SourceMed · Shelf #1765 in the ask interface

Suspend-before-low and threshold-suspend insulin pumps in type 1 diabetes (PROLOG, ASPIRE)

Shelf #1765 · Diabetes, glucose & metabolism

Stamped: development phase, October–November 2026 — exact date not recorded.

Evidence published: 2013–2018.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In patients with type 1 diabetes at risk of hypoglycaemia, do automated insulin-suspension pump features reduce hypoglycaemic events versus sensor-augmented pump therapy without suspension?

Balance of the evidence

LIKELY — two randomised trials of successive generations of the same idea agree. PROLOG (predictive low-glucose suspend, n=103, 6 weeks): median sensor time below 70 mg/dl was 2.6% with suspend-before-low versus 3.2% on sensor-augmented pump (difference −0.8 percentage points; P<0.001), a 31% relative reduction in mean time below range, with no increase in mean glucose. ASPIRE (threshold suspend, n=247, 3 months): the area-under-curve rate of nocturnal hypoglycaemic events was 37.5% lower (a relative measure; P<0.001), the event rate was 31.8% lower (also relative), with nocturnal events at 1.5 versus 2.2 per patient-week — the absolute rates are printed beside the relative ones. PROLOG recorded one severe hypoglycaemic event, in the sensor-augmented-pump period; ASPIRE recorded four patients with severe events, all in the control group, and no DKA in either. Mean glucose did not rise in PROLOG and HbA1c did not differ in ASPIRE — suspension is a hypoglycaemia-safety feature, not a glucose-lowering one. Generation boundary stated: threshold-suspend reacts at a threshold; suspend-before-low predicts; full hybrid closed loop is Shelf #560/#561 territory and is not claimed here.

MeSH: Adolescent · Adult · Aged · Algorithms · Blood Glucose/analysis · Blood Glucose Self-Monitoring/methods · Child · Cross-Over Studies · Diabetes Mellitus, Type 1/blood/diagnosis/drug therapy · Drug Delivery Systems · Female · Humans · Hypoglycemia/blood/chemically induced/prevention & control · Hypoglycemic Agents/administration & dosage · Insulin/administration & dosage · Insulin Infusion Systems · Male · Middle Aged · Monitoring, Ambulatory · Young Adult

Cards (3)

Card 1 — PROLOG: predictive low-glucose suspend cut median time below 70 mg/dl to 2.6% versus 3.2% on sensor-augmented pump

Study: Multicentre, randomised crossover trial (6 weeks; two 3-week periods)

Population: 103 people with type 1 diabetes (age 6–72 years; mean HbA1c 7.3%) took part in a randomised crossover comparison of the predictive low-glucose suspend system (t:slim X2 with Basal-IQ) versus sensor-augmented pump therapy; 99% completed both study periods. The abstract does not print the sex split.

Summary: PROLOG: predictive low-glucose suspend cut median time below 70 mg/dl to 2.6% versus 3.2% on sensor-augmented pump

Result: Median time below 70 mg/dl fell from 3.6% at baseline to 2.6% in the PLGS period versus 3.2% in the SAP period (difference −0.8 percentage points, 95% CI −1.1 to −0.5; P<0.001); the corresponding mean values (4.4% at baseline, 3.1% PLGS, 4.5% SAP) represent a 31% relative reduction. Mean sensor glucose did not increase (159 vs 159 mg/dl; P=0.40), nor did time above 180 mg/dl (32% vs 33%; P=0.12). One severe hypoglycaemic event occurred in the SAP period and none in the PLGS period. Mean pump suspension time was 104 minutes/day.

Watch out: Six weeks is short (two 3-week crossover periods), and a crossover of device features cannot blind participants. The 31% figure is the relative reduction in mean time below 70 mg/dl; the absolute median difference was 0.8 percentage points — the carded primary result.

