SourceMed · Shelf #2617 in the ask interface
COSIRA: the coronary sinus Reducer doubled the chance of a two-class angina improvement over a sham procedure (35% versus 15%) — in 104 patients, with no gain in exercise time
Shelf #2617 · Heart, vessels, blood pressure & clotting
POSSIBLE (benefit)
Stamped: 2026-11-28.
Evidence published: 2015–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
In refractory angina, does a coronary sinus reducer improve symptoms?
Balance of the evidence
The coronary sinus Reducer (Neovasc) is a balloon-expandable, hourglass-shaped implant that narrows the coronary sinus, intended to redistribute blood toward ischaemic myocardium in patients with refractory angina who have no revascularisation option. Its evidence turns on one small sham-controlled trial, COSIRA (Verheye 2015, PMID: 25651246): 104 patients with class III–IV angina and myocardial ischaemia, not candidates for revascularisation, were randomised to the device or a sham procedure. At 6 months, 35% of the device group versus 15% of the sham group had improved by at least two angina classes (p = 0.02), and 71% versus 42% by at least one class (p = 0.003); quality-of-life scores also favoured the device (Seattle Angina Questionnaire improvement 17.6 versus 7.6 points). But the objective measures did not move — no significant difference in exercise time or in dobutamine-echocardiography wall-motion index — and a sham arm in which 42% of patients improve by a class or more is a standing warning about expectation effects in angina. The wider literature pooled by Hochstadt (2022, PMID: 34968714) — 9 prospective studies, 846 patients, all but one uncontrolled — reports 76% improving by at least one class with 98% procedural success and 3% non-major complications; consistent with COSIRA, but uncontrolled series cannot separate device effect from placebo effect and natural history.
MeSH: Adult · Aged · Aged, 80 and over · Angina Pectoris/classification/therapy · Cardiac Catheterization · Cardiac Catheters · Coronary Angiography · Coronary Sinus/diagnostic imaging · Equipment Design · Female · Humans · Male · Middle Aged · Myocardial Revascularization/instrumentation · Patient Acuity · Quality of Life · Surgical Mesh
Cards (3)
Card 1 — COSIRA (Verheye 2015): improvement of at least two angina classes in 35% versus 15% with sham (p = 0.02); at least one class in 71% versus 42% — but exercise time and wall motion did not improve
Study: Randomised sham-controlled trial (COSIRA, Verheye 2015)
Population: 104 patients with Canadian Cardiovascular Society class III or IV angina and myocardial ischaemia who were not candidates for revascularisation, randomised to implantation of the coronary-sinus reducing device (the Reducer, Neovasc) or to a sham procedure. Primary endpoint: proportion improving by at least two CCS angina classes at 6 months. The trial was funded by Neovasc, the device manufacturer.
Summary: COSIRA (Verheye 2015): improvement of at least two angina classes in 35% versus 15% with sham (p = 0.02); at least one class in 71% versus 42% — but exercise time and wall motion did not improve
Result: Improvement of at least two CCS classes at 6 months: 35% (18/52) with the device versus 15% (8/52) with sham (p = 0.02) — about 20 percentage points absolute. Improvement of at least one class: 71% (37/52) versus 42% (22/52, p = 0.003). Seattle Angina Questionnaire quality-of-life improvement: 17.6 versus 7.6 points on a 100-point scale (p = 0.03). There were no significant between-group differences in exercise time or in dobutamine-echocardiography wall-motion index. At 6 months the device group had had 1 myocardial infarction; the sham group 1 death and 3 myocardial infarctions (numbers far too small to interpret).
Watch out: A 104-patient, manufacturer-funded trial with a subjective primary endpoint and a strikingly responsive sham arm (42% improved at least one class): the between-group difference is real by design, but its size should be discounted for expectation effects, and the failure of exercise time and wall motion to improve means the mechanism's objective footprint is unproven. Nothing here speaks to mortality, infarction, or durability beyond 6 months.
