibility of Results
Cards (4)
Card 1 — PRECISION 2019: the 90-day implantable sensor read within 15/15% of laboratory reference 85% of the time, MARD 9.6%, with no device-related serious adverse events (N = 35)
Study: Prospective accuracy study (35 participants)
Population: Prospective multicentre accuracy and safety study of the Eversense implantable continuous glucose monitoring system in adults with type 1 or type 2 diabetes through 90 days: 35 participants received the system. Accuracy was assessed as percentage agreement and mean absolute relative difference (MARD) against Yellow Springs Instrument reference values from 40 to 400 mg/dL; the primary safety endpoint was device- or insertion/removal-related serious adverse events through 90 days. An updated glucose algorithm was also applied retrospectively to the earlier PRECISE II study's data.
Summary: PRECISION 2019: the 90-day implantable sensor read within 15/15% of laboratory reference 85% of the time, MARD 9.6%, with no device-related serious adverse events (N = 35)
Result: 85% of sensor values were within 15/15% of reference and MARD was 9.6% (95% CI 8.9–10.4). All sensors remained functional through day 90 and no device- or procedure-related serious adverse events occurred. Applying the updated algorithm to PRECISE II data gave 87% of readings within 15/15% and a MARD of 8.5% (95% CI 8.0%–9.1%).
Watch out: Thirty-five participants at study clinics, with accuracy measured under in-clinic conditions against reference blood draws: real-world wear, insertion-site problems over repeated implants and performance in children are outside this record.
Source: Christiansen MP, Klaff LJ, Bailey TS et al.. A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study. Diabetes Technol Ther. 2019. PMID: 30925083. Registry ID: NCT02647905 (as stated in the record). MeSH: Adult · Aged · Blood Glucose / analysis* · Blood Glucose Self-Monitoring / instrumentation* · Diabetes Mellitus, Type 1 / blood* / drug therapy · Diabetes Mellitus, Type 2 / blood* / drug therapy · Female · Humans · Insulin Infusion Systems / adverse effects* · Male · Middle Aged · Prospective Studies · Reproducibility of Results
Card 2 — PROMISE 2022: the 180-day sensor's MARD was 9.1% (8.5% for the modified sensor); 92.9% of readings within 20%/20% of reference; primary sensors survived to 180 days in 65% (N = 181)
Study: Prospective accuracy study (181 participants)
Population: Prospective multicentre study at 8 United States sites: 181 participants with diabetes received a primary implantable sensor; 96 received a second sensor in the other arm (53 identical, 43 with a modified sacrificial boronic acid design). Accuracy was assessed against the YSI 2300 laboratory analyser at 10 clinic visits over 180 days, with confirmed alert detection rates, calibration stability and sensor survival.
Summary: PROMISE 2022: the 180-day sensor's MARD was 9.1% (8.5% for the modified sensor); 92.9% of readings within 20%/20% of reference; primary sensors survived to 180 days in 65% (N = 181)
Result: For primary sensors, 92.9% of readings were within 20%/20% of reference and overall MARD was 9.1%; confirmed alert detection was 93% at 70 mg/dL (3.9 mmol/L) and 99% at 180 mg/dL (10.0 mmol/L); 65% of primary sensors survived to 180 days. For the modified sensors, 93.9% of readings were within 20%/20% and MARD was 8.5%, alert detection was 94% and 99%, and 90% survived to 180 days. The median share of time on one calibration per day was 56% (primary) and 63% (modified). No device- or insertion/removal-related serious adverse events were reported.
Watch out: One in three primary sensors did not survive the intended 180 days — the accuracy figures describe sensors that kept working, and the improved survival belongs to a modified sensor design that was itself still investigational in this study.
Source: Garg SK, Liljenquist D, Bode B et al.. Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study. Diabetes Technol Ther. 2022. PMID: 34515521. Registry ID: NCT03808376. MeSH: Blood Glucose · Blood Glucose Self-Monitoring · Diabetes Mellitus, Type 1* / drug therapy · Diabetes Mellitus, Type 2* · Humans · Prospective Studies · Reproducibility of Results
Card 3 — ENHANCE 2025: the 365-day sensor's MARD was 8.8% with about one calibration per week; 90% of sensors survived a full year and no related serious adverse events occurred (N = 110)
Study: Prospective accuracy study (110 participants)
Population: Prospective multicentre study of the Eversense 365 implantable system in adults with diabetes: 110 participants were implanted and assessed at 14 in-clinic visits over 1 year against Yellow Springs Instrument reference measurements, including during hyperglycaemia and hypoglycaemia challenges.
Summary: ENHANCE 2025: the 365-day sensor's MARD was 8.8% with about one calibration per week; 90% of sensors survived a full year and no related serious adverse events occurred (N = 110)
Result: Overall MARD was 8.8%, with primarily one calibration per week. Confirmed alert detection was 96.6% at 70 mg/dL (3.9 mmol/L) and 97.9% at 180 mg/dL (10.0 mmol/L). 90% of sensors survived 365 days. Interoperable continuous glucose monitoring special controls were met, and no related serious adverse events were reported.
Watch out: Accuracy and survival in a manufacturer-run pivotal study of volunteers willing to have a sensor implanted and attend 14 clinic visits; the study cannot show the system improves glucose control, only that its readings track the laboratory reference.
Source: Bailey TS, Liljenquist DR, Denham DS et al.. Evaluation of Accuracy and Safety of the 365-Day Implantable Eversense Continuous Glucose Monitoring System: The ENHANCE Study. Diabetes Technol Ther. 2025. PMID: 39869118. Registry ID: NCT05131139. MeSH: Adult · Aged · Blood Glucose* / analysis · Blood Glucose Self-Monitoring* / instrumentation / adverse effects / methods · Continuous Glucose Monitoring · Diabetes Mellitus, Type 1* / blood / drug therapy · Female · Humans · Hypoglycemia / blood · Hypoglycemic Agents / administration & dosage / therapeutic use · Insulin* / administration & dosage · Insulin Infusion Systems* / adverse effects · Male · Middle Aged · Prospective Studies · Reproducibility of Results
Card 4 — Synthesis: across three generations, the implantable Eversense system reads glucose with a MARD of 8.8–9.6% against laboratory reference and a clean serious-harm record in 326 study participants — POSSIBLE (benefit) as an accuracy claim only
Study: Synthesis
Population: The carded record: PRECISION (35 participants), PROMISE (181) and ENHANCE (110); total documented N = 326, separate cohorts, additive. The three-way home-use comparison carded at #8605 is deliberately not re-carded here.
Summary: Synthesis: across three generations, the implantable Eversense system reads glucose with a MARD of 8.8–9.6% against laboratory reference and a clean serious-harm record in 326 study participants — POSSIBLE (benefit) as an accuracy claim only
Result: The accuracy figures are consistent across generations and wear durations, which supports POSSIBLE (benefit) for the narrow claim 'the sensor measures glucose accurately for its labelled life'. Nothing carded tests whether implanting it improves any clinical outcome, so the pack is Tier C and the stamp reaches no further.
Watch out: Children, pregnancy, and head-to-head accuracy against current Libre or Dexcom sensors are outside this record; the comparative accuracy question across systems is homed at #8605.
Source: Synthesis of the carded records above (PMIDs 30925083, 34515521, 39869118). Registry IDs: NCT02647905, NCT03808376, NCT05131139.