Magseed versus wire-guided localisation for non-palpable breast lesions: excision succeeded in all patients in both groups and re-excision did not differ (16% versus 14%) in a two-centre prospective non-randomised comparison (documented N = 200 patients)

Shelf #12126 · Cancer

POSSIBLE (the treatment made no difference)

Stamped: 2026-11-28.

Evidence published: 2019.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients having wide local excision for non-palpable breast lesions, does magnetic seed (Magseed) localisation, compared with wire-guided localisation, change successful excision, re-excision and specimen size?

Balance of the evidence

POSSIBLE (the treatment made no difference) — A single prospective non-randomised comparison from two academic breast units found magnetic seed localisation (Magseed, detected with the Sentimag probe, both named as tested in the record; the record names no manufacturer or company, and none is named on this pack) performed the same as wire-guided localisation on the outcomes that matter: intraoperative identification and excision succeeded in all patients in both groups, re-excision did not differ (16% versus 14%, P = 0.692), and specimen weight and volume were similar. Four of 104 Magseed cases as printed in the record were converted to wire localisation preoperatively (2 misplaced seeds, 2 undetected seeds) — a failure mode wires do not have, printed on the face. The stamp is Tier C: one non-randomised comparison, so equivalence here is 'no difference shown in a modest cohort', not proven interchangeability. Wave 17 Block A flagged magnetic-seed localisation as an unverified reserve; it is built now on this record.

MeSH: Breast Neoplasms / diagnostic imaging / pathology / surgery* · Carcinoma / diagnostic imaging / pathology / surgery* · Carcinoma, Intraductal, Noninfiltrating / diagnostic imaging / pathology / surgery* · Female · Humans · Magnets* · Mammography · Mastectomy, Segmental / methods* · Middle Aged · Neoadjuvant Therapy · Surgery, Computer-Assisted / methods* · Tumor Burden · Ultrasonography, Mammary

Cards (2)

Card 1 — Prospective comparison 2019: Magseed matched wire localisation on successful excision and re-excision (16% versus 14%); 4 of 104 seeds needed conversion to wire

Study: Prospective non-randomised comparative study (200 patients in two cohorts of 100)

Population: Prospective multi-centre non-randomised controlled study in two academic breast units (October 2017 to September 2018): consecutive patients undergoing image-guided wide local excision for non-palpable lesions, 100 planned for Magseed localisation and a cohort of 100 undergoing wire-guided localisation in the same period; demographic and disease characteristics were well balanced. Magseed is a paramagnetic seed visualised on mammography and ultrasound and detected intraoperatively with the Sentimag probe, as named in the record; the record names no manufacturer or company.

Summary: Prospective comparison 2019: Magseed matched wire localisation on successful excision and re-excision (16% versus 14%); 4 of 104 seeds needed conversion to wire

Result: Intraoperative identification and excision of the localised lesion was successful in all patients, confirmed by specimen radiography. Re-excision did not differ significantly (Magseed 16% versus wire 14%, P = 0.692). Specimen weight (39.6 g versus 44.5 g, P = 0.206) and volume (90.1 cm3 versus 95.6 cm3, P = 0.579) were similar. Four of 104 patients as printed in the record were converted preoperatively from Magseed to wire localisation (2 misplaced seeds, 2 undetected seeds). The authors conclude Magseed was as reliable and effective as wire localisation for identification, excision with tumour-free margins and specimen weight.

Watch out: Not randomised — unit- and period-based allocation can hide case-mix differences a balanced table does not capture, and 200 patients cannot exclude a modest re-excision difference; the 4/104 conversion denominator as printed exceeds the stated 100-patient Magseed accrual, which the record does not explain.

Source: Zacharioudakis K, Down S, Bholah Z, Lee S, Khan T, Maxwell AJ. Is the future magnetic? Magseed localisation for non palpable breast cancer. A multi-centre non randomised control study. Eur J Surg Oncol. 2019. PMID: 31288944. Registry ID: none stated in the record. MeSH: Breast Neoplasms / diagnostic imaging / pathology / surgery* · Carcinoma / diagnostic imaging / pathology / surgery* · Carcinoma, Intraductal, Noninfiltrating / diagnostic imaging / pathology / surgery* · Female · Humans · Magnets* · Mammography · Mastectomy, Segmental / methods* · Middle Aged · Neoadjuvant Therapy · Surgery, Computer-Assisted / methods* · Tumor Burden · Ultrasonography, Mammary

Card 2 — Synthesis: on this single non-randomised record, Magseed and wire localisation performed the same, with a small seed-specific conversion risk

Study: Synthesis

Population: The carded record: the 2019 two-centre comparison; total documented N = 200 patients (two cohorts of 100 as described in the record).

Summary: Synthesis: on this single non-randomised record, Magseed and wire localisation performed the same, with a small seed-specific conversion risk

Result: No outcome separated the techniques. POSSIBLE (the treatment made no difference), Tier C — a single non-randomised study cannot carry a firmer equivalence claim.

Watch out: Radioactive seed, radar and radiofrequency localisation are different wire-free techniques and are not covered by this record; scheduling and patient-preference advantages claimed for wire-free methods were not measured here.

Source: Synthesis of the carded record above (PMID 31288944).