Source: Verheye S et al. Efficacy of a device to narrow the coronary sinus in refractory angina. The New England Journal of Medicine. 2015. PMID: 25651246. Registry ID: NCT01205893. MeSH: Adult · Aged · Aged, 80 and over · Angina Pectoris/classification/therapy · Cardiac Catheterization · Cardiac Catheters · Coronary Angiography · Coronary Sinus/diagnostic imaging · Equipment Design · Female · Humans · Male · Middle Aged · Myocardial Revascularization/instrumentation · Patient Acuity · Quality of Life · Surgical Mesh
Card 2 — Hochstadt 2022 meta-analysis: across 9 prospective studies and 846 patients, 76% improved by at least one angina class and 40% by two — nearly all of it uncontrolled
Study: Systematic review and meta-analysis of prospective studies (Hochstadt 2022)
Population: A meta-analysis of prospective studies of coronary sinus narrowing for refractory angina published to June 2021: 9 studies including 846 patients — the single randomised sham-controlled trial (COSIRA) plus multiple uncontrolled prospective series. Outcomes: proportions improving by at least one or at least two Canadian Cardiovascular Society angina classes, procedural success, complications, Seattle Angina Questionnaire scores and 6-minute walk distance.
Summary: Hochstadt 2022 meta-analysis: across 9 prospective studies and 846 patients, 76% improved by at least one angina class and 40% by two — nearly all of it uncontrolled
Result: Improvement of at least one CCS class: 76% of patients (95% CI 73% to 80%). Improvement of at least two classes: 40% (95% CI 35% to 46%). Procedural success 98%, with no major and 3% non-major periprocedural complications. Seattle Angina Questionnaire scores and 6-minute walk distance improved significantly after implantation in the pooled before-and-after data.
Watch out: Proportions from uncontrolled series are not treatment effects: COSIRA's sham arm improved by at least one class in 42% of patients, so a pooled 76% improvement rate in uncontrolled data is compatible with a much smaller true device effect plus placebo response, regression to the mean, and publication selection. Use this card for procedural safety and consistency, not for efficacy.
Source: Hochstadt A et al. Effectiveness of Coronary Sinus Reducer for Treatment of Refractory Angina: A Meta-analysis. The Canadian Journal of Cardiology. 2022. PMID: 34968714. Registry ID: none in the record. MeSH: Angina Pectoris/diagnosis/physiopathology/surgery · Cardiac Catheterization/methods · Coronary Sinus/physiopathology/surgery · Equipment Design · Humans · Intractable Pain/physiopathology · Prosthesis Implantation/adverse effects/instrumentation/methods · Stents · Treatment Outcome
Card 3 — Synthesis: POSSIBLE — a genuine sham-controlled symptom benefit in a small trial, of uncertain size, with objective endpoints unmoved
Study: Synthesis
Population: One randomised sham-controlled trial (n = 104) and one meta-analysis of 9 prospective studies (846 patients, predominantly uncontrolled).
Summary: Synthesis: POSSIBLE — a genuine sham-controlled symptom benefit in a small trial, of uncertain size, with objective endpoints unmoved
Result: Sham-controlled: at least two angina classes better in 35% versus 15%; at least one class in 71% versus 42%; quality of life better by about 10 points more on a 100-point scale; exercise time and wall-motion index not significantly different. Uncontrolled pooled: 76% improve at least one class; procedural success 98%; non-major complications 3%.
Watch out: For 'no-option' patients the relevant comparison is often device versus continued medical therapy and time; this evidence supports offering the device within that conversation, with the size of the benefit honestly hedged. It does not support use in patients who still have revascularisation or drug options untried.
Source: Synthesis of Verheye S et al. The New England Journal of Medicine. 2015 (PMID: 25651246); Hochstadt A et al. The Canadian Journal of Cardiology. 2022 (PMID: 34968714).