Respiratory impedance by oscillometry: height drives the adult equations, women run higher resistance than men, and the 1994 equations still in laboratory use misread healthy older adults by −34% to +76%

Shelf #7153 · Lungs, breathing & sleep

PHYSIOLOGY — ESTABLISHED

Stamped: 2026-11-28.

Evidence published: 2013–2021.

Last source check: not yet re-checked since stamping.

Source correction status: not yet checked for source corrections.

Question

In healthy humans, what are normal respiratory impedance values by impulse oscillometry / forced oscillation (resistance at 5 and 20 Hz, reactance, area of reactance) by age, height and sex — and how do values differ between devices and populations?

Balance of the evidence

PHYSIOLOGY — ESTABLISHED. Textbook framing for oscillometry specifically is thin in Boron & Boulpaep's respiratory-mechanics chapters — this shelf rests chiefly on the primary measurement literature and says so. Three primary cohorts from three populations agree on the structure: in 567 healthy Chinese adults from 19 hospitals (Liang 2021), height — not age — was the dominant predictor, and both the equations and the manufacturer's ranges in routine use mis-stated the normal limits; in 397 never-smoking lung-healthy adults aged 45–91 from the population-based KORA cohorts (Schulz 2013), women had higher resistance and lower low-frequency reactance than men, R5−R20, reactance area and resonant frequency rose with age, and the Vogel & Smidt 1994 equations in widespread use erred by −34% to +76% where the new equations erred by −1% to +5% (relative differences); in 75 asymptomatic Norwegians over 70 (Aarli 2013), within-breath impedance (the inspiratory–expiratory reactance difference at 5 Hz) gained its first elderly reference values, with higher resistance than earlier studies reported. Values are carded per-device and per-population: IOS and FOT generations and their signal processing are not interchangeable.

MeSH: Aged · Cohort Studies · Electric Impedance · Female · Humans · Male · Middle Aged · Oscillometry/methods · Reference Values · Respiratory Function Tests

Cards (4)

Card 1 — Liang 2021: reference equations from 567 healthy Chinese adults (19 hospitals) — height, not age, drives IOS indices, and the manufacturer's normal ranges clearly differ from the derived limits

Study: Human normative cohort (multicentre primary population dataset), J Thorac Dis 2021

Population: 567 healthy adults with acceptable impulse oscillometry data, drawn from 1,318 subjects measured at 19 general hospitals across China between 2016 and 2018, with measurements performed to European Respiratory Society recommendations; the study was registered as part of a larger study (NCT03467880).

Summary: Liang 2021: reference equations from 567 healthy Chinese adults (19 hospitals) — height, not age, drives IOS indices, and the manufacturer's normal ranges clearly differ from the derived limits

Result: Sex-specific reference equations with lower and upper limits of normal were developed for the impulse oscillometry indices by multiple linear regression. Height, but not age, was the most influential contributor to the indices. The reference equations then in laboratory use predicted higher R5 and X5 than the new equations, and the normal ranges for R5 and X5 recommended by the equipment manufacturer were clearly different from the upper and lower limits derived from this population — so the device defaults and the population data disagree about who is normal.

Watch out: One population and one device family: the equations are Chinese-adult IOS values, and the headline lesson — manufacturer ranges ≠ population limits — means these numbers must not be exported to other devices or ancestries without local validation.

Source: Liang XL et al. Reference values of respiratory impedance with impulse oscillometry in healthy Chinese adults. J Thorac Dis. 2021. PMID: 34277060. Registry ID: ClinicalTrials.gov NCT03467880, from the record's databank fields. MeSH: not captured from the record

Card 2 — Schulz 2013 (KORA): in 397 lung-healthy adults aged 45–91, women had higher resistance and lower reactance than men; the 1994 Vogel & Smidt equations erred by −34% to +76% (relative) where new equations erred by −1% to +5%

Study: Human normative cohort (population-based primary dataset, advanced age), PLoS One 2013

Population: 397 never-smoking, lung-healthy subjects with normal spirometry (243 women, 154 men), identified within a population-based sample of 1,990 subjects aged 45–91 years from the KORA cohorts (Augsburg, Germany); sex-specific quantile regression models used age, height and body weight as predictors.

Summary: Schulz 2013 (KORA): in 397 lung-healthy adults aged 45–91, women had higher resistance and lower reactance than men; the 1994 Vogel & Smidt equations erred by −34% to +76% (relative) where new equations erred by −1% to +5%

Result: Women showed higher resistance values than men, while reactance at low frequencies (up to 20 Hz) was lower (P < 0.05). A significant age dependency was found for R5−R20 (the difference between resistance at 5 and 20 Hz), the integrated low-frequency reactance area (AX) and resonant frequency in both sexes, and for reactance at 5 Hz (X5) in women only. Against the healthy subjects, mean differences between observed and predicted resistance (5 and 20 Hz) and reactance (5 Hz) ranged from −1% to +5% (relative) with the new model, versus −34% to +76% (relative) with the then-current Vogel & Smidt 1994 equations.

