SourceMed · Shelf #2301 in the ask interface
Recurrent reflex syncope with a cardioinhibitory response: closed-loop-stimulation pacing cut recurrences in two blinded trials (SPAIN + BIOSync CLS)
Shelf #2301 · Heart, vessels, blood pressure & clotting
Stamped: development phase, October–November 2026 — exact date not recorded.
Evidence published: 2017–2021.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In patients aged 40 or older with recurrent reflex (vasovagal) syncope and documented cardioinhibition or tilt-induced asystole, does dual-chamber pacing with closed-loop stimulation reduce syncope recurrence versus sham or switched-off pacing?
Balance of the evidence
Pacing for vasovagal syncope failed for decades in unselected patients; these two trials show what changed — selecting the cardioinhibitory phenotype and using a sensor that reacts to the faint itself. Closed-loop stimulation (CLS) adjusts rate from intracardiac impedance, rising as the vasovagal cascade begins. SPAIN (46 patients aged at least 40 with high-burden syncope and a cardioinhibitory tilt test; randomised crossover between CLS pacing and sham pacing) found at least a 50% reduction in syncope episodes in 72% of patients during CLS periods versus 28% during sham (P=0.017); events occurred in 4 patients (8.7%) on CLS versus 21 (45.7%) on sham. BIOSync CLS (127 patients aged at least 40 with severe recurrent reflex syncope and tilt-induced asystole over 3 seconds; pacing ON versus OFF, double-blind) found syncope recurrence in 16% versus 53% over a median 11.2 months (hazard ratio 0.23; P=0.00005), with estimated 2-year recurrence 22% versus 68%, and minor device-related adverse events in 4%. LIKELY: in this narrow, test-selected phenotype — recurrent, severe, cardioinhibitory — CLS pacing works. The shelf's boundary is the phenotype: these results say nothing about pacing unselected fainters, the population in which pacing earned its poor reputation.
MeSH: Adult · Bradycardia · Cardiac Resynchronization Therapy · Female · Humans · Male · Middle Aged · Recurrence · Severity of Illness Index · Syncope, Vasovagal · Tilt-Table Test · Treatment Outcome · Cardiac Pacing, Artificial · Heart Arrest · Reflex · Syncope
Cards (3)
Card 1 — SPAIN 2017: syncope events in 8.7% of patients during CLS pacing vs 45.7% during sham; at least 50% burden reduction in 72% vs 28%
Study: Randomised double-blind crossover trial (SPAIN)
Population: 46 patients (mean age 56.3; 47.8% men) in Canada and Spain, aged at least 40, with high-burden vasovagal syncope (at least 5 lifetime episodes, at least 2 in the past year; mean 12 previous episodes) and a cardioinhibitory head-up tilt test (bradycardia under 40/min for 10 s or asystole over 3 s), randomised to dual-chamber CLS pacing for 12 months then sham pacing (DDI at 30/min) for 12 months, or the reverse order, crossing over early if 3 syncopes occurred within a month.
Summary: SPAIN 2017: syncope events in 8.7% of patients during CLS pacing vs 45.7% during sham; at least 50% burden reduction in 72% vs 28%
Result: The proportion of patients with at least a 50% reduction in syncopal episodes was 72% (95% CI 47 to 90) with DDD-CLS versus 28% (95% CI 9.7 to 53.5) with sham (P=0.017). Four patients (8.7%) had events during DDD-CLS and 21 (45.7%) during sham (hazard ratio 6.7; 95% CI 2.3 to 19.8 for sham versus CLS periods). Time to first syncope was significantly longer with CLS (29.2 vs 9.3 months; P<0.016). The authors conclude DDD-CLS significantly reduced syncope burden and prolonged time to first recurrence versus sham pacing in tilt-positive cardioinhibitory vasovagal syncope.
Watch out: A crossover of 46 highly selected patients; the hazard ratio is reported for sham versus CLS periods, so this card leads with the absolute rates. Tilt-positive cardioinhibitory patients only — the selection is the intervention's other half.
Source: Baron-Esquivias G et al. Dual-Chamber Pacing With Closed Loop Stimulation in Recurrent Reflex Vasovagal Syncope: The SPAIN Study. J Am Coll Cardiol. 2017. PMID: 28958328. Registry ID: NCT01621464, from the record. MeSH: Adult · Bradycardia · Cardiac Resynchronization Therapy · Female · Humans · Male · Middle Aged · Recurrence · Severity of Illness Index · Syncope, Vasovagal · Tilt-Table Test · Treatment Outcome
Card 2 — BIOSync CLS 2021: syncope recurrence 16% (pacing ON) vs 53% (OFF) — HR 0.23; estimated 2-year recurrence 22% vs 68%
Study: Randomised double-blind trial (BIOSync CLS)
Population: 127 patients aged at least 40 with at least two episodes of unpredictable severe reflex syncope in the previous year and tilt-induced syncope with an asystolic pause longer than 3 seconds, randomised to an active dual-chamber pacemaker with closed-loop stimulation (63) or the same device inactive (64); patients and outcome assessors were blinded. Primary endpoint: time to first recurrence of syncope; median follow-up 11.2 months.
Summary: BIOSync CLS 2021: syncope recurrence 16% (pacing ON) vs 53% (OFF) — HR 0.23; estimated 2-year recurrence 22% vs 68%
Result: Syncope recurred in 10 of 63 patients (16%) in the pacing group versus 34 of 64 (53%) in the control group (hazard ratio 0.23; P=0.00005). Estimated recurrence at 1 year: 19% versus 53%; at 2 years: 22% versus 68%. The combined endpoint of syncope or presyncope also favoured pacing (37% vs 63%; HR 0.44; P=0.002). Minor device-related adverse events were reported in five patients (4%). The authors conclude CLS pacing is highly effective in this population and support tilt testing to select candidates.
Watch out: Benefit is confined to the tilt-induced-asystole phenotype; the trial does not rehabilitate pacing for syncope generally. Device implantation in a condition that is benign in prognosis terms is a values judgement the numbers alone cannot make.
Source: Brignole M et al. Cardiac pacing in severe recurrent reflex syncope and tilt-induced asystole. Eur Heart J. 2021. PMID: 33279955. Registry ID: NCT02324920, from the record. MeSH: Adult · Cardiac Pacing, Artificial · Heart Arrest · Humans · Reflex · Syncope · Tilt-Table Test · Treatment Outcome
Card 3 — Synthesis: in the cardioinhibitory phenotype, CLS pacing prevents most recurrences — LIKELY
Study: Synthesis
Population: 46 patients in SPAIN and 127 in BIOSync CLS (Baron-Esquivias 2017; Brignole 2021).
Summary: Synthesis: in the cardioinhibitory phenotype, CLS pacing prevents most recurrences — LIKELY
Result: Recurrence roughly 9–16% with CLS pacing versus 46–53% without, across two blinded designs. LIKELY.
Watch out: Pairs with ISSUE-3 (Shelf #2302): two different selection strategies (tilt-induced asystole here; loop-recorder-documented asystole there) converging on the same lesson — phenotype first.
Source: Synthesis of Baron-Esquivias G et al. J Am Coll Cardiol. 2017 (PMID: 28958328) and Brignole M et al. Eur Heart J. 2021 (PMID: 33279955). Registry IDs as on the trial cards.