SourceMed · Shelf #3750 in the ask interface
Ankle syndesmosis injury: suture-button fixation beats screws on function scores (AOFAS ~95 vs ~87) and implant problems (breakage 0% vs 25.4%, removal 6.0% vs 22.4%) in pooled randomised trials
Shelf #3750 · Bones & joints
POSSIBLE (benefit) — two meta-analyses of randomised trials found better AOFAS scores and markedly fewer implant failures, removals and malreductions with suture-button fixation than with syndesmotic screws
Stamped: 2026-11-28.
Evidence published: 2019–2020.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
For acute ankle syndesmosis (diastasis) injuries, does suture-button fixation give better outcomes than syndesmotic screws?
Balance of the evidence
POSSIBLE (benefit) — consistent across two independent meta-analyses of the same small randomised literature. Shimozono 2019 (PMID: 30475639; 5 trials, 285 patients) found suture-button patients scored higher on the American Orthopaedic Foot & Ankle Society scale at a mean of 20.8 months (95.3 vs 86.7; P < 0.001), with no broken implants (0.0% vs 25.4%; P < 0.001), fewer implant removals (6.0% vs 22.4%; P = 0.01) and less joint malreduction (0.8% vs 11.5%; P = 0.05). Onggo 2020 (PMID: 30503612; 5 trials, 280 patients) confirms the pattern with a cooler head: the 1-year AOFAS advantage was a mean difference of 5.46 points (95% CI 0.40 to 10.51; p = 0.03) — statistically significant but, the authors note, of unlikely clinical significance — while implant failure was dramatically lower with the button (odds ratio 0.03, 95% CI 0.01 to 0.15; p < 0.001) and infection/wound problems were marginally, non-significantly higher (odds ratio 1.4, 95% CI 0.4 to 4.85; p = 0.60). POSSIBLE (benefit): the button's clearest wins are mechanical — nothing to break, rarely anything to remove, better reduction — with a small functional edge on top.
MeSH: Ankle Fractures/surgery · Ankle Injuries/surgery · Bone Screws · Device Removal · Female · Fracture Fixation, Internal/adverse effects/instrumentation/methods · Humans · Male · Postoperative Complications · Randomized Controlled Trials as Topic · Return to Sport · Suture Techniques/adverse effects · Sutures
Cards (3)
Card 1 — Shimozono 2019 meta-analysis (5 RCTs, 285 patients): AOFAS 95.3 vs 86.7 (P < 0.001); broken implants 0.0% vs 25.4%; removal 6.0% vs 22.4%; malreduction 0.8% vs 11.5%
Study: Meta-analysis of randomised controlled trials (Shimozono 2019)
Population: 5 randomised controlled trials comparing suture-button (n = 143) with syndesmotic screw fixation (n = 142) for ankle syndesmosis injuries; outcomes included American Orthopaedic Foot & Ankle Society (AOFAS) score at a mean follow-up of 20.8 months, implant breakage, implant removal and joint malreduction.
Summary: Shimozono 2019 meta-analysis (5 RCTs, 285 patients): AOFAS 95.3 vs 86.7 (P < 0.001); broken implants 0.0% vs 25.4%; removal 6.0% vs 22.4%; malreduction 0.8% vs 11.5%
Result: The suture-button group had a higher postoperative AOFAS score (95.3 vs 86.7; P < 0.001) and lower rates of broken implants (0.0% vs 25.4%; P < 0.001), implant removal (6.0% vs 22.4%; P = 0.01) and joint malreduction (0.8% vs 11.5%; P = 0.05) than the screw group. The authors conclude the suture-button technique results in improved functional outcomes and lower rates of broken implant and malreduction, warranting a grade A recommendation in their assessment.
Watch out: Five small trials; an 8.6-point AOFAS gap is larger than the second meta-analysis found at a fixed 1-year time point, and AOFAS is a clinician-completed score vulnerable to unblinded assessment — weight the hard mechanical outcomes (breakage, removal, malreduction) more than the score gap.
