SourceMed · Shelf #12537 in the ask interface

ECCO2R to avoid intubation in COPD exacerbation (ECLAIR): intubation was avoided in 56% of 25 cases, but major bleeding occurred in 36% and 90-day mortality was identical (28% versus 28%) in a case-control study (documented N = 50 in total)

Shelf #12537 · Lungs, breathing & sleep

MAYBE (weak evidence)

Stamped: 2026-11-28.

Evidence published: 2016.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with an acute COPD exacerbation and hypercapnic respiratory failure that has not responded to noninvasive ventilation, does veno-venous extracorporeal carbon dioxide removal (ECCO2R) avoid endotracheal intubation compared with invasive mechanical ventilation in matched historical controls?

Balance of the evidence

MAYBE (weak evidence) — the only record is a multicentre prospective case-control study with historical controls, not a randomised trial: 25 patients treated with a pump-driven veno-venous ECCO2R system (the iLA-Activve system named in the record) matched with 25 historical controls treated with invasive mechanical ventilation. Intubation was avoided in 14 of 25 ECCO2R patients (56.0%); seven were later intubated for progressive hypoxaemia and four for ventilatory failure despite ECCO2R and noninvasive ventilation. The harm side is substantial: relevant ECCO2R-associated adverse events occurred in 11 patients (44.0%), of whom 9 (36.0%) had major bleeding complications. There was no outcome advantage: mean time on invasive ventilation was 8.3 versus 13.7 days, intensive care stay 28.9 versus 24.0 days, hospital stay 36.9 versus 37.0 days, and 90-day mortality was 28.0% in both groups. Historical controls and a 25-patient treated group cannot show that ECCO2R helps patients live longer or leave hospital sooner — only that intubation can sometimes be avoided, at a high bleeding cost; the authors call for larger randomised studies. ECCO2R in ARDS (the REST trial, ask #1077) is a different population and question.

MeSH: Aged · Aged, 80 and over · Carbon Dioxide · Case-Control Studies · Extracorporeal Circulation / adverse effects* / methods · Female · Humans · Hypercapnia / etiology / therapy · Intensive Care Units · Intubation, Intratracheal / adverse effects* / statistics & numerical data · Male · Middle Aged · Noninvasive Ventilation / adverse effects · Pilot Projects · Prospective Studies · Pulmonary Disease, Chronic Obstructive / complications · Respiration, Artificial / adverse effects* / statistics & numerical data · Respiratory Insufficiency / etiology / therapy*

Cards (2)

Card 1 — ECLAIR 2016: ECCO2R avoided intubation in 14 of 25 NIV-refractory COPD patients — but 36% had major bleeding and mortality matched controls at 28%

Study: Case-control study, prospective cases with historical controls (25 cases, 25 controls)

Population: Multicentre prospective case-control study in five European intensive care units: 25 patients with acute hypercapnic respiratory failure refractory to noninvasive ventilation (respiratory acidosis, pH ≤ 7.35 and PaCO2 > 45 mm Hg, meeting predefined intubation criteria) were treated with a pump-driven veno-venous ECCO2R system (iLA-Activve; Novalung, Heilbronn, Germany — the device and maker as named in the record) and matched with 25 historical controls treated with invasive mechanical ventilation, matched on main diagnosis, age, SAPS-II score and pH.

Summary: ECLAIR 2016: ECCO2R avoided intubation in 14 of 25 NIV-refractory COPD patients — but 36% had major bleeding and mortality matched controls at 28%

Result: Intubation was avoided in 14 patients (56.0%) in the ECCO2R group, at a mean extracorporeal blood flow of 1.3 L/min; seven patients were intubated for progressive hypoxaemia and four for ventilatory failure despite ECCO2R and noninvasive ventilation. Relevant ECCO2R-associated adverse events occurred in 11 patients (44.0%), of whom 9 (36.0%) suffered major bleeding complications. Mean time on invasive ventilation was 8.3 versus 13.7 days, intensive care stay 28.9 versus 24.0 days and hospital stay 36.9 versus 37.0 days in the case and control groups, and 90-day mortality was 28.0% versus 28.0%.

Watch out: Historical controls are not a randomised comparison: differences in era and case selection can move every outcome here, and with 25 treated patients the identical mortality rates carry very wide uncertainty. The tested device is named as the record names it; other ECCO2R systems may differ in safety and effectiveness.

Source: Braune S, Sieweke A, Brettner F et al.. The feasibility and safety of extracorporeal carbon dioxide removal to avoid intubation in patients with COPD unresponsive to noninvasive ventilation for acute hypercapnic respiratory failure (ECLAIR study): multicentre case-control study. Intensive Care Med. 2016. PMID: 27456703. Registry ID: NCT01784367 (stated in the record). MeSH: Aged · Aged, 80 and over · Carbon Dioxide · Case-Control Studies · Extracorporeal Circulation / adverse effects* / methods · Female · Humans · Hypercapnia / etiology / therapy · Intensive Care Units · Intubation, Intratracheal / adverse effects* / statistics & numerical data · Male · Middle Aged · Noninvasive Ventilation / adverse effects · Pilot Projects · Prospective Studies · Pulmonary Disease, Chronic Obstructive / complications · Respiration, Artificial / adverse effects* / statistics & numerical data · Respiratory Insufficiency / etiology / therapy*

Card 2 — Synthesis: intubation avoidable in about half of cases in uncontrolled evidence, with major bleeding in about a third — MAYBE (weak evidence)

Study: Synthesis

Population: The carded record: the ECLAIR study (25 ECCO2R cases and 25 matched historical controls); total documented N = 50 across both groups.

Summary: Synthesis: intubation avoidable in about half of cases in uncontrolled evidence, with major bleeding in about a third — MAYBE (weak evidence)

Result: Feasibility is shown; effectiveness is not. No randomised trial of ECCO2R for this question appears in this record, and the mortality and length-of-stay results show no advantage, so the stamp is MAYBE (weak evidence) and.

Watch out: Patients who respond to noninvasive ventilation, and ECCO2R used in ARDS rather than COPD exacerbation (ask #1077), are outside this record.

Source: Synthesis of the carded record above (PMID 27456703). Registry ID: NCT01784367.