Fully magnetically levitated centrifugal LVAD (HeartMate 3) in advanced heart failure (MOMENTUM 3)
Shelf #1576 · Heart, vessels, blood pressure & clotting
Stamped: development phase, October–November 2026 — exact date not recorded.
Evidence published: 2019–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with advanced heart failure eligible for a durable LVAD, does the fully magnetically levitated centrifugal-flow pump, compared with the axial-flow pump, improve survival free of disabling stroke or reoperation for pump malfunction?
Balance of the evidence
LIKELY: MOMENTUM 3's final report (1,028 patients with advanced heart failure, randomised irrespective of bridge-to-transplant or destination-therapy intent) found the centrifugal-flow HeartMate 3 superior on the primary composite — alive at 2 years free of disabling stroke or reoperation to replace or remove a malfunctioning device — in 76.9% versus 64.8% on the axial-flow HeartMate II (P<0.001 for superiority), driven above all by pump replacement (2.3% vs 11.3%; RR 0.21 relative): the magnetically levitated rotor nearly abolishes pump thrombosis. The 5-year extended-phase follow-up of the same trial (an observational, per-protocol analysis of patients still on support at 2 years — carded as a follow-up report, not a second trial) found the composite at 54.0% versus 29.7% (HR 0.55 relative) and overall survival 58.4% versus 43.7% (HR 0.72 relative; P=0.003). One trial programme, randomised once against the incumbent device: the device comparison carries LIKELY; the 5-year survival figures carry the observational-extension caveat on their card. The topic draft pinned the trial's 2018 interim report (first 366 patients); the final report leads this shelf and the interim report is fenced as superseded.
MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Disease-Free Survival · Female · Heart Failure · Heart-Assist Devices · Humans · Intention to Treat Analysis · Kaplan-Meier Estimate · Male · Middle Aged · Prosthesis Design · Prosthesis Failure · Reoperation · Stroke
Cards (3)
Card 1 — MOMENTUM 3 final report 2019: HeartMate 3 beat axial-flow on survival free of disabling stroke or reoperation — 76.9% vs 64.8% at 2 years
Study: Randomised controlled trial (final report; noninferiority and superiority design)
Population: 1,028 patients with advanced heart failure randomly assigned to the fully magnetically levitated centrifugal-flow pump (516) or the mechanical-bearing axial-flow pump (512), irrespective of the intended goal of support. The abstract does not state mean age or sex mix.
Summary: MOMENTUM 3 final report 2019: HeartMate 3 beat axial-flow on survival free of disabling stroke or reoperation — 76.9% vs 64.8% at 2 years
Result: At 2 years, 397 patients (76.9%) in the centrifugal-flow group versus 332 (64.8%) in the axial-flow group remained alive and free of disabling stroke or reoperation to replace or remove a malfunctioning device (P<0.001 for superiority). The record prints a relative risk of 0.84 (95% CI 0.78 to 0.91) alongside these proportions — as printed, that ratio runs with the risk of failing the composite, opposite to the direction of the percentages; the card prints both exactly as they appear and flags the point rather than repairing it. Pump replacement was less common with the centrifugal pump (12 patients [2.3%] vs 57 [11.3%]; RR 0.21 relative, 0.11 to 0.38; P<0.001), and events per patient-year for stroke of any severity, major bleeding, and gastrointestinal haemorrhage were lower (rates not printed in the abstract).
Watch out: An earlier interim report of this trial (NEJM 2018, PMID 29526139; first 366 patients, 79.5% vs 60.2%) is superseded by this final report and is not counted as separate evidence. The comparator is a device no longer implanted in most programmes, so the shelf's modern relevance is the pump-thrombosis lesson, not a live device choice in most centres.
Source: Mehra MR et al. A Fully Magnetically Levitated Left Ventricular Assist Device — Final Report. The New England journal of medicine. 2019. PMID: 30883052. Registry ID: ClinicalTrials.gov NCT02224755, from the record's databank fields. MeSH: Adolescent · Adult · Aged · Aged, 80 and over · Disease-Free Survival · Female · Heart Failure · Heart-Assist Devices · Humans · Intention to Treat Analysis · Kaplan-Meier Estimate · Male · Middle Aged · Prosthesis Design · Prosthesis Failure · Reoperation · Stroke
Card 2 — MOMENTUM 3 five-year extended phase 2022 (follow-up of the same trial): survival 58.4% vs 43.7% — observational, per-protocol
Study: Extended-phase observational follow-up of the same randomised trial (not a second randomisation)
Population: Of the 1,020 patients in the trial's per-protocol population, 536 were still receiving LVAD support at 2 years; 477 (295 enrolled plus 182 providing limited data) contributed to the 5-year analysis. Median age in this follow-up set was 62 years and 18% were women.
Summary: MOMENTUM 3 five-year extended phase 2022 (follow-up of the same trial): survival 58.4% vs 43.7% — observational, per-protocol
Result: The 5-year Kaplan-Meier estimate of survival to transplant, recovery, or LVAD support free of debilitating stroke or reoperation was 54.0% with the centrifugal-flow pump versus 29.7% with the axial-flow pump (hazard ratio 0.55 relative, 95% CI 0.45 to 0.67; P<0.001). Overall Kaplan-Meier survival was 58.4% versus 43.7% (HR 0.72 relative, 0.58 to 0.89; P=0.003). Serious adverse events of stroke, bleeding, and pump thrombosis were less frequent in the centrifugal-flow group.
Watch out: This analysis conditions on still receiving LVAD support at 2 years and is per-protocol and observational — it extends the randomised finding in time but cannot carry the same causal weight, and survivorship selection runs in both arms. It is carded as the same trial's follow-up, never as independent confirmation.
Source: Mehra MR et al. Five-Year Outcomes in Patients With Fully Magnetically Levitated vs Axial-Flow Left Ventricular Assist Devices in the MOMENTUM 3 Randomized Trial. JAMA. 2022. PMID: 36074476. Registry IDs: ClinicalTrials.gov NCT02224755 and NCT03982979, from the record's databank fields. MeSH: Female · Follow-Up Studies · Heart Failure · Heart-Assist Devices · Humans · Kaplan-Meier Estimate · Male · Middle Aged · Stroke · Treatment Outcome
Card 3 — Synthesis: magnetic levitation wins the device comparison — LIKELY; five-year survival gap is follow-up evidence, not a second trial
Study: Synthesis
Population: Synthesis of the cards above: MOMENTUM 3 final report 2019 (1,028 patients) and its 5-year extended phase 2022 (477 patients analysed).
Summary: Synthesis: magnetic levitation wins the device comparison — LIKELY; five-year survival gap is follow-up evidence, not a second trial
Result: The randomised comparison is unambiguous and mechanistically coherent — eliminate the bearing and the thrombosis, and the composite follows. The 5-year data suggest durability and a survival separation, with the stated observational caveats.
Watch out: One programme, one randomisation: the shelf counts MOMENTUM 3 once. The practical question in most centres has moved on (the axial-flow comparator is largely historical), which the balance acknowledges.
Source: Synthesis of Mehra MR et al. The New England journal of medicine. 2019. PMID: 30883052, and Mehra MR et al. JAMA. 2022. PMID: 36074476. Registry IDs as on the individual cards.