Cardiac resynchronization therapy in HFrEF

Shelf #274 · Heart, vessels, blood pressure & clotting

Stamped: development phase, October–November 2026 — exact date not recorded.

Evidence published: 2004–2010.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In heart failure with a wide QRS complex, does resynchronizing the ventricles keep people alive and out of hospital — and how far down the symptom scale does the benefit reach?

Balance of the evidence

The most consistent device story in heart failure. In severe symptoms, CRT cut the composite of death or major hospitalization by about a third in CARE-HF (39% vs 55%; HR 0.63) — with death itself down from 30% to 20% — and COMPANION's CRT-defibrillator arm reduced mortality by a 36% relative margin. The benefit follows the QRS down the symptom scale but changes character: RAFT (NYHA class II–III) reduced both death or heart-failure hospitalization (33.2% vs 40.3%; HR 0.75) and death alone (HR 0.75), while MADIT-CRT (NYHA I–II) prevented heart-failure events (17.2% vs 25.3%; HR 0.66) with no mortality difference at all. The price is procedural — RAFT saw roughly twice as many 30-day device adverse events with CRT. The honest label is: LIKELY.

MeSH: Cardiac Resynchronization Therapy · Heart Failure · Defibrillators, Implantable · Pacemaker, Artificial · Humans

Cards (4)

Card 1 — CARE-HF (Cleland 2005): resynchronization cut death or cardiovascular hospitalization — 39% vs 55%

Study: Open-label RCT (CARE-HF), N Engl J Med 2005

Population: 813 patients with NYHA class III–IV heart failure, left ventricular systolic dysfunction, and cardiac dyssynchrony, on standard pharmacologic therapy; medical therapy plus cardiac resynchronization vs medical therapy alone; mean follow-up 29.4 months.

Summary: Primary endpoint 39% vs 55% (HR 0.63); death from any cause 20% vs 30% (HR 0.64) — 813 patients with NYHA class III–IV heart failure and cardiac dyssynchrony.

Result: The primary endpoint was reached by 159 patients in the resynchronization group vs 224 in the medical-therapy group ( 39% vs 55% ; HR 0.63, 95% CI 0.51–0.77; P<0.001). Deaths: 82 vs 120 ( 20% vs 30% ; HR 0.64, 0.48–0.85; P<0.002) — a 36% relative reduction in mortality. Resynchronization also reduced interventricular mechanical delay and end-systolic volume, increased ejection fraction, and improved symptoms and quality of life (P<0.01 for all comparisons).

Watch out: Open-label, as device trials must be, with a hospitalization-heavy composite — but the mortality component stands on its own here. Conducted before contemporary quadruple drug therapy, so today's absolute benefit stacks on better background treatment. A pre-registration-era trial: no registry ID exists in its record.

Source: Cleland JGF et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005. PMID 15753115.

Card 2 — COMPANION (Bristow 2004): CRT with a defibrillator cut death by a 36% relative margin; pacing alone missed significance

Study: Open-label RCT (COMPANION), N Engl J Med 2004

Population: 1,520 patients with NYHA class III or IV heart failure due to ischemic or nonischemic cardiomyopathy and QRS at least 120 ms, randomized 1:2:2 to optimal pharmacologic therapy (diuretics, ACE inhibitors, beta-blockers, spironolactone) alone, therapy plus CRT-pacemaker, or therapy plus CRT-pacemaker-defibrillator.

Summary: Death or hospitalization HR 0.81 (CRT-pacemaker) and 0.80 (CRT-defibrillator); risk of death reduced 24% relative (P=0.059) and 36% relative (P=0.003) respectively — 1,520 patients with NYHA class III–IV heart failure.

Result: Against drug therapy alone, CRT-pacemaker reduced the primary composite (time to death from or hospitalization for any cause) with HR 0.81 (P=0.014), and CRT-defibrillator with HR 0.80 (P=0.01). The combined endpoint of death from or hospitalization for heart failure fell by 34% relative in the pacemaker group (P<0.002) and 40% relative in the pacemaker-defibrillator group (P<0.001). Death from any cause: reduced by 24% relative with CRT-pacemaker (P=0.059 — not significant) and by 36% relative with CRT-defibrillator (P=0.003).

