In nonvalvular atrial fibrillation at stroke risk, is the dual-seal Amulet occluder noninferior to Watchman for occlusion success, safety, and stroke prevention?
Balance of the evidence
A large device-versus-device trial with a split practical answer: the same protection, a better seal, a rockier procedure. Amulet IDE randomised 1,878 patients with nonvalvular AF at increased stroke risk to the Amulet dual-seal occluder or Watchman. Amulet was noninferior on the primary safety composite (14.5% vs 14.7% at 12 months) and on effectiveness — ischaemic stroke or systemic embolism 2.8% versus 2.8% at 18 months — and the composite of stroke, systemic embolism or cardiovascular/unexplained death was 5.6% versus 7.7% (noninferior). Complete occlusion at 45 days was higher with Amulet (98.9% vs 96.8% with residual jet at most 5 mm; P=0.003 for superiority). The trade: procedure-related complications were higher with Amulet (4.5% vs 2.5%), mainly more pericardial effusions and device embolisations, falling with operator experience; major bleeding and death were similar. LIKELY: device choice can rest on anatomy and operator experience; neither device prevents stroke better at 18 months. (Desk note: this record's MeSH is sparse as published — copied verbatim, not supplemented.)
MeSH: Aged · Female · Humans · Male · Septal Occluder Device · Stroke
Cards (2)
Card 1 — Amulet IDE 2021: safety 14.5% vs 14.7% and stroke/embolism 2.8% vs 2.8% (noninferior); occlusion 98.9% vs 96.8% (superior); procedure complications 4.5% vs 2.5%
Study: Randomised controlled trial (Amulet IDE)
Population: 1,878 patients with nonvalvular AF at increased stroke risk, randomised 1:1 to percutaneous implantation of the Amplatzer Amulet occluder or the Watchman device. Primary endpoints: safety (procedure-related complications, all-cause death, or major bleeding at 12 months), effectiveness (ischaemic stroke or systemic embolism at 18 months), and LAA occlusion rate at 45 days.
Summary: Amulet IDE 2021: safety 14.5% vs 14.7% and stroke/embolism 2.8% vs 2.8% (noninferior); occlusion 98.9% vs 96.8% (superior); procedure complications 4.5% vs 2.5%
Result: Primary safety: 14.5% (Amulet) versus 14.7% (Watchman) (difference −0.14 points; 95% CI −3.42 to 3.13; P<0.001 for noninferiority); major bleeding 10.6% versus 10.0% and all-cause death 3.9% versus 5.1% within that composite. Procedure-related complications were higher with Amulet (4.5% vs 2.5%), largely pericardial effusion and device embolisation. Primary effectiveness: 2.8% versus 2.8% (difference 0.00; 95% CI −1.55 to 1.55; P<0.001 for noninferiority). Stroke/systemic embolism/cardiovascular-or-unexplained death: 5.6% versus 7.7% (noninferior). Major bleeding at 18 months: 11.6% versus 12.3% (P=0.32 for superiority). LAA occlusion at 45 days: 98.9% versus 96.8% (difference 2.03 points; P=0.003 for superiority).
Watch out: Noninferiority between two devices — neither is compared here with anticoagulation or with no treatment (see Shelves #272 and #2249 for those settings). The early-complication excess with Amulet fell with operator experience, so centre volume is part of the result.
Source: Lakkireddy D et al. Amplatzer Amulet Left Atrial Appendage Occluder Versus Watchman Device for Stroke Prophylaxis (Amulet IDE): A Randomized, Controlled Trial. Circulation. 2021. PMID: 34459659. Registry ID: NCT02879448, from the record. MeSH: Aged · Female · Humans · Male · Septal Occluder Device · Stroke (as published on the record).
Card 2 — Synthesis: equal protection, better seal, more early procedural complications — LIKELY
Study: Synthesis
Population: 1,878 patients (Lakkireddy 2021).
Summary: Synthesis: equal protection, better seal, more early procedural complications — LIKELY
Result: Safety and effectiveness noninferior; occlusion superior (98.9% vs 96.8%); procedure-related complications 4.5% vs 2.5%. LIKELY.
Watch out: A device-choice shelf, not a whether-to-occlude shelf.
Source: Synthesis of Lakkireddy D et al. Circulation. 2021 (PMID: 34459659). Registry ID as on the trial card.