SourceMed · Shelf #2294 in the ask interface
Remote monitoring of implanted devices: faster evaluation and fewer visits — outcome benefits inconsistent (IN-TIME + TRUST + REM-HF)
Shelf #2294 · Heart, vessels, blood pressure & clotting
Stamped: development phase, October–November 2026 — exact date not recorded.
Evidence published: 2010–2017.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients with heart failure and an implanted defibrillator or resynchronisation device, does automatic remote monitoring improve clinical outcomes, or mainly logistics?
Balance of the evidence
Three randomised trials, three different answers — because they tested three different intensities of monitoring. TRUST (1,339 ICD patients) tested logistics and won clearly: automatic daily home monitoring cut in-hospital device evaluations by 45%, handled 85.8% of mid-year follow-ups remotely, and cut median time from an arrhythmic event to physician evaluation from 36 days to under 2 days (P<0.001), with no morbidity penalty. IN-TIME (664 heart-failure patients, ejection fraction at most 35%, recent ICD or CRT-D) tested one manufacturer's daily multiparameter telemonitoring with a clinical response pathway: the composite clinical score (death, heart-failure hospitalisation, NYHA change, patient self-assessment) worsened in fewer monitored patients (18.9% vs 27.2%; odds ratio 0.63; 95% CI 0.43 to 0.90; P=0.013), and deaths were 10 versus 27 — a striking secondary split the trial was not powered to confirm. REM-HF (1,650 heart-failure patients with devices in England; weekly downloads with formalised follow-up) then found nothing: death or unplanned cardiovascular hospitalisation 42.4% versus 40.8% (hazard ratio 1.01; 95% CI 0.87 to 1.18; P=0.87), and no secondary endpoint differed. POSSIBLE (Tier C): adopt remote monitoring for what is proven — faster event detection, fewer routine visits — and treat outcome benefit as possible but unestablished, likely dependent on what the monitoring centre does with the data.
MeSH: Aged · Australia · Defibrillators, Implantable · Europe · Female · Heart Failure · Humans · Israel · Male · Monitoring, Ambulatory · Telemedicine · Treatment Outcome · Automation · Coronary Disease · Safety · Stroke Volume · Adult · Aged, 80 and over · Cardiac Resynchronization Therapy · Cardiac Resynchronization Therapy Devices · Middle Aged · Patient Compliance · Remote Consultation
Cards (4)
Card 1 — IN-TIME 2014: worsened composite clinical score 18.9% (telemonitoring) vs 27.2% (control), OR 0.63, P=0.013; deaths 10 vs 27
Study: Randomised controlled trial (IN-TIME)
Population: 664 patients with chronic heart failure (NYHA class II–III, ejection fraction at most 35%, optimal drug therapy, no permanent AF) with a recent dual-chamber ICD or CRT-D implant at 36 centres in Australia, Europe, and Israel, randomised after a 1-month run-in to automatic daily implant-based multiparameter telemonitoring plus standard care (333) or standard care alone (331). Mean age 65.5 years; mean ejection fraction 26%; 58.7% received CRT-D. Primary outcome at 1 year: a composite clinical score (all-cause death, overnight heart-failure admission, NYHA class change, patient global self-assessment).
Summary: IN-TIME 2014: worsened composite clinical score 18.9% (telemonitoring) vs 27.2% (control), OR 0.63, P=0.013; deaths 10 vs 27
Result: At 1 year, 63 of 333 telemonitoring patients (18.9%) versus 90 of 331 controls (27.2%) had a worsened composite score (P=0.013; odds ratio 0.63; 95% CI 0.43 to 0.90). Ten patients in the telemonitoring group versus 27 in the control group died during follow-up. The authors conclude daily multiparameter telemonitoring can significantly improve clinical outcomes in heart failure and is feasible for clinical practice. The trial was funded by the device manufacturer (Biotronik).
Watch out: The mortality split (10 vs 27) is a secondary observation inside a composite-score primary endpoint — treat it as a signal, not a finding. Single-manufacturer system with a staffed response pathway; the intervention is the pathway as much as the data feed (contrast REM-HF below).