Watch out: The trial reports its headline effects as relative reductions (the abstract gives no absolute event rates), so this card quotes them as relative, per the library's rule. The mortality claim that survived significance testing belongs to the defibrillator combination, not to resynchronization pacing alone. A pre-registration-era trial: no registry ID exists in its record.

Source: Bristow MR et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004. PMID 15152059.

Card 3 — RAFT (Tang 2010): adding CRT to an ICD in NYHA class II–III cut death or heart-failure hospitalization — 33.2% vs 40.3%

Study: RCT (RAFT), N Engl J Med 2010

Population: 1,798 patients with NYHA class II or III heart failure, LVEF 30% or less, and an intrinsic QRS of 120 ms or more (or paced QRS of 200 ms or more); ICD plus CRT (894) vs ICD alone (904), both with optimal medical therapy; mean follow-up 40 months.

Summary: Primary outcome 33.2% vs 40.3% (HR 0.75); death 20.8% vs 26.1% (HR 0.75); heart-failure hospitalization 19.5% vs 26.1% (HR 0.68); more device adverse events at 30 days (124 vs 58 patients) — 1,798 patients.

Result: The primary outcome occurred in 297 of 894 patients (33.2%) in the ICD-CRT group vs 364 of 904 (40.3%) in the ICD group — HR 0.75 (95% CI 0.64–0.87; P<0.001). Deaths: 186 vs 236 (20.8% vs 26.1% computed from the reported counts; HR 0.75, 0.62–0.91; P=0.003). Hospitalization for heart failure: 174 vs 236 patients (19.5% vs 26.1%; HR 0.68, 0.56–0.83; P<0.001). At 30 days after implantation, adverse events had occurred in 124 ICD-CRT patients vs 58 ICD patients (P<0.001).

Watch out: The control arm already carried a defibrillator, so RAFT isolates the resynchronization increment — and prices it: the early procedural harm is part of the trade. Percentages for death and heart-failure hospitalization are computed from the event counts in the trial report.

Source: Tang AS et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N Engl J Med. 2010. PMID 21073365 · NCT00251251.

Card 4 — MADIT-CRT (Moss 2009): in NYHA class I–II, CRT-D prevented heart-failure events (17.2% vs 25.3%) but not death

Study: RCT (MADIT-CRT), N Engl J Med 2009

Population: 1,820 patients with ischemic or nonischemic cardiomyopathy, LVEF 30% or less, QRS 130 ms or more, and NYHA class I or II symptoms, enrolled over 4.5 years; CRT-ICD (1,089) vs ICD alone (731), 3:2 randomization; average follow-up 2.4 years.

Summary: Primary endpoint 17.2% vs 25.3% (HR 0.66), driven by a 41% relative reduction in heart-failure events, concentrated at QRS 150 ms or more; no difference in mortality (about 3% per year in both groups) — 1,820 patients.

Result: The primary endpoint (death or a nonfatal heart-failure event) occurred in 187 of 1,089 patients (17.2%) in the CRT-ICD group vs 185 of 731 (25.3%) in the ICD-only group — HR 0.66 (95% CI 0.52–0.84; P=0.001). The superiority was driven by a 41% relative reduction in heart-failure events and was evident primarily in the prespecified subgroup with QRS 150 ms or more. There was no significant difference in the risk of death — about 3% per year in each group. CRT reduced left ventricular volumes and improved ejection fraction.

Watch out: Heart-failure events were diagnosed by physicians who knew the treatment assignments (a blinded committee adjudicated them) — a softer endpoint than death, in the mildest population on this shelf. The honest headline at this end of the symptom scale: CRT prevents decompensations, not deaths.

Source: Moss AJ et al. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med. 2009. PMID 19723701 · NCT00180271.