In non-valvular atrial fibrillation, does closing the left atrial appendage prevent stroke as well as long-term anticoagulation — and at what procedural price?

Balance of the evidence

Three randomized trials, one consistent answer on efficacy and one consistent price. Against warfarin, Watchman closure was noninferior for the composite of stroke, systemic embolism, and cardiovascular death in PROTECT-AF and PREVAIL — with more early safety events, mostly periprocedural. Against a modern DOAC, PRAGUE-17 again met noninferiority for its broad net-benefit composite (sHR 0.84; P=0.004 for noninferiority) in high-risk patients, with major device complications in 4.5%. None of the three shows closure prevents strokes better than anticoagulation; the case for the device is avoiding lifelong blood thinners in selected patients, bought with an upfront procedural risk. The honest label is: LIKELY — for noninferiority, not superiority.

MeSH: Atrial Fibrillation · Atrial Appendage · Warfarin · Factor Xa Inhibitors · Stroke · Humans

Cards (3)

Card 1 — PROTECT-AF (Holmes 2009): LAA closure noninferior to warfarin — efficacy 3.0 vs 4.9 events per 100 patient-years, safety events higher

Study: Randomized non-inferiority trial (PROTECT-AF), Lancet 2009

Population: 707 patients with non-valvular AF and at least one of: previous stroke or TIA, congestive heart failure, diabetes, hypertension, or age 75 or older (mean age 72 years; 30% women, per the trial's registry record); LAA closure with the Watchman device (463) vs warfarin (244), 2:1 randomization; 1,065 patient-years of follow-up.

Summary: Primary efficacy rate 3.0 vs 4.9 per 100 patient-years (RR 0.62, 95% CrI 0.35–1.25; probability of noninferiority >99.9%); primary safety events 7.4 vs 4.4 per 100 patient-years (RR 1.69) — 707 patients.

Result: Primary efficacy events: 3.0 per 100 patient-years with closure vs 4.9 per 100 patient-years with warfarin (rate ratio 0.62, 95% credible interval 0.35–1.25) — the probability of noninferiority exceeded 99.9%. Primary safety events (major bleeding, pericardial effusion, device embolization) were more frequent with closure: 7.4 vs 4.4 per 100 patient-years (RR 1.69, 1.01–3.19), mainly periprocedural complications.

Watch out: Bayesian credible intervals and a generous two-fold non-inferiority margin, with an early-generation device on its learning curve — the safety penalty was concentrated around the implant procedure. Event rates are per patient-year, not simple percentages.

Source: Holmes DR et al. Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial. Lancet. 2009. PMID 19683639 · NCT00129545.

Card 2 — PREVAIL (Holmes 2014): safer implants than PROTECT-AF, but the first efficacy endpoint missed its noninferiority criterion

Study: Randomized trial (PREVAIL), J Am Coll Cardiol 2014

Population: 407 patients with non-valvular AF (mean age 74 years; 30% women, per the trial's registry record); Watchman closure (269) vs chronic warfarin (138), 2:1 randomization; coprimary endpoints assessed at 18 months.

Summary: First coprimary endpoint rate 0.064 vs 0.063 (rate ratio 1.07; upper 95% CrI 1.89 — noninferiority not achieved); stroke or systemic embolism after day 7 noninferior; early safety events 2.2% — 407 patients.

Result: First coprimary efficacy endpoint (stroke, systemic embolism, or cardiovascular/unexplained death): rate 0.064 in the device group vs 0.063 in the control group (rate ratio 1.07, 95% credible interval 0.57–1.89) — the prespecified noninferiority criterion was NOT achieved, because the upper credible bound crossed 1.75. The second coprimary endpoint (stroke or systemic embolism more than 7 days after randomization) did achieve noninferiority (0.0253 vs 0.0200). Early safety events occurred in 2.2% of the Watchman arm, meeting the prespecified safety goal; on a broader definition, adverse events remained lower than in PROTECT-AF (4.2% vs 8.7%; P=0.004), and pericardial effusions requiring surgery fell from 1.6% to 0.4%.

Watch out: A trial often summarized as positive that formally missed its first noninferiority endpoint — event rates were low and numerically similar, but with 407 patients the credible interval was too wide to prove the prespecified claim. Read it as supportive of the device, not definitive.

Source: Holmes DR Jr et al. Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. J Am Coll Cardiol. 2014. PMID 24998121 · NCT01182441.

Card 3 — PRAGUE-17 (Osmancik 2020): closure noninferior to a DOAC in high-risk AF — composite 10.99% vs 13.42% per year

Study: Randomized non-inferiority trial (PRAGUE-17), J Am Coll Cardiol 2020

Population: 402 high-risk patients with non-valvular AF (mean CHA2DS2-VASc 4.7 ± 1.5) randomized to LAA closure (201; implantation successful in 90.0%) or DOAC (201; apixaban in 95.5%); median follow-up 19.9 months.

Summary: Primary composite annual rates 10.99% with closure vs 13.42% with DOAC (sHR 0.84, 0.53–1.31; P=0.004 for noninferiority); major device-related complications 4.5% — 402 high-risk patients.

Result: Annual rates of the primary outcome (stroke, TIA, systemic embolism, cardiovascular death, major or nonmajor clinically relevant bleeding, or procedure-/device-related complications): 10.99% with closure vs 13.42% with DOAC (subdistribution HR 0.84, 95% CI 0.53–1.31; P=0.44; P=0.004 for noninferiority). No component differed between groups: all-stroke/TIA sHR 1.00 (0.40–2.51), clinically significant bleeding sHR 0.81 (0.44–1.52), cardiovascular death sHR 0.75 (0.34–1.62). Major LAA-closure-related complications occurred in 9 patients (4.5%).

Watch out: The composite charges device complications to the closure arm and bleeding to both — a fair net-benefit test, but it means 'noninferior' here describes the whole package, not stroke prevention alone. A small trial with wide intervals on every component, and two different closure devices were used.

Source: Osmancik P et al. Left Atrial Appendage Closure Versus Direct Oral Anticoagulants in High-Risk Patients With Atrial Fibrillation. J Am Coll Cardiol. 2020. PMID 32586585 · NCT02426944.