Cardiac resynchronization in narrow QRS: the EchoCRT harm

Shelf #714 · Heart, vessels, blood pressure & clotting

Stamped: development phase, October–November 2026 — exact date not recorded.

Evidence published: 2013.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In heart failure with a narrow QRS but echocardiographic dyssynchrony, does CRT help — or harm (EchoCRT)?

Balance of the evidence

LIKELY (harm / no benefit): EchoCRT (809 randomised patients with QRS <130 ms and echocardiographic dyssynchrony) was stopped for futility and found no reduction in death or heart-failure hospitalisation with CRT switched on (28.7% vs 25.2%; hazard ratio 1.20) and more deaths (11.1% vs 6.4%; hazard ratio 1.81, 95% CI 1.11 to 2.93 — a relative figure inside the 0.5–2.0 band, printed with its absolute rates). A large trial stopped early with a null primary and a mortality signal in the wrong direction carries LIKELY for 'do not do this'. This is the mirror image of Shelf #274, where CRT in wide-QRS heart failure is carded; the wide-QRS trials are not re-carded here. Heart-failure figures are printed with mortality beside them throughout.

MeSH: Cardiac Resynchronization Therapy/methods · Echocardiography · Electrocardiography · Heart Failure, Systolic/diagnosis,mortality,therapy · Hospitalization/statistics & numerical data · Kaplan-Meier Estimate · Stroke Volume

Cards (2)

Card 1 — EchoCRT 2013: CRT in narrow QRS did not reduce death or heart-failure hospitalisation (28.7% vs 25.2%) and deaths were higher (11.1% vs 6.4%) — trial stopped for futility

Study: Randomised trial (all patients implanted; CRT capability switched on or off), stopped early for futility

Population: 809 patients at 115 centres with NYHA class III or IV heart failure, left ventricular ejection fraction 35% or less, QRS duration less than 130 ms, and echocardiographic evidence of left ventricular dyssynchrony, randomised to CRT-on (404) or CRT-off (405); mean follow-up 19.4 months at study closure. The abstract does not state mean age.

Summary: EchoCRT 2013: CRT in narrow QRS did not reduce death or heart-failure hospitalisation ( 28.7% vs 25.2% ) and deaths were higher ( 11.1% vs 6.4% ) — trial stopped early.

Result: The primary outcome (death from any cause or first hospitalisation for worsening heart failure) occurred in 116 of 404 CRT patients and 102 of 405 controls (28.7% vs 25.2%; hazard ratio 1.20, 95% CI 0.92 to 1.57; P=0.15) — printed with its mortality beside it: deaths were 45 vs 26 (11.1% vs 6.4%; hazard ratio 1.81, 95% CI 1.11 to 2.93; P=0.02 — a relative figure). The study was stopped on March 13, 2013 for futility on the data and safety monitoring board's recommendation.

Watch out: The mortality excess is a secondary outcome in a trial stopped for futility — but with a null primary leaning the wrong way (hazard ratio 1.20) there is no reading of this trial in which narrow-QRS CRT helps. Echocardiographic dyssynchrony measures of that era were themselves unreliable, which is part of the lesson.

Source: Ruschitzka F et al. Cardiac-Resynchronization Therapy in Heart Failure with a Narrow QRS Complex. N Engl J Med. 2013;369(15):1395-1405. PMID: 23998714. Registry ID: ClinicalTrials.gov NCT00683696, from the record's databank fields.

Card 2 — Synthesis: a narrow QRS closes the CRT question — do not implant for dyssynchrony alone — LIKELY harm, proposed (Tier C)

Study: Synthesis

Population: Synthesis of the card above: EchoCRT (809 randomised patients).

Summary: A narrow QRS closes the CRT question — do not implant for dyssynchrony alone: LIKELY harm.

Result: EchoCRT asked whether a mechanical picture (echo dyssynchrony) could substitute for the electrical criterion (wide QRS) that every positive CRT trial was built on. It could not: death or heart-failure hospitalisation 28.7% vs 25.2%, deaths 11.1% vs 6.4%. The positive wide-QRS evidence stays home on Shelf #274 (CARE-HF/COMPANION/RAFT/MADIT-CRT) and is pointed to, not re-carded. As a large single trial with a null primary and a concordant mortality signal, it carries LIKELY for the negative claim;.

Watch out: This shelf is about QRS under 130 ms with echo dyssynchrony; it does not argue against CRT in wide QRS, bundle branch block, or pacing-induced cardiomyopathy upgrade situations covered elsewhere.

Source: Synthesis of Ruschitzka F et al. N Engl J Med. 2013;369(15):1395-1405. PMID: 23998714. Registry ID as on the card above.