Failing on right-ventricular pacing: upgrading to CRT-D cut death or heart-failure hospitalisation from 32% to 10% (BUDAPEST-CRT Upgrade)
Shelf #2296 · Heart, vessels, blood pressure & clotting
Stamped: development phase, October–November 2026 — exact date not recorded.
Evidence published: 2023.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In heart-failure patients with reduced ejection fraction, a high right-ventricular pacing burden, and a wide paced QRS, does upgrading to cardiac resynchronisation therapy with defibrillator beat continuing with an ICD?
Balance of the evidence
The largest effect size in the devices literature this year — read it with its composite in view. BUDAPEST-CRT Upgrade randomised 360 symptomatic HFrEF patients (NYHA class II–IVa) with an existing pacemaker or ICD, right-ventricular pacing burden of at least 20%, and paced QRS of at least 150 ms, to CRT-D upgrade (215) or ICD (145). The primary outcome — all-cause mortality, heart-failure hospitalisation, or failure to reduce left ventricular end-systolic volume by at least 15% at 12 months — occurred in 32.4% of the CRT-D group versus 78.9% of the ICD group (odds ratio 0.11; 95% CI 0.06 to 0.19; P<.001). That composite is dominated by the remodelling term (most unchanged ventricles 'fail' it), so the clinically decisive figure is the hard secondary: death or heart-failure hospitalisation occurred in 10% versus 32% (hazard ratio 0.27; 95% CI 0.16 to 0.47; P<.001). Procedure- or device-related complications were similar (12.3% vs 7.8%). LIKELY (Tier C): in a paced patient whose ventricle has failed under a heavy right-ventricular burden, upgrade to resynchronisation — the hard-endpoint separation is too large to be a remodelling artefact. The shelf quotes both numbers so readers can weight them.
MeSH: Humans · Heart Failure · Male · Female · Cardiac Resynchronization Therapy · Aged · Defibrillators, Implantable · Stroke Volume · Treatment Outcome · Middle Aged · Hospitalization
Cards (2)
Card 1 — BUDAPEST-CRT Upgrade 2023: primary composite 32.4% vs 78.9% (OR 0.11); death or heart-failure hospitalisation 10% vs 32% (HR 0.27)
Study: Randomised controlled trial (BUDAPEST-CRT Upgrade)
Population: 360 symptomatic HFrEF patients (NYHA class II–IVa) with a pacemaker or ICD, right-ventricular pacing burden of at least 20%, and paced QRS duration of at least 150 ms, randomised 3:2 at multiple centres to CRT-D upgrade (215) or ICD (145). Primary outcome: composite of all-cause mortality, heart failure hospitalisation, or less than 15% reduction in left ventricular end-systolic volume at 12 months (assessed in 179 and 128 patients respectively); secondary outcomes included all-cause mortality or heart failure hospitalisation. Median follow-up 12.4 months.
Summary: BUDAPEST-CRT Upgrade 2023: primary composite 32.4% vs 78.9% (OR 0.11); death or heart-failure hospitalisation 10% vs 32% (HR 0.27)
Result: The primary outcome occurred in 58 of 179 patients (32.4%) in the CRT-D arm versus 101 of 128 (78.9%) in the ICD arm (odds ratio 0.11; 95% CI 0.06 to 0.19; P<.001). All-cause mortality or heart failure hospitalisation occurred in 22 of 215 (10%) versus 46 of 145 (32%) (hazard ratio 0.27; 95% CI 0.16 to 0.47; P<.001). Procedure- or device-related complications were similar (25 of 211 [12.3%] vs 11 of 142 [7.8%]). The authors conclude CRT-D upgrade reduced the combined risk of mortality, heart-failure hospitalisation, and absence of reverse remodelling in this population.
Watch out: The 78.9% control rate on the primary composite is mostly failure-to-remodel, an endpoint resynchronisation is mechanically favoured to win; the HR 0.27 hard-endpoint result is the figure to cite. One-year median follow-up; registry ID was not printed in the fetched record and is recorded as none rather than reconstructed.
Source: Merkely B et al. Upgrade of right ventricular pacing to cardiac resynchronization therapy in heart failure: a randomized trial. Eur Heart J. 2023. PMID: 37632437. Registry ID: none in the fetched record. MeSH: Humans · Heart Failure · Male · Female · Cardiac Resynchronization Therapy · Aged · Defibrillators, Implantable · Stroke Volume · Treatment Outcome · Middle Aged · Hospitalization
Card 2 — Synthesis: upgrade the failing paced ventricle — the hard endpoints separate widely — LIKELY
Study: Synthesis
Population: 360 randomised patients (Merkely 2023).
Summary: Synthesis: upgrade the failing paced ventricle — the hard endpoints separate widely — LIKELY
Result: Death or heart-failure hospitalisation 10% vs 32% (HR 0.27 relative); primary composite 32.4% vs 78.9%. LIKELY.
Watch out: Complements BLOCK HF (Shelf #2295): de novo biventricular pacing prevents the problem; this trial shows upgrading treats it once established.
Source: Synthesis of Merkely B et al. Eur Heart J. 2023 (PMID: 37632437).