Source: Forlenza GP et al. Predictive Low-Glucose Suspend Reduces Hypoglycemia in Adults, Adolescents, and Children With Type 1 Diabetes in an At-Home Randomized Crossover Study: Results From the PROLOG Trial. Diabetes Care. 2018. PMID: 30089663. Registry ID: ClinicalTrials.gov NCT03195140, from the record's databank fields. MeSH: Adolescent · Adult · Aged · Algorithms · Blood Glucose/analysis · Blood Glucose Self-Monitoring/methods · Child · Cross-Over Studies · Diabetes Mellitus, Type 1/blood/diagnosis/drug therapy · Drug Delivery Systems · Female · Humans · Hypoglycemia/blood/chemically induced/prevention & control · Hypoglycemic Agents/administration & dosage · Insulin/administration & dosage · Insulin Infusion Systems · Male · Middle Aged · Monitoring, Ambulatory · Young Adult

Card 2 — ASPIRE: threshold-suspend reduced nocturnal hypoglycaemic event exposure by 37.5% (relative), with events at 1.5 versus 2.2 per patient-week

Study: Multicentre, randomised controlled trial (3 months; nocturnal hypoglycaemia phase)

Population: 247 patients with type 1 diabetes and documented nocturnal hypoglycaemia (121 threshold-suspend, 126 sensor-augmented pump control). The abstract does not print mean age, diabetes duration, HbA1c, or the sex split.

Summary: ASPIRE: threshold-suspend reduced nocturnal hypoglycaemic event exposure by 37.5% (relative), with events at 1.5 versus 2.2 per patient-week

Result: The co-primary outcome of nocturnal hypoglycaemic-event area under the curve was 980 ± 1200 versus 1568 ± 1995 mg/dl × minutes on control — 37.5% lower (a relative measure; P<0.001). Nocturnal hypoglycaemic events occurred 31.8% less frequently (relative): 1.5 ± 1.0 versus 2.2 ± 1.3 events per patient-week (P<0.001). The percentages of nocturnal sensor values below 50, 50 to <60, and 60 to <70 mg/dl were each significantly reduced (P<0.001 for each range; the record prints no percentages). HbA1c changes were similar in the two groups (the record prints no figures). Four patients, all in the control group, had a severe hypoglycaemic event; no patient had diabetic ketoacidosis.

Watch out: The headline percentages are relative reductions in event exposure; the absolute night-level rates (1.5 vs 2.2 events per patient-week) are the scale of benefit a patient experiences. Threshold-suspend is the older, reactive generation of the feature — carded alongside PROLOG's predictive version with the generational distinction stated, not merged.

Source: Bergenstal RM et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J Med. 2013. PMID: 23789889. Registry ID: ClinicalTrials.gov NCT01497938, from the record's databank fields. MeSH: Adolescent · Adult · Aged · Area Under Curve · Blood Glucose/analysis · Diabetes Mellitus, Type 1/blood/drug therapy · Female · Glycated Hemoglobin/metabolism · Humans · Hypoglycemia/chemically induced/prevention & control · Hypoglycemic Agents/administration & dosage/adverse effects · Insulin/administration & dosage/adverse effects · Insulin Infusion Systems/adverse effects · Male · Middle Aged · Young Adult

Card 3 — Synthesis: automatic suspension prevents hypoglycaemia without improving HbA1c — LIKELY

Study: Synthesis

Population: Synthesis of the cards above: PROLOG (Forlenza 2018), 103 patients, and ASPIRE (Bergenstal 2013), 247 patients — two trials, two technology generations.

Summary: Synthesis: automatic suspension prevents hypoglycaemia without improving HbA1c — LIKELY

Result: Across both generations, automated insulin suspension reduces hypoglycaemic exposure in type 1 diabetes patients at risk, day and night, with no HbA1c cost. The benefit is safety, and it is consistent.

Watch out: Both trials are short (6 weeks, 3 months) and industry-conducted in selected at-risk populations; neither addresses severe-hypoglycaemia rates as a powered endpoint.

Source: Synthesis of Forlenza GP et al. Diabetes Care. 2018 (PMID: 30089663) and Bergenstal RM et al. N Engl J Med. 2013 (PMID: 23789889). Registry IDs as on the trial cards.