Microaxial flow pump (Impella CP) in STEMI-related cardiogenic shock (DanGer Shock)

Shelf #1536 · Heart, vessels, blood pressure & clotting

Stamped: development phase, October–November 2026 — exact date not recorded.

Evidence published: 2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with STEMI and cardiogenic shock, does a microaxial flow pump (Impella CP) plus standard care, compared with standard care alone, reduce all-cause mortality at 180 days?

Balance of the evidence

POSSIBLE: DanGer Shock (360 randomised; 355 analysed) is the first randomised trial of a mechanical support device to show a mortality benefit in infarct shock: death from any cause at 180 days occurred in 45.8% with the microaxial flow pump versus 58.5% with standard care (HR 0.74 relative, 95% CI 0.55 to 0.99; P=0.04) — an absolute difference of 12.7 percentage points. The price was substantial: the composite safety end point (severe bleeding, limb ischaemia, haemolysis, device failure, or worsening aortic regurgitation) occurred in 24.0% versus 6.2% (RR 4.74 relative), and renal-replacement therapy in 41.9% versus 26.7%. One trial of 355 analysed patients, with a P value of 0.04 and heavy harms, earns POSSIBLE — a real but fragile and costly benefit claim after the null device trials on Shelves #1534 and #1535.

MeSH: Aged · Female · Humans · Male · Heart-Assist Devices · Incidence · Shock, Cardiogenic · ST Elevation Myocardial Infarction · Treatment Outcome · Assisted Circulation

Cards (2)

Card 1 — DanGer Shock 2024: Impella CP lowered 180-day mortality in STEMI shock (45.8% vs 58.5%; HR 0.74) but quadrupled the composite of severe device-related harms

Study: International, multicentre, randomised trial

Population: 360 patients with STEMI and cardiogenic shock underwent randomisation to a microaxial flow pump (Impella CP) plus standard care or standard care alone; 355 were included in the final analysis (179 pump; 176 standard care). The median age was 67 years, and 79.2% were men.

Summary: DanGer Shock 2024: Impella CP lowered 180-day mortality in STEMI shock (45.8% vs 58.5%; HR 0.74) but quadrupled the composite of severe device-related harms

Result: Death from any cause at 180 days occurred in 82 of 179 (45.8%) patients in the pump group and 103 of 176 (58.5%) in the standard-care group (HR 0.74 — a relative measure — 95% CI 0.55 to 0.99; P=0.04). A composite safety end-point event (severe bleeding, limb ischaemia, haemolysis, device failure, or worsening aortic regurgitation) occurred in 43 patients (24.0%) versus 11 (6.2%) (RR 4.74 relative, 2.36 to 9.55). Renal-replacement therapy was administered to 75 (41.9%) pump patients versus 47 (26.7%) (RR 1.98 relative, 1.27 to 3.09).

Watch out: The record's abstract reports the safety outcomes only as a composite plus renal-replacement therapy; it does not print separate rates for bleeding, limb ischaemia or haemolysis, so the topic draft's itemised harm list is not carded beyond the composite. P=0.04 with an interval reaching 0.99 makes the mortality result statistically fragile, and the trial enrolled a selected shock population over a decade.

Source: Møller JE et al. Microaxial Flow Pump or Standard Care in Infarct-Related Cardiogenic Shock. The New England journal of medicine. 2024. PMID: 38587239. Registry ID: ClinicalTrials.gov NCT01633502, from the record's databank fields. MeSH: Aged · Female · Humans · Male · Heart-Assist Devices · Incidence · Shock, Cardiogenic · ST Elevation Myocardial Infarction · Treatment Outcome · Assisted Circulation

Card 2 — Synthesis: the first device win in infarct shock — real mortality benefit, heavy harm price, one trial — POSSIBLE

Study: Synthesis

Population: Synthesis of the card above: DanGer Shock 2024 (355 patients analysed with STEMI and cardiogenic shock).

Summary: Synthesis: the first device win in infarct shock — real mortality benefit, heavy harm price, one trial — POSSIBLE

Result: Against the null backdrop of IABP (#1534) and VA-ECMO (#1535), DanGer Shock's 12.7-point absolute mortality reduction is the outlier result in shock devices — and it arrives with a 24.0% rate of severe device-related complications and a 15-point excess of dialysis. Adoption turns on patient selection and centre expertise, neither of which a single trial can specify.

Watch out: Do not generalise across devices: DanGer tested one pump, placed early by protocol, in STEMI shock; it rehabilitates neither balloon pumps nor ECMO, whose own trials are null.

Source: Synthesis of Møller JE et al. The New England journal of medicine. 2024. PMID: 38587239. Registry ID as on the individual card.