Watch out: The error ranges are relative differences between observed and predicted values in this cohort, printed with that meaning; the cohort is Caucasian and aged 45+, so the equations fill the older-age gap rather than replacing adult equations at younger ages.

Source: Schulz H et al. Reference values of impulse oscillometric lung function indices in adults of advanced age. PLoS One. 2013. PMID: 23691036. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Cohort Studies · Female · Humans · Male · Middle Aged · Oscillometry/methods · Reference Standards · Respiratory Function Tests

Card 3 — Aarli 2013: first within-breath impedance reference values in 75 asymptomatic elderly over 70 — whole-breath and inspiratory/expiratory reactance at 5 Hz by forced oscillation

Study: Human normative cohort (within-breath FOT/IOS, elderly), Clin Respir J 2013

Population: 75 healthy non-smoking participants aged over 70 years with normal spirometry and at least two valid forced-oscillation measurements, selected from 148 elderly participants drawn as an age- and sex-stratified random sample from the population of Bergen, Norway; measurements used an impulse oscillometry system.

Summary: Aarli 2013: first within-breath impedance reference values in 75 asymptomatic elderly over 70 — whole-breath and inspiratory/expiratory reactance at 5 Hz by forced oscillation

Result: Predictive equations for men and women were created for whole-breath and within-breath impedance parameters by linear multiple regression, including the within-breath reactance difference at 5 Hz (ΔX5 — the inspiratory-to-expiratory swing used to detect expiratory flow limitation). Resistance measurements were higher than those reported in previous studies, and frequency dependence of resistance was significantly larger in this elderly sample.

Watch out: N = 75 after validity filtering from 148 selected — a small, single-city elderly sample; within-breath indices are device- and algorithm-specific (the ΔX5 method derives from the Dellacà forced-oscillation work), so these values travel with the technique, not across oscillometry platforms.

Source: Aarli BB et al. Reference values for within-breath pulmonary impedance parameters in asymptomatic elderly. Clin Respir J. 2013. PMID: 22822726. Registry ID: none in the record. MeSH: Aged · Aged, 80 and over · Airway Resistance/physiology · Electric Impedance · Exhalation/physiology · Female · Forced Expiratory Flow Rates/physiology · Humans · Inhalation/physiology · Linear Models · Male · Oscillometry/methods, standards · Predictive Value of Tests · Reference Values · Reproducibility of Results · Respiratory Function Tests/standards · Spirometry

Card 4 — Synthesis: oscillometry norms are real but local — height- and sex-structured, age-sensitive in the reactance indices, and only valid for the device and population that produced them

Study: Synthesis

Population: Synthesis of the three carded human cohorts above (Liang 2021; Schulz 2013; Aarli 2013). Boron & Boulpaep framing for oscillometry is thin, as flagged in the balance; the forty-year review of adult impedance reference values in the wider literature is consistent with this per-population picture but is a review and is not carded.

Summary: Synthesis: oscillometry norms are real but local — height- and sex-structured, age-sensitive in the reactance indices, and only valid for the device and population that produced them

Result: The three cohorts triangulate one discipline. Structure: resistance falls and reactance rises with height (Liang — height the dominant adult predictor), women run higher resistance and lower low-frequency reactance than men (Schulz), and the frequency-dependence and reactance indices (R5−R20, AX, resonant frequency, X5) drift with age even when R5 itself does not (Schulz; Aarli's elderly within-breath values). Validity: in every cohort the inherited standard failed somewhere measurable — manufacturer ranges (Liang), the 1994 equations' −34% to +76% relative errors in older adults (Schulz), absent within-breath norms over 70 (Aarli). REFUSAL, recorded per the standing rule: a verified Indian-adult IOS reference cohort (Moitra et al., Int J Tuberc Lung Dis 2020) was refused as an evidence card because its PubMed record carries an EMPTY abstract and its numbers cannot be verified from the record; it is named here as refused, not carded, pending a full-text pull.

Watch out: Never pool IOS and FOT values or apply one population's limits to another: device generation, signal processing and population each shift the normal range, which is this shelf's central finding rather than a caveat.

Source: Synthesis of Liang XL et al., J Thorac Dis 2021 (PMID: 34277060); Schulz H et al., PLoS One 2013 (PMID: 23691036); Aarli BB et al., Clin Respir J 2013 (PMID: 22822726). Registry IDs as on the individual cards.