Resynchronisation by the conduction system: left-bundle-branch pacing beat biventricular pacing on death or heart-failure hospitalisation in one randomised trial (HeartSync-LBBP)

Shelf #2300 · Heart, vessels, blood pressure & clotting

Stamped: development phase, October–November 2026 — exact date not recorded.

Evidence published: 2026.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In heart failure with reduced ejection fraction and left bundle-branch block, does left-bundle-branch pacing achieve better long-term outcomes than conventional biventricular pacing for cardiac resynchronisation?

Balance of the evidence

One randomised trial with a startlingly large effect — carded with its size and setting in plain view. HeartSync-LBBP randomised 200 patients with ejection fraction of 35% or less and left bundle-branch block at six centres in China to left-bundle-branch pacing (LBBP) or biventricular pacing (BiVP), with a median follow-up of 36 months. The primary endpoint (death from any cause or heart-failure hospitalisation) occurred in 8% with LBBP versus 28% with BiVP (hazard ratio 0.26; 95% CI 0.12 to 0.57; P<.001), driven by hospitalisation (7.0% vs 28.0%; HR 0.23) — mortality alone did not differ significantly (2.0% vs 5.0%; HR 0.40; 95% CI 0.08 to 2.04). Implant success was 98% versus 94%; echocardiographic response rates were similar (86% vs 81%), while super-response was more frequent with LBBP (55% vs 36%; P=.007). The earlier pilot, His-SYNC (41 patients, His-bundle pacing versus BiVP), is named here but NOT carded: its PubMed record is a two-page report carrying no abstract, so no figures from it are quoted anywhere on this shelf. POSSIBLE (Tier C): conduction-system pacing is the most promising challenger biventricular pacing has had — one 200-patient, single-country trial with an effect this size needs independent replication before stronger language; the authors themselves conclude that further trials are warranted.

MeSH: Aged · Female · Humans · Male · Middle Aged · Bundle-Branch Block · Cardiac Pacing, Artificial · Cardiac Resynchronization Therapy · China · Echocardiography · Heart Failure · Prospective Studies · Stroke Volume · Treatment Outcome · Ventricular Function, Left

Cards (2)

Card 1 — HeartSync-LBBP 2026: death or heart-failure hospitalisation 8% (LBBP) vs 28% (BiVP) — HR 0.26; hospitalisation 7% vs 28%; mortality alone not significant

Study: Randomised clinical trial (HeartSync-LBBP)

Population: 200 patients with left ventricular ejection fraction of 35% or less and left bundle-branch block at six centres in China (enrolled October 2020 to March 2022; 136 men, 64 women), randomised 1:1 to left-bundle-branch pacing or biventricular pacing. Primary endpoint: time to death from any cause or heart failure hospitalisation. Secondaries: all-cause death, heart-failure hospitalisation, echocardiographic response (LVEF increase of at least 5 points), and super-response (increase of at least 15 points or LVEF reaching at least 50%). Median follow-up 36 months.

Summary: HeartSync-LBBP 2026: death or heart-failure hospitalisation 8% (LBBP) vs 28% (BiVP) — HR 0.26; hospitalisation 7% vs 28%; mortality alone not significant

Result: Implant success: 98% (LBBP) versus 94% (BiVP) (P=.28). The primary endpoint was significantly lower with LBBP (8% vs 28%; hazard ratio 0.26; 95% CI 0.12 to 0.57; P<.001). All-cause mortality did not differ significantly (2.0% vs 5.0%; HR 0.40; 95% CI 0.08 to 2.04; P=.25); heart-failure hospitalisation was lower with LBBP (7.0% vs 28.0%; HR 0.23; 95% CI 0.10 to 0.52; P<.001). Echocardiographic response rates were similar (86.0% vs 81.0%; P=.34); super-response was higher with LBBP (55.0% vs 36.0%; P=.007). The authors conclude LBBP was superior on the composite in this population and that further trials are warranted.

Watch out: A single 200-patient trial from one country, with a composite driven by hospitalisation and an effect size (HR 0.26) larger than resynchronisation itself achieved against medical therapy — exactly the profile that demands replication. The His-SYNC pilot (PMID 31078637) is not carded: its PubMed record carries no abstract, so this shelf quotes no figures from it.

Source: Chen X et al. Long-Term Outcomes of Left Bundle-Branch Pacing vs Biventricular Pacing in Heart Failure: The HeartSync-LBBP Randomized Clinical Trial. JAMA Cardiol. 2026. PMID: 41811342. Registry ID: ChiCTR2000036554, from the record. MeSH: Aged · Female · Humans · Male · Middle Aged · Bundle-Branch Block · Cardiac Pacing, Artificial · Cardiac Resynchronization Therapy · China · Echocardiography · Heart Failure · Prospective Studies · Stroke Volume · Treatment Outcome · Ventricular Function, Left

Card 2 — Synthesis: a challenger with one strong trial — POSSIBLE

Study: Synthesis

Population: 200 patients with HFrEF and left bundle-branch block (Chen 2026).

Summary: Synthesis: a challenger with one strong trial — POSSIBLE

Result: Death or heart-failure hospitalisation 8% vs 28% (HR 0.26 relative); mortality alone not significant. POSSIBLE.

Watch out: Watch this space: this is the devices question most likely to be re-litigated by larger trials in the next few years; the shelf should be revisited when they report.

Source: Synthesis of Chen X et al. JAMA Cardiol. 2026 (PMID: 41811342). Registry ID as on the trial card.