Small-annulus TAVR: clinical events matched (9.4% versus 10.6%), but valve dysfunction was 9.4% on the self-expanding valve versus 41.6% on the balloon-expandable (SMART)

Shelf #3361 · Heart, vessels, blood pressure & clotting

SPLIT — clinical outcomes noninferior (9.4% versus 10.6%) but bioprosthetic-valve dysfunction far lower with the self-expanding valve (9.4% versus 41.6%) in SMART

Stamped: 2026-11-28.

Evidence published: 2024.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In severe aortic stenosis with a small annulus, does a balloon-expandable valve match a self-expanding valve for valve function and clinical outcomes?

Balance of the evidence

SPLIT — the patients fared the same; the valves did not. SMART (Herrmann HC et al., PMID: 38587261) randomised 716 patients with severe aortic stenosis and an annulus area of 430 mm² or less (mean age 80; 87% women) to a self-expanding supraannular valve or a balloon-expandable valve. The clinical coprimary endpoint — death, disabling stroke, or heart-failure rehospitalisation through 12 months — occurred in an estimated 9.4% with the self-expanding valve versus 10.6% with the balloon-expandable valve (difference −1.2 percentage points; p<0.001 for noninferiority). The valve-function coprimary endpoint split wide open: bioprosthetic-valve dysfunction in 9.4% versus 41.6% (difference −32.2 points; p<0.001 for superiority), mean gradient 7.7 versus 15.7 mmHg, effective orifice area 1.99 versus 1.50 cm², and moderate or severe prosthesis–patient mismatch at 30 days in 11.2% versus 35.3%. SPLIT: balloon-expandable matches on 12-month clinical outcomes and loses decisively on haemodynamics — whether that loss becomes clinical over time is the unanswered part. Tier C: single-trial, industry-funded.

MeSH: Aged · Aged, 80 and over · Female · Humans · Male · Aortic Valve/surgery · Aortic Valve Stenosis/surgery/complications · Bioprosthesis/adverse effects · Heart Failure · Heart Valve Prosthesis · Kaplan-Meier Estimate · Postoperative Complications/etiology · Prosthesis Design · Prosthesis Failure · Stroke/etiology · Transcatheter Aortic Valve Replacement/adverse effects

Cards (2)

Card 1 — Herrmann 2024 (SMART): death, disabling stroke, or heart-failure admission 9.4% versus 10.6% (noninferior) — valve dysfunction 9.4% versus 41.6%, gradient 7.7 versus 15.7 mmHg (superior) for the self-expanding valve

Study: Randomised trial with coprimary non-inferiority and superiority endpoints (Herrmann 2024, SMART)

Population: 716 patients with symptomatic severe aortic stenosis and aortic-valve annulus area ≤430 mm², treated at 83 sites in 13 countries (mean age 80 years; 87% women; mean STS Predicted Risk of Mortality 3.3%), randomised 1:1 to TAVR with a self-expanding supraannular valve or a balloon-expandable valve; endpoints through 12 months. The trial was funded by Medtronic, maker of the self-expanding valve tested.

Summary: Herrmann 2024 (SMART): death, disabling stroke, or heart-failure admission 9.4% versus 10.6% (noninferior) — valve dysfunction 9.4% versus 41.6%, gradient 7.7 versus 15.7 mmHg (superior) for the self-expanding valve

Result: The Kaplan–Meier estimate of death, disabling stroke, or rehospitalisation for heart failure through 12 months was 9.4% with the self-expanding valve and 10.6% with the balloon-expandable valve (difference −1.2 percentage points; 90% CI −4.9 to 2.5; p<0.001 for noninferiority). Bioprosthetic-valve dysfunction occurred in an estimated 9.4% versus 41.6% (difference −32.2 points; 95% CI −38.7 to −25.6; p<0.001 for superiority). Mean gradient at 12 months was 7.7 versus 15.7 mmHg; effective orifice area 1.99 versus 1.50 cm²; haemodynamic structural valve dysfunction 3.5% versus 32.8%; moderate or severe prosthesis–patient mismatch at 30 days 11.2% versus 35.3% (all p<0.001). Major safety endpoints appeared similar.

Watch out: The dysfunction composite is built largely from haemodynamic measures (gradient, mismatch) that favour a supraannular design by construction, in a trial funded by that design's manufacturer — the 32-point gap is real arithmetic on a partly definitional endpoint, while the clinical composite it should predict was identical at 12 months. Whether mismatch at this level shortens valve life or lives is exactly what 12 months cannot show.

Source: Herrmann HC et al. Self-Expanding or Balloon-Expandable TAVR in Patients with a Small Aortic Annulus. The New England Journal of Medicine. 2024. PMID: 38587261. Registry ID: NCT04722250. MeSH: Aged · Aged, 80 and over · Female · Humans · Male · Aortic Valve/surgery · Aortic Valve Stenosis/surgery/complications · Bioprosthesis/adverse effects · Heart Failure · Heart Valve Prosthesis · Kaplan-Meier Estimate · Postoperative Complications/etiology · Prosthesis Design · Prosthesis Failure · Stroke/etiology · Transcatheter Aortic Valve Replacement/adverse effects

Card 2 — Synthesis: pick the self-expanding valve for the numbers on the echo; the patients, at one year, could not tell the difference

Study: Synthesis

Population: One randomised trial (SMART, 716 small-annulus TAVR patients, 87% women, self-expanding supraannular versus balloon-expandable valve, 12-month endpoints).

Summary: Synthesis: pick the self-expanding valve for the numbers on the echo; the patients, at one year, could not tell the difference

Result: Clinical composite: 9.4% versus 10.6% (noninferior). Valve dysfunction: 9.4% versus 41.6%. Gradient: 7.7 versus 15.7 mmHg. Mismatch at 30 days: 11.2% versus 35.3%. SPLIT, Tier C: single-trial, sponsor-funded, surrogate-heavy superiority claim.

Watch out: The population is overwhelmingly women with small annuli — the result does not transfer to average annuli, where the haemodynamic gap narrows. No data here beyond 12 months, and no mortality difference was shown or powered for.

Source: Synthesis of Herrmann HC et al. The New England Journal of Medicine. 2024 (PMID: 38587261).