Source: Hindricks G et al. Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet. 2014. PMID: 25131977. Registry ID: NCT00538356, from the record. MeSH: Aged · Australia · Defibrillators, Implantable · Europe · Female · Heart Failure · Humans · Israel · Male · Monitoring, Ambulatory · Telemedicine · Treatment Outcome
Card 2 — TRUST 2010: home monitoring cut in-hospital evaluations by 45% and time-to-evaluation from 36 days to under 2 — morbidity unchanged
Study: Randomised controlled trial (TRUST)
Population: 1,339 ICD patients randomised 2:1 to automatic daily home monitoring (HM) or conventional in-office follow-up, with checks at 3, 6, 9, 12, and 15 months (HM patients used remote checks at 6, 9, and 12 months, with office visits only if needed). Mean age about 64; ejection fraction about 29%; 72–74% primary prevention. Outcomes: scheduled and unscheduled evaluations, morbidity, and time from event occurrence to physician evaluation.
Summary: TRUST 2010: home monitoring cut in-hospital evaluations by 45% and time-to-evaluation from 36 days to under 2 — morbidity unchanged
Result: HM reduced total in-hospital device evaluations by 45% without affecting morbidity. In the HM group, 85.8% of all 6-, 9-, and 12-month follow-ups were performed remotely only. Median time to evaluation was under 2 days with HM versus 36 days with conventional follow-up (P<0.001) for all arrhythmic events. The authors conclude HM is safe and enables more rapid detection of actionable events than conventional monitoring.
Watch out: A logistics and safety trial by design — no outcome hypothesis was tested, so TRUST must not be cited for clinical benefit. Its MeSH record contains indexing artefacts (e.g., 'Apoptosis'); MeSH on this card is trimmed to the relevant terms.
Source: Varma N et al. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation. 2010. PMID: 20625110. Registry ID: NCT00336284, from the record. MeSH: Aged · Automation · Coronary Disease · Defibrillators, Implantable · Female · Follow-Up Studies · Heart Failure · Humans · Male · Middle Aged · Safety · Stroke Volume
Card 3 — REM-HF 2017: 1,650 patients — death or unplanned cardiovascular hospitalisation 42.4% vs 40.8% (HR 1.01); no endpoint differed
Study: Randomised controlled trial (REM-HF)
Population: 1,650 patients with heart failure and a cardiac implanted electronic device at nine centres in England, randomised (2011–2014) to active remote monitoring (weekly downloads with formalised remote follow-up protocols) or usual care. Mean age 70 years; 86% men; median follow-up 2.8 years. Primary endpoint: first event of death from any cause or unplanned hospitalisation for cardiovascular reasons.
Summary: REM-HF 2017: 1,650 patients — death or unplanned cardiovascular hospitalisation 42.4% vs 40.8% (HR 1.01); no endpoint differed
Result: The primary endpoint occurred in 42.4% of the active-monitoring group and 40.8% of the usual-care group (hazard ratio 1.01; 95% CI 0.87 to 1.18; P=0.87). No secondary endpoint — all-cause death, cardiovascular death, or unplanned hospitalisation — differed significantly. Adherence was high (dropout 4.3%). The authors conclude remote monitoring with weekly downloads and a formalised follow-up approach does not improve outcomes in this population.
Watch out: REM-HF used weekly downloads across manufacturers, not daily multiparameter alerts — its null constrains low-intensity monitoring most directly. It is the largest of the three trials and the only one powered as a pure outcome trial.
Source: Morgan JM et al. Remote management of heart failure using implantable electronic devices. Eur Heart J. 2017. PMID: 28575235. Registry ID: ISRCTN96536028, from the record. MeSH: Adult · Aged · Aged, 80 and over · Cardiac Resynchronization Therapy · Cardiac Resynchronization Therapy Devices · Defibrillators, Implantable · Female · Heart Failure · Humans · Male · Middle Aged · Monitoring, Ambulatory · Patient Compliance · Remote Consultation
Card 4 — Synthesis: logistics proven, outcomes possible — monitoring intensity and response pathway decide — POSSIBLE
Study: Synthesis
Population: 3,653 patients across the three trials (Hindricks 2014; Varma 2010; Morgan 2017).
Summary: Synthesis: logistics proven, outcomes possible — monitoring intensity and response pathway decide — POSSIBLE
Result: Faster evaluation (<2 vs 36 days) and 45% fewer in-hospital evaluations (TRUST); one positive composite trial with a mortality signal (IN-TIME); one large purely null outcome trial (REM-HF). POSSIBLE.
Watch out: Presenting any single one of these trials alone would mislead in both directions — the shelf's value is holding all three together.
Source: Synthesis of Hindricks G et al. Lancet. 2014 (PMID: 25131977), Varma N et al. Circulation. 2010 (PMID: 20625110), and Morgan JM et al. Eur Heart J. 2017 (PMID: 28575235). Registry IDs as on the trial cards.