SourceMed · Shelf #561 in the ask interface
Hybrid closed-loop insulin delivery in type 1 diabetes
Shelf #561 · Diabetes, glucose & metabolism
LIKELY
Stamped: development phase, October–November 2026 — exact date not recorded.
Evidence published: 2018–2022.
Last source check: not yet re-checked since stamping.
Source correction status: not yet checked for source corrections.
Question
Does automated (closed-loop) insulin delivery improve time-in-range versus sensor-augmented or standard pump therapy?
Balance of the evidence
LIKELY: Randomised trials agree that automated insulin delivery improves glycaemic control in type 1 diabetes — hybrid closed-loop raised time-in-range by 10.8 percentage points versus sensor-augmented pump therapy over 12 weeks, and the iLet bionic pancreas lowered HbA1c by an adjusted -0.5% versus standard care (with CGM) at 13 weeks in both adults and youth, with time-in-range up by 11% in adults — without increasing CGM-measured hypoglycaemia. Trials are short (12-13 weeks) and outcomes are glycaemic measures, not complications.
MeSH: Blood Glucose Self-Monitoring · Diabetes Mellitus, Type 1/drug therapy · Glycated Hemoglobin/analysis · Hypoglycemia/prevention & control · Hypoglycemic Agents/administration & dosage · Infusion Pumps, Implantable · Insulin/administration & dosage
Cards (4)
Card 1 — Hybrid closed-loop raised time-in-range by 10.8 percentage points versus sensor-augmented pump therapy
Study: Multicentre randomised controlled trial
Population: 86 patients with type 1 diabetes aged 6 years and older, on insulin pumps with suboptimal control (HbA1c 7.5-10.0%), at 4 UK and 2 US centres (closed-loop n=46, control n=40); the abstract does not state mean age or sex mix.
Summary: Tuber 2018: time in range 65% vs 54% — a 10.8-point advantage [8.2 to 13.5] for day-and-night hybrid closed loop.
Result: Time in the target range 3.9-10.0 mmol/L (≈70-180 mg/dL) was 65% (SD 8) with closed-loop versus 54% (SD 9) with sensor-augmented pump therapy (mean difference in change 10.8 percentage points, 95% CI 8.2 to 13.5; p<0.0001). HbA1c fell to 7.4% (SD 0.6) in the closed-loop group versus 7.7% (SD 0.5) in controls after the intervention (mean difference in change 0.36%, 95% CI 0.19 to 0.53; p<0.0001). Time below 3.9 mmol/L (≈70 mg/dL) and above 10.0 mmol/L (≈180 mg/dL) was shorter with closed-loop. No severe hypoglycaemia occurred; one diabetic ketoacidosis occurred in the closed-loop group due to infusion set failure.
Watch out: Open-label, 12-week trial after a 4-week run-in during which both groups' HbA1c already improved; outcomes are sensor measures, not clinical events.
Source: Tauschmann M et al. Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-week randomised trial. Lancet. 2018 Oct 13;392(10155):1321-1329. PMID: 30292578. Registry ID: NCT02523131, from the record's databank fields.
Card 2 — iLet bionic pancreas (adults): HbA1c down an adjusted 0.5%, time-in-range up 11% versus standard care with CGM
Study: Multicentre randomised controlled trial (pivotal trial, adult report)
Population: 161 adults with type 1 diabetes, 18-79 years old, baseline HbA1c 5.5%-13.1% (bionic pancreas n=107, standard care n=54); before the study 32% used multiple daily injections, 27% a pump without automation, 5% a pump with predictive low-glucose suspend and 36% a hybrid closed-loop system; the abstract does not state sex mix.
Summary: iLet bionic pancreas (adults): adjusted HbA1c difference −0.5% ; time in range up 11% (2.6 h/day) at 13 weeks.