Source: Shimozono Y et al. Suture Button Versus Syndesmotic Screw for Syndesmosis Injuries: A Meta-analysis of Randomized Controlled Trials. The American journal of sports medicine. 2019. PMID: 30475639. MeSH: Ankle Fractures/surgery · Ankle Injuries/surgery · Bone Screws · Device Removal · Female · Fracture Fixation, Internal/adverse effects/instrumentation/methods · Humans · Male · Postoperative Complications · Randomized Controlled Trials as Topic · Return to Sport · Suture Techniques/adverse effects · Sutures
Card 2 — Onggo 2020 meta-analysis (5 RCTs, 280 patients): AOFAS mean difference +5.46 at 1 year (p = 0.03); implant failure OR 0.03 (p < 0.001); infection OR 1.4, not significant
Study: Meta-analysis of randomised controlled trials (Onggo 2020)
Population: 5 randomised controlled trials totalling 280 patients (140 suture button, 140 syndesmotic screw) with acute ankle diastasis/syndesmotic injuries, comparing clinical outcomes and complication profiles of the two constructs (searches to March 2018).
Summary: Onggo 2020 meta-analysis (5 RCTs, 280 patients): AOFAS mean difference +5.46 at 1 year (p = 0.03); implant failure OR 0.03 (p < 0.001); infection OR 1.4, not significant
Result: Suture-button fixation gave a significantly higher AOFAS score at 1 year (mean difference 5.46, 95% CI 0.40 to 10.51; p = 0.03) — which the authors judge unlikely to be clinically significant — and a significantly lower implant failure rate (odds ratio 0.03, 95% CI 0.01 to 0.15; p < 0.001). Infection and wound issues were marginally higher with the suture button (odds ratio 1.4, 95% CI 0.4 to 4.85; p = 0.60). No other parameter differed significantly; the authors conclude both constructs yield similar clinical outcomes, with fewer implant failures for the button.
Watch out: The two meta-analyses draw on substantially the same five trials, so this is corroboration, not independent replication; neither record addresses cost (buttons cost more per implant, screws cost more in removals) or outcomes beyond about two years.
Source: Onggo JR et al. Suture button versus syndesmosis screw constructs for acute ankle diastasis injuries: A meta-analysis and systematic review of randomised controlled trials. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2020. PMID: 30503612. MeSH: Ankle Injuries/surgery · Ankle Joint/surgery · Bone Screws · Fracture Fixation, Internal/methods · Humans · Randomized Controlled Trials as Topic · Suture Techniques · Sutures
Card 3 — Synthesis: pick the button for the hardware outcomes; expect similar knees — ankles — either way on symptoms
Study: Synthesis
Population: Two meta-analyses of the same small randomised literature (5 trials each; 285 and 280 patients) comparing suture-button with syndesmotic screw fixation for acute ankle syndesmosis injuries.
Summary: Synthesis: pick the button for the hardware outcomes; expect similar knees — ankles — either way on symptoms
Result: Function: AOFAS higher with the button by 5.46 points at 1 year (0.40–10.51) in one analysis and 95.3 versus 86.7 at ~21 months in the other. Hardware: broken implants 0.0% versus 25.4%; implant failure odds ratio 0.03; removal 6.0% versus 22.4%; malreduction 0.8% versus 11.5%. Infection: no significant difference. POSSIBLE (benefit), driven by implant-related outcomes with a small functional advantage.
Watch out: Syndesmosis injuries accompanying fractures versus isolated ligamentous diastasis may behave differently, and the pooled trials mix them; long-term arthritis outcomes by reduction quality are beyond both records.
Source: Synthesis of Shimozono Y et al. The American journal of sports medicine. 2019 (PMID: 30475639); Onggo JR et al. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2020 (PMID: 30503612).