Result: Mean HbA1c fell from 7.6% to 7.1% with the bionic pancreas versus 7.6% to 7.5% with standard care (adjusted difference -0.5%, 95% CI -0.6% to -0.3%, P<0.001) at 13 weeks. Mean time-in-range 70-180 mg/dL increased by 11% (2.6 h/day) and mean CGM glucose fell by 16 mg/dL versus standard care (P<0.001). CGM-measured hypoglycaemia did not differ significantly between groups. Severe hypoglycaemic events occurred in 7 participants (6.5%) in the bionic pancreas group and 2 (1.9%) in the standard care group (P=0.40).
Watch out: The comparator was standard care plus CGM, and over a third of participants were already using hybrid closed-loop before the study, so this is not a comparison against injections alone; 13-week duration.
Source: Kruger D et al. A Multicenter Randomized Trial Evaluating the Insulin-Only Configuration of the Bionic Pancreas in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2022 Oct;24(10):697-711. PMID: 36173236. Registry ID: NCT04200313, from the record's databank fields.
Estimate type: adjusted estimate, as stated in the card.
Card 3 — iLet bionic pancreas (youth 6-17): HbA1c down an adjusted 0.5%, largest gains at high baseline HbA1c
Study: Multicentre randomised controlled trial (pivotal trial, youth report — same trial as the adult report)
Population: 165 youth with type 1 diabetes, 6-17 years old, baseline HbA1c 5.8%-12.2% (bionic pancreas n=112, standard care n=53); the abstract does not state sex mix.
Summary: iLet (youth 6–17): adjusted HbA1c difference −0.5% , with the largest gains at high baseline HbA1c.
Result: Mean HbA1c fell from 8.1% to 7.5% with the bionic pancreas versus 7.8% at both baseline and 13 weeks with standard care (adjusted difference -0.5%, 95% CI -0.7% to -0.2%, P<0.001). In participants with baseline HbA1c 9.0% or higher (n=34), mean HbA1c fell from 9.7% to 7.9% with the bionic pancreas versus 9.7% to 9.8% with standard care. Mean time-in-range increased by 10% (2.4 h/day) versus standard care (P<0.001). A severe hypoglycaemia event occurred in 3 participants (2.7%) in the bionic pancreas group and 1 (1.9%) in the standard care group.
Watch out: Same pivotal trial as the preceding card (NCT04200313 counts once); 13 weeks only, and the control group also used real-time CGM.
Source: Messer LH et al. Positive Impact of the Bionic Pancreas on Diabetes Control in Youth 6-17 Years Old with Type 1 Diabetes: A Multicenter Trial. Diabetes Technol Ther. 2022 Oct;24(10):712-725. PMID: 36173237. Registry ID: NCT04200313, from the record's databank fields.
Estimate type: adjusted estimate, as stated in the card.
Card 4 — Synthesis: automated delivery improves time-in-range and HbA1c without more sensor-measured hypoglycaemia
Study: Synthesis
Population: Synthesis of the cards above: Tauschmann hybrid closed-loop trial (86 participants, age 6+) and the iLet pivotal trial (326 participants across its adult and youth reports, one trial).
Summary: Automated delivery adds ~11 points of time-in-range and ~0.5% HbA1c without more sensor-measured hypoglycaemia — LIKELY.
Result: Across both programmes, automation added roughly 11 percentage points of time-in-range and about 0.4-0.5% of HbA1c reduction over 12-13 weeks versus sensor-augmented or standard care with CGM, with no increase in CGM-measured hypoglycaemia. This shelf is stamped LIKELY: two independent randomised programmes agree on direction and size. It is distinct from #560 (monitoring only) and #564 (pumps with no automation).
Watch out: All follow-up is 12-13 weeks; nothing here speaks to years-long use, device burden, or complication outcomes, and severe hypoglycaemia event counts were small and numerically higher in the iLet arm.
Source: Synthesis of Tauschmann M et al. Lancet. 2018;392(10155):1321-1329. PMID: 30292578; Kruger D et al. Diabetes Technol Ther. 2022;24(10):697-711. PMID: 36173236; Messer LH et al. Diabetes Technol Ther. 2022;24(10):712-725. PMID: 36173237. Registry IDs as on the individual cards.